what are the difference between calibration sensitivity and analytical sensitivity

calibration sensitivity is the slope of a calibration curve at the concentration of interest while analytical sensitivity is a response to the noise ratio

2
New cards

how to calculate calibration sensitivity

y=mx+b (m is the calibration sensitivity)

graph concentration vs mean analytical signal

3
New cards

how to find the analytical sensitivity

As (concentation) = calibration sensitivity/SD [ ]

4
New cards

how to find the coefficient of variation for the mean of each of the replicate sets

CV = SD/Average x 100

5
New cards

how to find the detection limit for method?

3.3*SD(blank)/calibration sensitivity

6
New cards

voltage in a series

Vt = Vc + Vb + Va

7
New cards

Voltage in a parallel circuit

1/Veff = 1/Va + 1/Vb

8
New cards

Capacitance in a series

1/Ceff = 1/Ca + 1/Cb

9
New cards

capacitance equals

Q/V

10
New cards

if a spectrum is the result of a single scan while another is the result of multiple scans, how man individual spectra were summed to increase S/N from 4.2 to 43?

a) formula?

b) answer

a) (S/N)a = sqrt(n) (S/N)b

b) 105

11
New cards

what type of noise can be reduced by decreasing the temperature of a measurement

thermal noise

12
New cards

what type of noise can be reduced by decreasing the frequency used for the measurement

flicker noise

13
New cards

what type of noise can be reduced by decreasing the bandwidth of the measurement

environmental noise

14
New cards

What would be an appropriate source and detector to determine the emission lines for lithium from 200 to 780 nm?

flame and photo multiplier tube (PMT)

15
New cards

you've been asked to measure the UV absorbance of a sample containing poly cyclic aromatic hydrocarbons, a byproduct of the incomplete combustion of organic matter. What would be an appropriate material for the sample container?

fused silica or quartz

16
New cards

which of the following changes will result in a decreased effective bandwidth?

a) increasing the number of blazes in the monochromatic grating

b) switching from a short-pass filter to a long-pass filter

c) increasing sample concentration

d) decreasing the amount of stray light

a

17
New cards

what would be the distance between blazes for a grating where the first order diffraction line for 600 nm light is reflected at 10 degrees and the angle of incidence light is 45 degrees?

a) formula

b) answer

a) n(wavelength) = distance*(sin(i)+sin(r))

b) 680 nm

18
New cards

which of the following would be considered a discontinuous wavelength selector?

a) quartz prism

b) Czerny-Turner Grating (monochormator)

c)interference filter

d)glass prism

c

19
New cards

which characteristics are desirable for a radiation transducer?

a) high sensitivity to short wavelength light

b) low S/N ratio

c) consistent high output when not illuminated

d) constant response over large wavelength range

e) all of the above

d

20
New cards

what is the appropriate process for atomization for AAS and AES? (3 steps)

desolvation, vaporization, atomization

21
New cards

the doppler effect is one reason for line broadening in atomic spectroscopy. What is the effect on a measured signal?

movement of excited atoms toward the detector causes a shortening in the wavelength of the emitted light

22
New cards

it is better to select an excitation source for AAS with ____bandwidth

smaller

23
New cards

what is the general layout for AES?

sample --> source --> wavelength selector --> detector

24
New cards

a)name the electrical signals that are considered analog?

b) how is the information encoded in an analog signal?

a) current, voltage, charge, and power

b) an analog signal is a signal where the amplitude of the signal changes with time, so the information is encoded either through the amplitude or frequency of the signal (and the signal can be wavelength, magnetic field strength, T,etc)

25
New cards

what is a figure of merit?

it is a numerical value characterizing the quantitative instrument performance and include precision, bias, sensitivity, detection limit, dynamic range and selectivity

26
New cards

How long would it take to discharge a 0.015 microF capacitor to 1% of its full charge through a resistance of 10 mega Ohms?

a) formula

b) answer

a) vc= Vc e^(-t/RC)

b) t = 0.69 seconds

27
New cards

select the type of ionization technique used in MS that would result in significant fragmentation

a) MALDI

b) electron impact

c) high performance

d) chemical ionization

c

28
New cards

which is true for chemical ionization?

a) considered a "hard ionization" technique

b) analyte reacts with electron beam

c) compared to other ionization techniques, less fragmentation occurs

d)compared to other ionization techniques, results in the most repeatable small fragments

c

29
New cards

true or false

formation of ions in electrospray occurs by spraying the sample through a capillary, where it is exposed to a high voltage source, then charged sample undergoes desolvation. Once the Rayleigh limit is achieved, coulombic repulsions overcome any surface tension of the liquid resulting in small, multiply charged species

true

30
New cards

which of the following detectors would be appropriate for measuring the Raman Spectra for samples irradiated with 455 nm light?

