primary, secondary, filament
3 primary sections of the xray circuit
primary circuit
consists of the main power switch (connected to
the incoming power supply), circuit breakers, the autotransformer, the timer circuit, and the primary side of the step-up transformer
secondary circuit
consists of the secondary side of the step-up
transformer, the mA meter, a rectifier bank, and the x-ray tube (except
for the filaments).
filament circuit
a rheostat, a step-down transformer, and the filaments
main power switch
simply an on-off switch for the unit and is connected to the power supply of the facility
line compensator
wired autotransformer that automatically adjusts the power supply to keep it at 220 volts
circuit breaker
protects against short circuits and electric shock
autotranformer
adjustable transformer controlled by the kVp selector on the operation console
step-up transformer
used to increase the voltage from the autotransformer to the kilovoltage necessary for xray production
step-up transformer
this is the divivding line between the primary and secondary circuit
timer circuit
exposure timer that makes it easier to control a low voltage than and very on
secondary circuit
consists of the secondary side of the step-up transformer, the mA meter, a rectifier bank, and the x-ray tube (except for the filaments)
filament circuit
consists of a rheostat, a step-down transformer, and the filaments.
mA meter
a device placed in the secondary circuit that monitors x-ray tube current
rectifiers
converts ac to dc in the secondary unit
anode to cathode
in the xray tube current must flow from….
solid state rectifier
rectifier commonly used in todays xray circuit that is made of 2 semi conducting crystals (p-type and n-type)
rheostat
begins the filament circuit and is controlled by the mA selector
mA
“tube current” that reflects the rate of flow of electrons passing through the xray tube during an exposure
exposure timer
works in concert with the rheostat
rheostat
controls filament temperate and the rate at which electrons are boiled off of the filament
exposure timer
determines the duration of the rheostat
together these two determine the quantity of electrons boiled off of the filament
step-down transformer
used in the filament circuit to increase the current by reducing the voltage that is applied to the filament
filaments
the final piece of the filament circuit
small focal spot
there are 2 filaments one signifying a large focal spot and a
boil off electrons
the goal of the filament circuit is to ….. out of the filament wire
5-7
a large filament current of …. amps is required to produce a tube current in the range of milliampres
thermionically emitted electrons
…. are needed for xray production and represent one more step in our manipulation of electricity
filament circuit
in this circuit we are boiling off electrons off a very small wire and must precisely control the current that is applied
damaged
if the current is too high on the small wire of the filament circuit, the tiny wire will be
kVp
this adjusts the autotransformer and determines the number of turns on the secondary side necessary to produce a voltage throught self induction
step up….mutual induction
the …. transformer increases this voltage by a fixed amount and
through ….. produces the kilovoltage selected on the
operating console
solid state rectifiers
used to route electricity through the xray tube correctly
technique
the process begins when the radiographer selects a
anode, cathode
After passing through the rectifiers, the electricity creates a large
positive charge on the ….. of the x-ray tube and a large negative
charge on the ….. focusing cup
filament circuit, rheostat
the ……. draws electricity from the autotransformer, which then travels to the …..
step down transformer
from the rheostat, electricity then travels to the
filament
the adjusted current from the step down transformer then travels directly to the ….
focusing cup
the filament is located with in the
boiled off
the filament heats up to the point where the electrons are boiled off
cloud of electrons
the filament circuit creates a
anode, cathode
the kilovoltage applied to the xray tube created a large positive charge on the …. and a large negative charge on the ….
kinetic energy
the large positive change attracts the electrons boiled off, giving them
crowded together
the large negative charge on the cathode serves to keep the electrons…
timer circuit
The electrons travel across to the anode and interact there to produce x-rays until the …. terminates the process