1/30
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Connect the following events with the corresponding components of a scintillation detector
photomultiplier tube
the multiplying action of successive steps yields a very large number of electrons in response to each photoelectron
Name the following features that can be seen in the energy spectrum obtained from a sodium iodide crystal detector
photopeak
energy of photon source
true or false: a detector that can distinguish separate photopeak’s at 200 and 205keV has a better energy resolution than a detector that can only distinguish photopeak’s as close as 200 and 250keV
true
match each of the listed terms below to one of the following definitions
Z-pulse
electrical output from the preamplifier and amplifier connected to a photomultiplier tube in response to a photon interaction in the crystal
match each of the listed terms below to one of the following definitions
pulse height analyzer
used to select Z-pulses within the upper and lower limits of the energy window
match each of the listed terms below to one of the following definitions
energy window
range of acceptable photon energies
a single 15% energy window is set symmetrically on the 159keV photopeak for an iodine-123 imaging study in a gamma camera the lower the upper energy windows are
147 and 171keV
scintillation detectors convert
gamma ray energy into visible light
photomultiplier tubes
convert visible light into electrons that are then accelerated to provide amplification
True or False: a parallel-hole collimator providing both high sensitivity and high resolution is difficult to design principally because the requirements for sensitivity conflict with those for resolution
true
starting with a standard low-energy all-purpose collimator with will happen to the images if you alter the collimator deign by
lengthening the bores
higher resolution lower sensitivity
starting with a standard low-energy all-purpose collimator what will happen to the images if you alter the collimator deign by
thickening the septa
less septal penetration with higher energy phonons
starting with a standard low-energy all-purpose collimator what will happen to the images if you alter the collimator deign by
increasing the number of holes while decreasing the size of each hole
higher resolution lower sensitivity
true or false: in a converging-hole collimator the holes angle inward toward the organ to be imaged in consequence the organ appears larger at the face of the crystal
ture
how many pixels of data can be stored in 64X64 image matrix
4096
the usual gamma camera head contains all of the following elements except
crystal
photomultiplier tubes
collimator
positioning circuit
pulse-heigh analyzer
focusing assembly
focusing assembly
match the part of the imaging system with its function
positioning algorithm
positioning algorithm - improves image resolution by factoring in the output of each PMT to better localize the position of the photon interaction in the imaging field
match the part of the imaging system with its function
pulse height analyzer
accepts summed PMT output (Z-pulses) that correspond to the photon energy of the source
match the part of the imaging system with its function
gantry
supports the camera heads
match the part of the imaging system with its function
collimator
restricts photon entry to the crystal so that each point in the image corresponds to a unique point in the source
match the part of the imaging system with its function
crystal
converts photon energy to light energy
match the part of the imaging system with its function
Photomultiplier tube
converts light energy to a small electrical pulse
What is the pixel size for an image acquired with a field of view of size 32cm × 32cm and a matriz of size 128 × 128
2.5 × 2.5mm (320mm/128=2.5mm)
Connect the following events with the corresponding components of a scintillation detector
preamplifier
the output of the photomultiplier tube is amplified to more readily detectable level of current
Connect the following events with the corresponding components of a scintillation detector
crystal
light photons are released in response to gamma photons
Connect the following events with the corresponding components of a scintillation detector
photocathode surface of photomultiplier tube
photoelectrons are released in response to light photons
Name the following features that can be seen in the energy spectrum obtained from a sodium iodide crystal detector
Compton peak or edge
maximum Compton electron energy
Name the following features that can be seen in the energy spectrum obtained from a sodium iodide crystal detector
Iodine escape peak
can be seen at 28keV below the photpeak
Name the following features that can be seen in the energy spectrum obtained from a sodium iodide crystal detector
annihilation peaks
can be seen at 511keV and1020keV below the photopeak of the high energy photon
Name the following features that can be seen in the energy spectrum obtained from a sodium iodide crystal detector
Coincidence peak
peak near 420keV emitted by a In-111 source
Name the following features that can be seen in the energy spectrum obtained from a sodium iodide crystal detector
lead x-ray peak
72keV