a)thermocouple

b)bolometer

c)photographic plate

d)charge-coupled device array (CCD)

D

31
New cards

which of the following detectors would be appropriate for measuring the IR spectra?

a) CCD array

b) photographic plate

c) photomultiplier tube

d) pyroelectric detector

d

32
New cards

which of the following sources would be appropriate excitation source for collecting IR spectra?

a) hollow cathode tube

b) nernst glower

c) Xenon lamp

d) Nd: YAG laser

C

33
New cards

define the term effective bandwidth

the wavelength range where the intensity is half the maximal value. This parameter will be reported for an instrument to inform the user regarding the resolution of data collected using the instrument

34
New cards

for grating, how many lines per millimeter would be required for the first-order diffraction line for 400 nm to be observed at a reflection angle of 7 degrees when the angle of incidences is 45?

a) formula?

b) answer

a) n(wavelength) = distance (sin(i) +sin (r))

1/d = lines/mm

b) 2072 lines/mm

35
New cards

consider an IR grating with 90 lines per mm and 15 mm of illuminated area. calculate the first-order resolution of this grating

a) formula

b) answer

a) wavelength/change wavelength = n* N

b)1350

36
New cards

what source, wavelength selector, container, and transducer should be used in these scenarios:

a) the investigation of the fine structure of absorption bands in the region 450 to 750 nm

b)obtaining absorption spectra in the far IR (20 to 50 micro meters)

c) a portable device for determining the iron content of natural water based on the absorption of the radiation by the read Fe(SCN) complex

d) the routine determination of nitrobenzene in air samples based on its absorption peak at 11.8 micrometers

e)determining the wavelengths of flame emission lines for metallic elements in the region from 200 to 780 nm

f) spectroscopic studies of the VUV region

g)spectroscopic studies in the near IR

a) W lamp, grating, glass window, PMT

b) globar, interferometer, KBr window, pyroelectric

c) W lamp, filter, T1Br or T1I window, Thermocouple

d) flame, grating, flame, PMT

e) Ar lamp, grating, KBr window, PMT

f) W lamp, grating, glass window, photoconductivity

37
New cards

why is specrtofluometry potentially more sensitive than spectrophotometry?

spectrophotometric measurements rely on a difference in signal whereas specrtofluometric measurements are absolute

38
New cards

explain why the wavelength of light fluoresced by a molecule is loner than the exciting light, but light fluoresced by an atom is the same wavelength as the exciting light

in AE, the atoms are typically in the gas phase and will not interact with one other. Therefore, there is no vibration or rotational relaxation observed when an atom fluoresces. However, molecules can vibrate and rotate, and will interact with other molecules, dissipating a little energy with each interaction. Light emitted will be lower energy than light absorbed

39
New cards

Formulas for calculating Stokes shift

convert wavelength to wave number

subtract individual peak from combined peak

40
New cards

absorbance from transmittance

A= -log(T)

41
New cards

What is the difference between H and D discharge lamps as a source for UV detection

H and D lamps differ only in the gases that are used in the discharge. D lamps generally produce higher intensity radiation

42
New cards

What is the difference between filters and monochromators as wavelength selectors

filters provide low resolution wavelength selection often suitable for quantitative work, but not for qualitative analysis or structural studies. Mononchromators produce high resolution (narrow bandwidths) for both qualitative and quantitative work. Photodiodes and PMTS.

43
New cards

What is the difference between photodiodes and PMTs

a photodiode consists of a photo-sensitive pn-junction diode that is normally reverse-biased. An incident beam of photons causes a photocurrent proportional to the photon flux. A PMT is a vacuum tube is a vacuum tue consisiting of a photoemissive cathode, a series of intermediate electrodes called dynodes and a collection anode. Each photoelectron emitted by the photocathode is accelerated in the E field to the first positively charged dynode where it can produce several secondary electrons. Results in a cascade multiplication of 10^6 or more electrons emitted per emitted photoelectron. Photomultipliers are more sensitive than photodiodes, but require a high voltage power supply compared to the low voltage supplies required by photodiodes. Photomultipliers are larger and require extensive shielding. Photodiodes are better suited for small, portable instruments because of their size and ruggedness.

