Section 2 Questions

0.0(0)
studied byStudied by 0 people
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/125

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

126 Terms

1
New cards

You are refilling a nondisposable nebulizer supplying O2 to a postoperative patient via aerosol face mask. The capillary tube dislodges and falls on the floor. The most appropriate action would be to

switch the patient to a humidifier

wipe the capillary tube with alcohol and reconnect

reinsert the capillary tube

replace the entire nebulizer with a clean unit

replace the entire nebulizer with a clean unit

2
New cards

Which of the following nondisposable items must be sterilized between use on different patients?

bronchoscope forceps

endoscope

ventilator tubing

face mask

bronchoscope forceps

Any reusable item posing a critical infection risk must be sterilized between use on different patients. Critical items are those introduced into the bloodstream or other parts of the body. These include surgical devices, intravascular catheters, implants, heart-lung bypass components, dialysis components and bronchoscope forceps/brushes. Devices that contact intact mucous membranes such as endoscopes or ventilator circuits are categorized as semi-critical and should be processed using a high-level disinfection method. Items that touch only intact skin or do not contact the patient (such as a face mask) present a a non-critical infection risk and require only detergent washing or low to intermediate level disinfection.

3
New cards

A respiratory therapist is caring for a patient who has a sign posted outside their hospital room door that indicates, "Respiratory Isolation". The therapist should observe all of the following EXCEPT:

Patient should use a mask when visitors are present

Use N95 respirator

Universal precautions

Do not allow visitors to enter the room

Do not allow visitors to enter the room

4
New cards

Which of the following types of equipment processing would you recommend for a reusable external facemask?

soap and water wash

high-level disinfection

low-level disinfection

sterilization

low-level disinfection

Since a face mask only contacts a patient's intact skin, it is considered a noncritical item. Low-level disinfection is the appropriate type of equipment processing for noncritical items.

5
New cards

A patient with a tracheostomy has 40% O2 and aerosol delivered by a tracheostomy mask. If the patient develops an Escherichia coli infection of the stoma and airways, the most likely source is

The patient's own upper airway bacteria

Nebulizer water used past its expiration date.

Inadequate hand washing by the health care workers.

Improper manufacturer sterilization of the mask.

Inadequate hand washing by the health care workers.

This organism is found in the lower intestine. It is often spread from improper hand washing after use of the toilet.

6
New cards

Which type of disinfection inactivates all microorganisms except bacterial spores?

intermediate-level

high-level

surface active

low-level

high-level

By definition, only high-level disinfection inactivates all microorganisms except bacterial spores. Some high-level disinfectants may also destroy spores if the exposure time is sufficient.

7
New cards

Which of the following types of equipment processing would you recommend for reusable PFT mouthpieces/tubing?

high-level disinfection

soap and water wash

sterilization

low-level disinfection

high-level disinfection

Since PFT mouthpieces/tubing may contact a patient's mucous membranes, they are considered semicritical items. High-level disinfection is the appropriate type of equipment processing for semicritical items. Alternatively, one can use disposable HEPA filtration and simply change the mouthpiece and filter between patients.

8
New cards

Which of the following is an acceptable way to dispose of liquid biohazardous/infectious waste, such as blood or body fluids?

pour into sanitary sewer (drain or toilet)

process in hot water cycle (pasteurization)

transport in containers to sanitary landfill

place in a biohazard bag for disposal

pour into sanitary sewer (drain or toilet)

Liquid biohazardous/infectious waste can be disposed of either 1) via the sanitary sewer system (pouring down the drain or toilet) or 2) via steam autoclaving (liquid cycle). If disposing into the sewer, state or local regulations may require pretreatment with a disinfectant like sodium hypochlorite (typically 1 part bleach to 9 parts liquid waste). If autoclaving, be sure to open the collecting vessel before initiating the cycle. After autoclaving the treated liquid can be poured down the drain. Appropriate PPE (e.g., gloves, gown, mask, and protective eyewear) should be worn when handling liquid waste, which must be transported in sealed leak-free containers, not bags. Only treated solid wastes are placed in sanitary landfills.

9
New cards

Which of the following are acceptable for home-care use for the cleaning of non-electric respiratory therapy equipment?

Acetic acid

Sonacide

Cidex

Bleach

Acetic acid

Acetic acid is another name for common household vinegar. Vinegar is a safe and effective cleaning agent for home use. Cidex is not a product that would be appropriate for use in the home. It is not safe to use bleach on respiratory therapy equipment in the home. The use of sonacide is not intended for home use.

