1/222
Flashcards for Nuclear Medicine Instrumentation Review
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Scintillation cameras
Produce projection images of the distribution of radioactivity in patients; sometimes called gamma cameras or Anger cameras.
Photopeak width
is measured as full width half maximum(FWHM)
Energy resolution
The broadening of photopeak (FWHM) is termed
100%
The keV is 100 how many percentage is detected
92%
The keV is 140 how many percentage is detected
54%
The keV is 200 how many percentage is detected
22%
The keV is 300 how many percentage is detected
6%
The keV is 500 how many percentage is detected
10%
When the gamma ray is completely absorbed by the NaI , some of this energy converted into light how many percent?
Photopeak
is when an incident gamma ray is completely absorbed
10mm thick
For gamma ray imaging, NaI scintillators are ~ thick
Ga67 and 111IN
Medium-energy collimators
Septa
The lead strips between the holes are called
55 PMTS
Scintillation cameras typically use
Parallel
Scintillation rays emerging from the patient
pass through a lead collimator that only allows photons traveling ____ to the collimator holes to reach the scintillator
4-6mm
Pinhole the typical aperture or diameter
multihole/parallel hole collimator
The workhorse of the collimator
Low energy collimator
used with 99m Tc and 201 TI have thin septa.
PHA
can be set to allow only selected energies to be counted, and reduce the number of Compton scatter photons in the image
PHA analysis
maximizes the number of photopeak events while minimizing the detected photons that would degrade image quality (i.e., Compton scatter)
NM images
can be viewed in real time on a display monitor during the acquisition
Nm images
can be viewed in real time on a display monitor during the acquisition
3 or 6
Projection images are obtained at selected angles, typically every ____ or ___ degrees
Iterative reconstruction algorithm
New used algorithms
CT SCAN
scan can be a low-dose scan for image coregistration and attenuation correction only
Higher dose CT scan
scans may be acquired for diagnostic imaging
lutetium oxyorthosilicate (LSO) or gadolinium oxyorthosilicate (GSO)
Modern PET scanners use detectors made of
Lutetium yttrium oxyorthosilicate
is also used, which is LSO doped with a small amount of yttrium
BGO AND LSO
have similar gamma ray absorption properties
GSO
is a markedly poorer absorber of 511-keV gamma rays
GSO AND LSO
are inorganic scintillators that emit more light than BGO
improves energy resolution
Increased light output of GSO and LSO
Increased light
output of GSO and LSO improves energy resolution
organic scintillators
emit their light much faster than those of BGO.
Shorter decay times
improve count rate performance (reduced dead time)
thick detectors
To efficiently detect 511-keV annihilation photons, _______are used (i.e., 20–30 mm)
20-30 mm
To efficiently detect 511-keV annihilation photons, thick detectors are used (?mm)
coincidence event
Two interactions occurring within a specified time interval τ (coincidence timing window) are called a
prompt coincidence
• ACD produces an output whenever two events are recorded within a specified coincidence timing window. Generically, any such events are called
extrinsic flood images
are obtained with the collimator in place and will assess the system performance including the collimator
intrinsic floods
are performed without the collimator and assess the performance of NaI crystal and associated light detectors
quadrant bar phantom
Resolution (i.e., the ability to separate two points) is checked using a
four sets of parallel bars
Quadrant bar phantoms have ________, with each rotated through 90 degrees, with dimensions of 3.5, 3.0, 2.5, and 2.0 mm
90
Quadrant bar phantoms have four sets of parallel bars, with each rotated through ___degrees, with dimensions of 3.5, 3.0, 2.5, and 2.0 mm
3.5, 3.0, 2.5, and 2.0 mm
Quadrant bar phantoms have four sets of parallel bars, with each
rotated through 90 degrees, with dimensions of ______
Bar pattern phantom
also check for linearity (i.e., ability to image straight lines)
high energy collimators
Are required for 131 I imaging and have the thickest septa
Iodine escape peak
occurs at approximately 30 keV below the photopeak and is the result of characteristic K-shell x-rays from iodine that escape the crystal
PHOTOMULTIPLIER TUBES
Light output from the NaI crystal is detected by
window
measured by percent, determines the acceptable range of energies around the peak for subsequent counting
128×128
A typical NM matrix size is
64 × 64
Cardiac imaging uses matrix size of
1024 × 1024
Whole body imaging uses matrix size
two byte memory
The number of counts in each pixel in a NM
image is stored using a
10-⁴, or only 0.01%
Collimator sensitivity is low with approximately
500,000
The counts in images can vary over a wide
range, with a median value of
Single Photon Emission computed tomography(SPECT)
provides computed tomographic views of the three-dimensional distribution of radioisotopes in the body.