44
New cards

What is the difference between single-beam and double-beam instruments for absorbance measurements.

ingle-beam and double-beam instruments for absorbance measurements.A single-beam spectrophotometer employs one beam of radiation that irradiates one cell. To obtain the absorbance, the reference cell is replaced with the sample cell containing the analyte. With a double-beam instrument, the reference cell and sample cell are irradiated simultaneously or nearly so. Double-beam instruments have the advantages that fluctuations in source intensity are cancelled as is drift in electronic components. The double-beam instrument is readily adapted for spectral scanning. Single-beam instruments have the advantages of simplicity and lower cost. Computerized versions are useful for spectral scanning

45
New cards

what is the difference between conventional and multi-channel spectrophotometers

Multichannel spectrophotometers detect the entire spectral range essentially simultaneously and can produce an entire spectrum in one second or less. They do not use mechanical means to obtain a spectrum. Conventional spectrophotometers use mechanical methods (rotation of a grating) to scan the spectrum. An entire spectrum requires several minutes to procure. Multichannel instruments have the advantage of speed and long-term reliability. Conventional spectrophotometers can be of higher resolution and have lower stray light characteristics

46
New cards

Provide 5 characterisics of absorption measurements that make it amenable to quantitative measurements in nearly every field

easy sample prep

selective

linear response to concentration

sensitive

non-destructive

47
New cards

what are shims in NMR spectrometer, and what are their purposes

In NMR,it is critical to have a uniform magnetic field (the more uniform the magnetic field, the higher probability that all proton spins will have the same Larmor frequency). Shims are metal coils carrying electric current. This current can be adjusted to modify the magnetic field

48
New cards

how are ions created in:

a) electron ionization

b)chemical ionization

a) a 70 eV electron strikes the compound creating a M+

b) the compound reacts with a proton donor to make M+

49
New cards

How are ions seperated in a:

a) single quadropole

b) triple quadrupole

a) Two pairs of parallel rods make the quadrupole. Ions are scanned by varying the direct current voltage (Vdc) and radio frequency voltage (Vrf).Similar tothe ion trap, quadrupoles "trap"ions of a specific mass (only ions with the appropriate mass will be able to travel along the rods). The main differenceis the shape of the electrodes (rods vs ring for quadrupole vs ion trap)

b)three-quadrupole systemthat results in higher ion selectivity.Within the 2ndquadrupole, analyte ions are fragmented via collisions with a neutral gas (collision induced dissociation

50
New cards

what is the detection range for:

a)ion trap

b) single quadrupole

c) triple quadrupole

d) TOF

a) less than 20,000 g/mol

b)1500 to 3000 g/mol

c) 1500 to 3000 g/mol

d)50000 g/mol

51
New cards

describe how increasing the rod length would affect the ion separation for a quadrupole

increases the resolution. The longer the rod, the more time to allow ions with a close but not exact target to get off track

52
New cards

if broadening is mainly due to ____, how should flow rate be changed to improve resolution?

a)longitudinal diffusion

b)finite equilibrium time

c)multiple flow paths

a) increase flow rate

b)decrease flow rate

c) no effect

53
New cards

what are the relative advantages and disadvantages of packed and open tubular columns in GC

packed columns have a high sample capacity. Open tubular columns have a better separation efficiency (smaller plate height), shorter analysis time and increased sensitivity

54
New cards

why do open tubular columns provide greater resolution than packed columns in GC

there is no band broadening due to the multiple path (A) term in the van Deemeter equation

55
New cards

When would you use ___ injection in GC?

a) split

b) splitless

c) on-column

a) is good for high concentration samples, scans where you want high resolution, or if you have a dirty sample

b) good for trace analysis

c) good for quantitative analysis or thermally sensitive samples

56
New cards

determining resolution from t and W

resolution = 2(Tr1-Tr2)/(WA +WB)

57
New cards

list the sources of error associated with pH measurements with a glass electrode

uncertainty of buffer solutions used, junction potential errors, junction potential drifts, alkaline or acid errors at extreme pH values, equilibration time, hydration of glass, temperature difference of calibration solution and analyte solutions, dirty electrode

58
New cards

explain the principle operation of an ion-selective electrode

analyte ions equilabrate between the outer solution and the membrane ligand, creating a slight charge imbalance. Changes in the analyte ion concentration can change the potential difference across the solution-membrane interface, causing a junction potential

59
New cards

Which of the following would decrease retention in ion chromatography:

a) increasing the column length

b) increasing flow rate of mobile phase

c) decrease the sample size

d) increase ionic strength of mobile phase

d

60
New cards

which of the following situations correctly describes the effect of increased flow rate of the mobile phase?

a) increased broadening due to the A constant in the van Deemeter equation

b)increased broadening due to the B constant in the van Deemeter equation

c) increased broadening due to the C constant in the van Deemeter equation

b

A constant = path traveled by analyte

B constant = diffusion of analyte via longitudinal and eddy diffusion

C constant = equilibration time of analyte

61
New cards

the use of a larger diameter stationary phase particle for HPLC would result in which of the following:

a) increased retention time

b) increased back pressure

c) increased plate height

d) increased resolution

c

62
New cards

for a size-exclusion chromatography:

a) a smaller stationary phase particle decreases resolution

b) a larger stationary phase particle decreases resolution

c) a smaller pore size of the stationary phase particles increases resolution

d) a larger pore size of the stationary phase particles increases resolution

c