Which of the following will degrade the display of color Doppler information on the image?

Increased sample size

2
New cards

Which type of resolution improves with decreased image sector size?

Temporal

3
New cards

Frequencies used in TTE imaging are usually in the range of?

2.5 - 4 MHz

4
New cards

Frequencies used in TEE imaging are usually in the range of?

5 - 7 MHz

5
New cards
<p>What acoustic phenomenon causes the abnormal appearance of the LV?</p>

What acoustic phenomenon causes the abnormal appearance of the LV?

Refraction

6
New cards

Why can’t TEE probes be thermally sterilized?

If the probe is heated above the Curie temperature the PZT properties will be lost

7
New cards

Doppler shift ____ to the blood flow velocity

Directly proportional

8
New cards

Increased overall gain levels on an ultrasound exam can cause?

Underestimation of valve area on planimetry

9
New cards

On the ROI on a 3D system the x-axis represents?

Width of the box

10
New cards

On the ROI on a 3D system the y-axis represents?

Height of the box

11
New cards

How do you correct beam width artifact on a Doppler tracing?

Use a smaller sample size

12
New cards

Which of the following describes a purpose for applying acoustic gel to the face of the transducer?

To reduce the difference in impedance between the matching layer and the skin

13
New cards

The standard sector width for TTE is?

60 degrees

14
New cards

Why is the temperature of the TEE probe monitored?

If the PE element is heated over 40 degrees celsius it would cause damage to the tissues of the esophagus/stomach

15
New cards

Which of the following correctly describes the calibration procedure for the ultrasound display motion?

Adjust brightness and contrast levels until the first block on the color bar blends into the overall screen color

16
New cards

Which of the following controls adjusts the number of shades of gray displayed on the 2D image?

Dynamic range

17
New cards
<p>What Doppler artifact is demonstrated on the image? </p>

What Doppler artifact is demonstrated on the image?

Beam width

18
New cards

Which operator control can be used to adjust the image contrast?

Dynamic range

19
New cards

____ adjusts the brightness of the entire image

Overall gain

20
New cards

____ adjusts the brightness of specific portions of the image

TGCs

21
New cards
<p>Which of the following is a potential cause for the abnormal swirling appearance of the UEA on the video?</p>

Which of the following is a potential cause for the abnormal swirling appearance of the UEA on the video?

Low MI settings

22
New cards

If you increase the power output on the machine?

Beam intensity and beam penetration increases

23
New cards

Increased overall gain levels on an ultrasound exam can cause?

Underestimation of valve area on planimetry

24
New cards

While performing a Doppler evaluation of the AV, a significant amount of wall motion artifact is displayed on the Doppler tracing. How can you eliminate the appearance of the artifactual information on the Doppler tracing while maintaining accurate display of the aortic waveform?

Increase the high pass filter settings

25
New cards

Which of the following controls adjusts the number of shades of gray displayed on the 2D image?

Dynamic range

26
New cards

You suspect a calcification is present within the LV. You activate harmonics and add a focal zone to the field of view. Which type of resolution will degrade with these changes?

Temporal

27
New cards

No flow is detected across the septum in the subcostal view when evaluating an adult patient with a known ASD. Which of the following could explain the lack of flow displayed across the defect?

Color Doppler PRF is set too high

28
New cards

What effect will a MI of 1.0 have on an echo exam with UEA?

The increased beam intensity will burst the UEA bubbles quickly

29
New cards

What system control can reduce the appearance of clutter on a color Doppler image?

Decreasing color priority

30
New cards
<p>How would you correct the issue with the Doppler tracing?</p>

How would you correct the issue with the Doppler tracing?

Increase the PRF

31
New cards

The interference of reflected sound waves is necessary for image creation in what technique?

Speckle tracking

32
New cards

TEE transducers are considered ____ devices that should be disinfected using ____

Semi-critical and high level disinfectant

33
New cards

Which of the following controls adjusts the number of shades of gray displayed on the 2D image?

Dynamic range

34
New cards

Which operator control can be used to adjust the image contrast?

Dynamic range

35
New cards

Which of the following is related to degraded axial resolution on the image?

Decrease in probe frequency

36
New cards

Which operator control can be used to adjust the image contrast?

Dynamic range

37
New cards

3D surface rendering is preferred for evaluating?

Valve characteristics

38
New cards

3D volume rendering is preferred for evaluating?

Segmental wall motion

39
New cards

While scanning an echo, you change the gray map to a new choice and continue to scan. This would be termed a change in?

Preprocessing

40
New cards

Adjustment of which of the following imaging controls would be best to assist in improving visualization of the descending AO posterior to the LA on an obese patient?

Move focal zone to the level of AO

41
New cards

Which of the following is most commonly associated with a false positive diagnosis of a vegetation on the native AV?

Beam width

42
New cards

What type of magnification provides the highest frame rate and best image resolution?

Preprocessing

43
New cards

A high frame rate (> 30 frames per second) is preferred for cardiac imaging?

For accurate display of cardiac motion

44
New cards

What is the best way to reduce or eliminate clutter and cross talk on a Doppler tracing?

Increase the wall filter

45
New cards
<p>How can a sonographer correct for blossoming on a spectral Doppler display? </p>

How can a sonographer correct for blossoming on a spectral Doppler display?

Turn the Doppler gain down

46
New cards

Which operator control can be used to adjust the image contrast?

Dynamic range

47
New cards

While scanning an echo, you change the gray map to a new choice and continue to scan. This would be termed a change in?

Preprocessing

48
New cards

Which of the following is related to degraded axial resolution on the image?

Decrease in probe frequency

49
New cards

Which of the following adjustments will improve temporal resolution on color Doppler?

Decrease sample size

50
New cards

What is the best way to reduce or eliminate clutter and cross talk on a Doppler tracing?

Increase the wall filter

51
New cards

Which of the following cardiac exam techniques provides the best spatial resolution of the cardiac structures?

TEE

52
New cards

Which of the following will aid in visualization of the endocardium?

Increased transducer frequency

53
New cards

Which acoustic property causes the echogenic appearance of the pericardium related to surrounding structures?

Specular reflection

54
New cards

Which artifact is caused by non-specular reflections from a rough surface and wave interference?

Speckle

55
New cards

What refers to shadowing caused by mechanical prosthetic valve apparatus that inhibits flow display on a color Doppler image?

Masking

56
New cards

Which of the following technical factors can negatively affect the 2D planimetry evaluation of the MV?

Lateral resolution and gain settings

57
New cards
<p>What color Doppler artifact is demonstrated on the image? </p>

What color Doppler artifact is demonstrated on the image?

Flash

58
New cards

Which of the following will improve the lateral resolution on the image?

Increase the focal zone number

59
New cards

Aliasing occurs when the Doppler shift?

Exceeds ½ the PRF

60
New cards

Which of the following adjustments would improve the appearance of an UEA on the echo image?

Decrease output power

61
New cards

On an apical four chamber view, if you move the color Doppler sample box from the mitral level to the pulmonary veins

Frame rate decreases