10
New cards

You should "double bag" solid or semi-solid infectious waste when:

the waste originates from a reverse isolation area

the waste comes from an immunocompromised patient

the waste originates from a contact isolation area

the waste could easily penetrate the primary bag

the waste could easily penetrate the primary bag

11
New cards

The organisms most commonly associated with contamination of nebulizers are:

fungi

adenoviruses

gram+ bacilli

gram- bacilli

gram- bacilli

The organisms most commonly associated with contamination of aerosol generators are gram- bacilli, particularly Pseudomonas aeruginosa. More recently, disturbing case reports of nosocomial infection by Legionella pneumophila (the cause of the highly virulent Legionnaires' Disease) have been disseminated.

12
New cards

After a 30-minute soak in alkaline gluteraldehyde (Cidex), viable microorganisms are found on the equipment. Which of the following is a possible reason?

high resistance of organism or insufficient soak time

inappropriate sterilization solution type

insufficient aerate

device requires autoclaving

high resistance of organism or insufficient soak time

Alkaline glutaraldehyde, or Cidex, is a liquid that can disinfect and sterilize equipment. There are several factors that dictate its effectiveness. They are: strength of solution, soak time, and resilience or resistance of the organism. Aeration time does not decrease the effectiveness of the solution.

13
New cards

A chemical indicator tape found inside the sealed, packaged disposable mask helps determine which of the following?

if the package has remained closed

if the equipment has been properly sterilized

if the equipment is sterile

if microorganisms remain viable on the equipment

if the equipment has been properly sterilized

A chemical indicator tape inside a sterialized package does not actually indicate the presence or absence of viable spores. It does, however, indicate if the equipment has been through the sterilization process.

14
New cards

A respiratory therapist is working in a 69-bed rural hospital with limited resources. In order to disinfect a re-usable plastic mouthpiece used for pulmonary function testing, the respiratory therapist should

soak in Sonacide for 30 minutes.

send for steam autoclave.

soak in Cidex for 10 hours.

soak in 60% alcohol for 10 hours.

soak in Cidex for 10 hours.

Cidex (alkaline glutaraldehyde), will kill bacteria in 10 min, and everything in 10 hours. It works by using a caustic pH between 7.5 and 8.5. It should be changed every 14 days.

15
New cards

A respiratory therapist is disinfecting reusable mouthpieces in Sonacide. For how long should the equipment soak to achieve disinfection?

6 minutes

10 minutes

2 minutes

8 minutes

10 minutes

Acid glutaraldehyde, like its counterpart glutaraldehyde, has the ability to disinfect and sterilize. However, soak times are different. The acid version can disinfect in 10 min. but can sterilize in only one hour. In this question, we are asked to merely disinfect the reusable mouthpieces.

16
New cards

Which of the following types of equipment processing would you recommend for a reusable device that was introduced into a patient's bloodstream?

sterilization

low-level disinfection

detergent washing

high-level disinfection

sterilization

17
New cards

Your hospital requires all liquid infectious waste be transported and processed centrally before disposal. To transport several full suction collection bottles from ICU for processing you would:

place sealed bottles in a puncture-resistant biohazard bag

seal and label each bottle and transport them one at a time

place sealed bottles into a closable leak-proof biohazard container

transfer all bottle contents into a large secondary carted vessel

place sealed bottles into a closable leak-proof biohazard container

18
New cards

A respiratory therapist is attempting to sterilize a non-disposable IPPB circuit from a patient who has confirmed active tuberculosis. The sterilizing material is acid gluteralderhyde (Sonacide). For how long must the therapist soak the circuit to achieve sterilization?

unable to achieve sterilization with this chemical

10 hours

10 minutes

20 minutes

10 hours

19
New cards

Which of the following types of equipment processing would you recommend for a reusable device that only contacted a patient's intact skin?

high-level disinfection

sterilization

soap and water wash

low-level disinfection

low-level disinfection

20
New cards

Which of the following devices is LEAST LIKELY to transmit pathogenic organisms to the patient?

ultrasonic nebulizer

small volume drug nebulizer

wick-type humidifier

large volume jet nebulizer

wick-type humidifier

21
New cards

A respiratory therapist is working in a 69-bed rural hospital with limited resources. In order to disinfect a re-usable plastic mouthpiece used for pulmonary function testing, the respiratory therapist should

soak in Sonacide for 30 minutes.

send for steam autoclave.

soak in 60% alcohol for 10 hours.

soak in Cidex for 10 hours.

soak in Cidex for 10 hours.

