1/39
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
quality assurance
the routine, periodic evaluation of an ultrasound system to guarantee optimal image quality
how often do quality assurance evaluations have to be performed?
periodically and routinely
what are the 4 requirements for a quality assurance program?
assessment of system components
repairs
preventative maintenance
record keeping
what are the 5 goals of a quality assurance program?
guarantee proper operation of the system
detect gradual changes
minimize downtime
reduce the number of non-diagnostic exams
reduce the number of repeat scans
what 3 devices enable the sonographer to perform quality assurance?
tissue equivalent phantom
doppler phantom
beam profile/slice thickness
what are the proper methods for quality assurance programs?
test under known, defined conditions
use constant instrument settings
use a phantom with measurable characteristics
image in an identical environment
objective standards
completely unbiased
not affected by an individual’s previous experience, preference or taste
factual, repeatable and able to be counted
subjective standards
influenced by an individual’s experience or beliefs
often cannot be verified using concrete facts and figures
what type of standards should quality assurance be based on?
objective standards
tissue equivalent phantom
has ultrasonic features similar to soft tissue
what are tissue equivalent phantoms used to evaluate?
used to evaluate gray scale and tissue texture, and multi-focus and adjustable-focus phased array transducers
tissue equivalent phantoms are similar to soft tissue in the following ways:
speed of sound is 1,540 m/s
attenuation
scattering characteristics
echogenicity
what are tissue-equivalent phantoms made out of?
graphite-filled aqueous gels or urethane rubber materials
graphite particles act as ultrasound scatterers in materials
doppler phantom
evaluate doppler systems
include a circulation pump which propels fluid through vessels embedded in a tissue equivalent phantom
what do doppler phantoms assess?
assesses the characteristics of all doppler modalities included pulsed, continuous wave, color and power mode
types of doppler phantoms:
vibrating string
moving belt
flow phantom
slice thickness phantom
contains a diffuse scattering plane that is at an angle to the incident sound beam
mimics soft tissue
what do slice thickness phantoms assess?
the slice thickness and its effect upon image accuracy
sensitivity
the ability of a system to display low-level echoes
what are the 2 forms of sensitivity?
normal
maximum
normal sensitivity
these settings are those at which all the pins, solid masses and cystic structures in the test phantom are accurately displayed
output power, TGC and amplification are adjusted to establish this type of sensitivity
at what type of sensitivity are all subsequent quality assurance and performance measurements made?
at the normal sensitivity
maximum sensitivity
evaluated with the output power and amplification of the system set to the maximum practical levels
maximum visualization depth is used to assess:
sensitivity, and should not differ from one routine evaluation to the next
dead zone
the region close to the transducer where images are inaccurate
extends from the transducer to the shallowest depth from which meaningful reflections appear
info within this region is unreliable and may not be used in the diagnostic setting
the dead zone results from:
the transducer ringing and the time it takes the system to switch from the transmit to the receive mode
what type of transducers have a thinner dead zone?
higher frequency transducers
an increasingly deeper dead zone may indicate:
cracked crystal
detached backing material
longer pulse duration
registration accuracy
the ability of the system to place reflections in proper positions while imaging from different orientations
range accuracy
AKA vertical depth calibration
describes the system’s accuracy in placing reflectors at correct depths located parallel to the sound beam
horizontal calibration
the system’s ability to place echoes in their correct position when the reflectors are perpendicular to the sound beam
digital calipers should be checked in both the:
vertical and horizontal directions
focal zone
the depth at which the intensity is the highest and the beam is the narrowest
the focus of phased array transducers must be carefully evaluated since they are adjustable by the sonographer
axial resolution
the smallest distance at which two pins positioned parallel to the sound beam are displayed as two distinct echoes
axial resolution is evaluated by:
scanning sets of successively closer spaced pins within the phantom
lateral resolution
the minimum distance at which two side-by-side pins are displayed as two distinct images
what is another method to evaluate lateral resolution?
measure the width of reflections on the display that are created by points targets in the phantom
compensation operation or uniformity
uniformity describes the system’s ability to display similar reflectors in the phantom
with proper TGC, identical reflectors should have the same appearance on the monitor, regardless of their depth
mock cysts and solid masses
the tissue equivalent phantom is used to evaluate the dimension, texture and filling of cysts
the system should accurately display hollow structures as anechoic and solid structures and hyperechoic
display, hardcopy output and gray scale dynamic range
adjusting the system’s output power and amplification should produce changes in the gray scale display
it is important to compare the relationship between the image on the system’s screen with the output of all other display devices such as remote viewing stations