the amount of heat given off is expressed in...
heat units (HU)
number of HU produced depends on the...
type of generator and the exposure factors used
alternating current
alternating betwee positive and negative flow
xray generator wave forms
single phase recified current
3 phase 6 pulse wave form
3 phase 12 pulse current
high frequency wave form
(ordered in low frequency to high)
each type of generator produces...
different voltage wave form
ripple
fluctuation in voltage output
the higher the frequency the higher the ripple
tube heat load capacity
-every tube has a specific heat load capacity
-amount of heat tube can withstand w/o damage tube components
heat is produced according to:
technical factors selected
-kVp, mA, time
-focal spot size selected
small focal spot typically not used above
200 mA
generator type
-more powerful generator produces xrays more efficent
-also creates more heat units (typically 1.4 times more)
generator factor indicates ______
heat production
the more _______ the generator, the _____ the heat units produced within the tube during xray production
efficient
higher
equation for heat unit
HU= kVp x mA x time x generator factor
single phase unit
1
three phase 6 pulse
1.35
three phase 12 pulse
1.41
high frequency
1.45
3 types of rating charts
tube rating chart
anode cooling chart
housing cooling charts
tube rating chart
-determine technical factors that can be used without overloading tube
-any combinations of facotrrs at or under the curve is safe
tube chart
y axis
x axis
curved lines
kVp
time (sec)
mA
find intersection of kVp and time on graph
technical factors
miliamperage, kilovoltage, time
(most important things we set on control panel)
housing cooling charts (dont really focus on)
similar to anode cooling chart
anode will reach heat limits before housing
kVp on graphs is usually between
30-150 kVp
cooling anode chart
calculates the time it takes for the anode to cool from a given exposure or time it will take to cool off enough to make another exposure
anode cooling chart
y axis
x axis
curved line
total heat capacity
time (mins)
starting point HU to end time of complete cooling
equation/ calculating cooling chart
kVp x mA x time x general factor