1/75
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
|---|
No study sessions yet.
FILTRATION
The most important patient protection characteristics of a radiographic imaging system
white x-ray
low energy x-rays is aka
reduce patient dose
primary function of filtration
filter out low energy x-rays
secondary function of filtration
inherent filtration
added filtration
total filtration
3 TYPES OF FILTRATION
inherent filtration
filtration that is a result of the composition of the tube and housing
added filtration
The filtration that is added to the port of the x-ray tube.
total filtration
the sum of inherent and added filtration
0.5
inherent filtration constant
half-value layer (HVL)
Filtration is Measured By:
COLLIMATION
X-ray field must coincide with the light field of the variable-aperture light-localizing collimator
tape measure
Use to measure collimation
marked on the x-ray tube housing
Location of Focal Spot:
Collimation
The restriction of the size and shape of the x-ray beam in order to reduce patient exposure.
Focal Spot Size (FSS)
It principally determines the spatial resolution of a radiographic imaging system
pinhole camera
star pattern
slit camera
FOCAL-SPOT SIZE is measured by:
Pinhole Camera
It is difficult to use
It requires excessive exposure time
star pattern
It is easy to use
slit camera
The standard for measurement of focal-spot size
3 mm
Star Pattern Limitation for Focal-Spot Sizes less than _____
kVp Calibration
Type of calibration that must be done whenever high-voltage generator components have changed significantly.
kVp Calibration
The x-ray generator must be properly calibrated
ion chamber
photodiode
Device used in kVp Calibration
photodiode
Converts light into electron
cutie pie
ion chamber is aka
4%
Variation in kVp necessary to affect image OD & radiographic contrast
Exposure Time
responsible for patient dose & OD
ion chamber
photodiode
Device used in Exposure timer accuracy
EXPOSURE LINEARITY
The ability of a radiographic unit to produce a constant radiation output for various combination of mA & exposure time
dosimeter
EXPOSURE LINEARITY is determined by:
EXPOSURE REPRODUCIBILITY
The ability of a radiographic unit to produce a precisely the same radiation exposure when all the techniques are changed & returned to previous value
1. Select a combination of technique factors & hold them constant for a series of 10 exposure
2. Make a series of at least three exposures at the same technique
factors, having changed technique controls between exposure
Two Methods of Evaluation for EXPOSURE REPRODUCIBILITY:
Film Illuminators
It should be analyzed photometrically
photometer
A device used to measure light intensity at areas of the illuminator
Screen-Film Contact Test
Cassette holders must be examined every month for worn closures, light leaks, and warping.
Screens should be cleaned monthly with commercially available cleaners
Radiographing a wiremesh pattern
Analyzing the image for areas of blur
Method of Evaluation for Screen-Film Contact
Replaced Felt or Foam Pressure Pad
if blur appears
Replaced Cassette
if replacement of foam or pressure pad does not correct the problem
Processor Cleaning
RT removes the transport and crossover racks & cleans them & the processing tanks with appropriate fluids
feed/entrance rack
developer rack
cross-over rack
fixer rack
water rack
dryer rack
RACKS IN AUTOMATIC PROCESSING:
cross-over rack
prevents contamination of developer from fixer
power of hydrogen
pH
90 seconds
Dry-To-Dry Time:
500 films/hr
Processing Rate:
Hydrometer
measures specific gravity
hygrometer
measures humidity
specific gravity
measures the activity of developer
35C / 95F
High development temperature
31C / 87F
Wash water temperature
SCHEDULED MAINTENANCE
Routine procedures that are performed usually weekly or monthly
PREVENTIVE MAINTENANCE
A planned program of parts replacement at regular intervals
It requires that a part be replace before it fails
Non-Scheduled Maintenance
The worst kind A failure in the system that necessitates processor repair
proper program of scheduled and preventive maintenance
Non-Scheduled Maintenance Minimize By:
1.000
specific gravity of water
1.030
specific gravity of developer
10-10.5 pH level
Normal pH level of solution
contaminated
if the pH level of solution decreased 0.4 the solution is _____
oxidized
if the pH level of solution decreased 0.8 the solution is _____
oxidation
loss of electron
replace solution
once ph level is too low or too high:
over diluted
if the specific gravity is low =
too concentrated
if the specific gravity is high =
reject film analysis
To provide the knowledge necessary to set up a system which will give a detailed analysis of the rejected films and the reason for the rejections, in order to put in place remedial action.
monthly
reject film analysis frequency
save money
save time
reduce radiation dose
identify errors
provide data for comparison
provide statistics
BENEFITS of
reject film analysis
10%
A reject rate of ___ or more should be considered unacceptable
5% to 10%
A reject rate of ___ justifies continued monitoring.
reject film analysis
To provide sufficient knowledge of exposure selection and manipulation to enable accurate selection of exposure in order to produce maximum information on the resultant radiographs
number of rejected films/number of examination x 100%
reject rate formula
number of repeated films/number of examination x 100%
repeat rate formula
artifacts
cut off
mirrored
over/underexposure
hyporetention
positioning error
cause of reject or repeat films
kVp
Controls largely the penetrating power and to a lesser degree, the intensity of radiation and therefore film density and patient dose. It affects the contrast
mas
Controls the intensity of radiation and therefore the film density and patient dose.
time
Controls the length of time of radiation flow, and therefore film density and patient dose. The longer the exposure the higher the film density and patient dose
SID
The greater the distance of xray beam travels the less effective it will be, the less distance it travels the more effective it will be
Film-focal distance
other name for SID