22
New cards

During a procedure in ICU, blood spills onto the outer plastic casing of an in-use ventilator. After cleaning the surface using gloves and disposable gauze pads, how would you decontaminate it?

wash the surface with a detergent and rinse with sterile water

swab the area with a tuberculocidal disinfectant or bleach solution

spray the area with a concentrated isopropyl alcohol solution

take the ventilator out of service and have it sterilized

swab the area with a tuberculocidal disinfectant or bleach solution

23
New cards

A patient has been receiving an aerosolized bronchodilator through a handheld nebulizer over the past day. Which of the following would a respiratory therapist recommend to minimize the risk of her getting a nosocomial infection?

  1. Add a broad-spectrum antibiotic to the nebulizer with each treatment.

  2. Change the handheld nebulizer and mouthpiece every 24 hours.

  3. Discard medications and saline solution 24 hours after opening.

  4. Respiratory therapists should wash their hands before giving a treatment

1 and 2 only

1, 2, 3 and 4

2, 3 and 4 only

3 and 4 only

2, 3, and 4 only

24
New cards

Which of the following are acceptable for home-care use for the cleaning of non-electric respiratory therapy equipment?

Sonacide

Acetic acid

Bleach

Cidex

Acetic acid

Acetic acid is another name for common household vinegar. Vinegar is a safe and effective cleaning agent for home use. Cidex is not a product that would be appropriate for use in the home. It is not safe to use bleach on respiratory therapy equipment in the home. The use of sonacide is not intended for home use.

25
New cards

You need to transport several contaminated nondisposable ventilator circuits from a patient unit to an equipment processing area. You would:

transport the circuits in leak- and puncture-resistant biohazard bags

place circuits in biohazard bags, then transport in a closed, rigid container

wash and disinfect the circuits at point of origin before transporting

bag the circuits, then transport on a cleaned and sanitized cart

place circuits in biohazard bags, then transport in a closed, rigid container

26
New cards

At its point of generation, solid infectious waste should be:

placed in standard garbage bags and sent out for incineration

placed in the unit's utility sink and soaked in disinfectant

placed in a color-coded, puncture-resistant bag or container

broken into small pieces and flush into the sanitary sewer

placed in a color-coded, puncture-resistant bag or container

27
New cards

A bronhoscope has been used on a patient with a staphylococcus infection. To properly sterilize the equipment, perform a

3 minute soak in one-day-old Sonacide

steam autoclave

15 minute soak in 13-day-old Cidex

1 hour soak in 30-day-old Sonacide

15 minute soak in 13-day-old Cidex

Cidex, or alkaline glutaraldehyde, is most commonly used for bronchoscopy sterilization. A 10-hour soak is best, but all bacteria will be destroyed within 10 min. In this example, the primary concern is Staphylococcus, a bacteria. Alkaline glutaraldehyde must be changed no later than every 14 days.

28
New cards

An intubated patient has a size 7.5 mm endotracheal tube in place. Diffuse rhonchi are auscultated. A respiratory therapist is using a 12 Fr suction catheter set to a pressure of -100 cm H2O, attempting to suction for 15 seconds. Only scant secretions are suctioned, and breath sounds do not improve after the procedure. The respiratory therapist should increase the

catheter size to a 14 Fr.

catheter size to an 18 Fr.

suction pressure

suction duration.

suction pressure

29
New cards

While performing a routine ventilator a check on a patient with a balloon-tipped flow-directed pulmonary artery catheter in place, the respiratory therapist notices the inflection points on the waveform indicate the tip of the pulmonary artery catheter is improperly placed in the right ventricle. To correct this problem, the therapist should recommend

removing the catheter and inserting a new one

recording the pulmonary artery pressure as shown on the waveform

inflating the balloon and advancing the catheter

twisting the catheter until pulmonary artery pressures are observed

inflating the balloon and advancing the catheter

The pulmonary catheter should be terminated in the pulmonary artery, which is beyond the right ventricle. Therefore, advancing the catheter is appropriate. The catheter may be advanced by sailing it into position, which is done by inflating the catheter balloon and allowing blood flow to carry the catheter into position and then deflating the balloon once it is in the proper position.

