1/47
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Scintillation Counter
Instrument that detects radiation using a crystal or other material that emits light when struck by ionizing radiation
Measured during a QC test on a dose calibrator
Accuracy, Linearity, Geometry, and Constancy
Frequency of testing Accuracy of a dose calibrator
Yearly
Frequency of testing Linearity on a dose calibrator
Quarterly (every 3 mo)
Frequency of testing Geometry on a dose calibrator
Installation or after moving the dose calibrator
Measured by a QC test on a scintillation detector
Constancy, Calibration, Chi-Square, Energy Resolution
Radiation Source used to test a surgical gamma probe
Co-57 due to its energy (122 keV) and half-life (270 days)
Radionuclide to be used with a collimator that has thinner septa between its holes
P-32, because it is a Beta emitter and not used in diagnostic imaging (24, 305 keV)
Do before returning a package with radioactive materials
Wipe test/survey to ensure no surface contamination
Limit for surface exposure rate at which you must notify the NRC and final delivery carrier of radioactive materials
200 mR/hr
Limit for exposure rate at 1 meter away from surface at which you must notify the NRC and final delivery carrier of radioactive materials
10 mR/hr
Measuring a window of a an energy level (eg 10% for 159 keV)
5% below and 5% above
In this case, 151.05 and 166.95 keV
Formula for sensitivity
(Source cpm - background cpm) / Source activity in uCi
Composition and function of a PLES phantom
Two lead bars of the same width and spacing. Can take two images at 90o of each other to achieve linearity over an entire FOV.
NRC time limit on personnel radiation exposure records
Indefinite
Orthogonal hole transmission phantom
Produces a “mesh-like” QC image when used with a scintillation camera
Camera movement that improves Spatial Resolution during SPECT imaging
Elliptical and/or Body Contour orbits
Distance at which exposure rate (mR/hr) is measured according to transport index
1 meter from the surface of a package
Maximum radiation dose that one would receive in an unrestricted area
2 mrem/hr
Formula for measuring exposure rate on a GM meter
Meter reading x Adjustment setting (knob)
Frequency of surveying areas where therapeutic radiopharm’s are prepared or administered
End of Day whenever a therapeutic dose has been prepared/administered
Therapeutic doses are quite high and require written directive as well as a daily survey
Frequency of calibrating a survey meter (GM meter, scintillation probe, ionization chamber)
Yearly
Difference between a Direct Radiation Survey and a Wipe Check Radiation survey
Direct surveys provide an instant reading, while a Wipe Check will need to go in a well counter to get a reading
NRC limit for the yearly occupational radiation dose of the lens/eye
15,000 mRem/year (15 rem/year)
If a worker receives close to the limit some occupational radiation dose in one year, what should be done?
Review ALARA protocols and dedicate as many protective measures as possible during work hours
Done before daily use of the GM meter
Check the batteries
Average energy needed to create an ion pair in an ionization chamber
34 eV per pair (20 - 45 eV depending on type of gas in chamber)
Geiger Effect
Ionization of the entire fixed volume of gas in a detector chamber due to secondary ionization
This occurs in the Geiger-Muller region of the voltage region chart (highest applied voltage and collected charge)
Pulses produced by the collection of ions in the Proprtional Region vs those produced by an instrument operating in the Ionization Region
The Proportional Region of the gas chamber operates at a higher current, so the pulses from this region tend to be larger and more ions are created
Region at which gas amplification occurs in the chamber of a radiation detector
Proportional Region of the gas curve
Pulse Height
Spectrometer representation of an amount of radiation detected in the chamber of a radiation detector. It is dependent on the number of ions produced by interaction of the gas and incoming radiation
Jaszczak Phantom
20 cm plexigalss cylinder with rods and spheres inserted. Used to determine SPECT system performance.
Full Width at Half Maximum (FWHM)
Measure of Spatial Resolution, based on possible positions of the source when the count rate is half of its maximum
A smaller FWHM is indicative of a higher resolution, so a gamma camera can better differentiate between two radioactive sources
Cause of ring artifacts in tomographic ring construction
Abnormalities of the detector OR rotational mechanics in SPECT imaging
These are usually seen with uncorrected non-uniformities in the flood field or defects in the collimator
Frequency of measuring the Activity Linearity of a dose calibrator
Quarterly (every 3 mo)
Time for retaining records of surveys
3 years
Best reseolution for a parallel-hole collimator
At the surface of the collimator. Greater distance between the PHC and your object degrades the test image.
Frequency of checking Center of Rotation (COR) for SPECT imaging
At least weekly, until the COR system is shown to be several over a time frame of several months
Best collimator for high resolution images of a small organ, gland or joint
Pinhole Collimator
The PC is a large, hollow cone with a single aperture at the point. This allows it to magnify images of small organs, etc.
Activity needed to evaluate intrinsic uniformity of a planar imaging camera (using Tc 99m)
200 uCi
An intrinsic flood test is performed without a collimator, and at a distance of at least 5x the largest camera dimension from the camera face
Two most important Quality Assurance measurements for a scintillation camera system
Field Uniformity to measure detector performance
Spatial Resolution to measure image quality
Light generated by a NAI crystal in a scintillation detector
A result of the photon energy absorbed in the crystal from output from a photomultiplier tube (PMT)
Scintillator material creates light when met with gamma rays. This light is converted to an electrical signal by the PMT
Basic function of ANY collimator
Limit access of photons to the instrument, thereby “filtering” the gamma rays that will pass through to the scintillator crystal
True/False: All survey meters/radiation detectors (that we use) contain gas detecors
True!
Common cause of field non-uniformity
Misadjusted photopeak
Geometry
Measurement of the sensitivity of the dose calibrator to different source configurations
Correction factors are used when comparing dose activities in different sample volumes or syringe sizes
Limit for activity levels of multiple radioactive materials in a package shipped back to a pharmacy
The lowest activity of any of the materials in the package
Scintillation
Light emitted by some material after it absorbs ionizing radiation