Parallel hole collimator
are commonly used for SPECT imaging
180 and 360
Scintillation cameras rotate ___ or __ degrees around the patient.
Scan projection
were originally used as inputs for filtered-back projection reconstruction algorithms to compute tomographic images
Multiheaded cameras
are used to increase
system sensitivity and reduce scan times
elliptical orbits
(i.e body contouring)for scintillation camera traveling around the patient allows the distance to the patient to be minimized
111In octreotide
for neuroendocrine imaging
111In labeled ProstaScint
for prostate imaging
67Ga
for infections
Bismuth Germanate
Most early-generation PET scanners had detectors made of
Collimators
Essential for providing spatial information in planar NM imaging.
Count
The registration of a single gamma ray by the detector; ~500,000 are acquired for a typical scintillation camera image.
Collimators
Typically made of lead and contain multiple holes; lead strips between holes are called septa.
Pinhole collimator
Simplest form of collimator, consisting of a single, small hole or aperture.
Multihole collimator
Provides substantially better geometric sensitivity compared to the pinhole collimator; consists of a very large number of parallel holes.
Converging multihole collimator
Provides both enhanced spatial resolution and improved sensitivity by focusing the direction of the holes at a point some distance from the collimator surface.
Diverging collimators
Project an image size that is smaller than the object size, and FOV increases with distance.
Collimator sensitivity
The fraction of gamma rays reaching it from all directions that pass through the holes.
NaI Scintillators
Detect gamma rays emerging from patients and are generally rectangular; very fragile and easily damaged.
Detection efficiency
Percentage of incident gamma rays absorbed in the scintillator.
Pulse height analyzer (PHA)
Electronic device used to determine which portion of the detected spectrum is used to create images.
Analog to digital converters
Used to generate the digital information for NM images.
Single photon emission computed tomography (SPECT)
Provides computed tomographic views of the three-dimensional distribution of radioisotopes in the body.
PET camera
Contains rings of detectors (scintillators) surrounding the patient.
Random coincidence (Accidental coincidence)
occur when annihilation photons from two unrelated positron annihilation events are detected in two different detectors, within the coincidence timing window, and recorded as a single coincidence event.
Sinogram
Line of response data are used to create a
filtered back projection
was popular in the early days of PET
but is now rarely used
depth independent
Attenuation in PET is____ and depends only on the total thickness of tissue traveled.
CT images
are used to generate attenuation correction factors for use in PET
ordered-subset expectation maximization
Images are reconstructed using statistically based algorithms such as
Time of flight (TOF) PET
Measuring the difference in arrival times of the two annihilation photons from an annihilation is used
TOF information
can be used in the reconstruction process to improve image quality including improved spatial resolution as well as enhanced lesion contrast
TOF PET
can identify the location of an annihilation event with an uncertainty that corresponds to a Full Width Half Maximum of ~7.5 cm
7.5 cm
TOF PET can identify the location of an annihilation event with an uncertainty that corresponds to a Full Width Half Maximum of
PET images
typically have several million counts
Several million
PET images typically have ______ counts
F18
The most common positron emitter used for PET imaging is
110 minutes
HALF LIFE OF F18
flourodeoxylglucose
18F in the form of ________ is the most commonly used agent