30
New cards

A post-operative patient is receiving positive-pressure ventilation with an IPPB while recovering from anesthesia. The mandatory rate is 10/min. Inspiratory pressure is set to 18 cm H2O. Gradually the rate increases to 16/min while the monometer needle is showing a significant negative deflection before inhalation begins. The pressure monometer no longer rises smoothly during inhalation. What should the respiratory therapist do?

sedate the patient

increase sensitivity

decrease inspiratory flow rate

wean the patient

increase sensitivity

The gradual increase in respiratory rate indicates that the patient is waking up from anesthesia, as expected. The negative deflection before inhalation indicates that the machine sensitivity is too low-requiring the patient to work too hard to trigger inspiration. The manometer not rising smoothly indicates the flow rate is also too low. It is not appropriate to sedate the patient but rather allow him to wake up with more applicable settings.

31
New cards

Which of the following alarms would be most important for a patient receiving positive pressure ventilation by a pressure-cycled ventilator?

high pressure

low FIO2

I:E ratio

low volume

low volume

32
New cards

Adding an aerosol-holding chamber to a metered-dose inhaler

requires more complex patient education regarding technique.

requires greater patient coordination.

minimizes deposition of particles that are too large.

disallows the patient to receive medication by mask

minimizes deposition of particles that are too large.

An aerosol holding chamber, commonly known as a spacer, not only improves medication delivery when added to a metered-dose inhaler, but also reduces the significant amount of coordination required to properly self-administer an MDI and filters out particles that are too large for pulmonary effectiveness.

33
New cards

In working with an electrically powered apparatus, you notice a slight tingling sensation when the metal parts of the equipment are touched. In this case, you should:

use this device only with non-electrically sensitive patients

immediately take the piece of equipment out of service

use an extension cord to increase electrical resistance

use a 'cheater' adapter to bypass the ground connection

immediately take the piece of equipment out of service

A tingling sensation which occurs when the metal parts of a piece of equipment are touched usually indicates improper grounding and the possibility of a serious leakage current. In these cases, you must ensure that the faulty equipment is immediately taken out of service.

34
New cards

A large drop in the humidifier water level in an infant ventilator circuit will result in:

a decrease in the volume lost due to compression

a decrease in the volume received by the infant

an increase in the circuit's mechanical deadspace

an increase in the volume received by the infant

a decrease in the volume received by the infant

Due to the very small tidal volumes delivered to infants, increases in a ventilator circuit's internal volume, as will occur when the humidifier water level drops, cause large increases in compressed volume loss. The larger the compressed volume loss, the lower the actual volume received by the infant.

35
New cards

A patient with CHF requires 5 L/min humidified oxygen by nasal cannula continuously. The patient is being discharged home. Which of the following modalities is not indicated?

molecular sieve device

pulse-dose conservation device

simple mask

portable oxygen tanks

simple mask

36
New cards

A doctors orders continuous heated aerosol for a newborn in an oxyhood. You would recommend a servo-controlled heated membrane humidifier instead for which of the following reasons?

  1. Minimal risk of fluid imbalance

  2. Low and safe noise level

  3. Low likelihood of infection

1 and 3

2 and 3

1 and 2

1, 2 and 3

1, 2, and 3

Newborn infants are at high risk for infection. For this reason, medical gas humidification should always be achieved with an evaporative humidifier that produces only molecular water, not water aerosol (which can spread bacteria and cause overhydration). In order to assure maintenance of a neutral thermal environment, the humidifier selected should allow precise control over gas temperature at the point of delivery. Last, neonatal exposure to high noise levels has been associated with subsequent auditory damage. Thus a 'quiet' device should be used. A servo-controlled heated humidifier meets all three criteria.

37
New cards

An air-entrainment jet nebulizer is set at 50%. The respiratory therapist analyzes the oxygen coming from the end of the tubing and finds it is 60%. Which of the following most likely explains this finding?

poorly calibrated analyzer

leak in the tubing

excess water in the tubing

low flow through the jet orifice

excess water in the tubing

When administering oxygen by any device that has a venturi mechanism, back pressure on the venturi will slow the speed of gas, decrease room air entrainment, and result in an increase in FIO2. Of the options offered, only excess water in the tubing would cause this type of back pressure.

38
New cards

A patient receiving volume-controlled ventilation has a balloon-tipped pulmonary artery catheter in place. The respiratory therapist notices the PA waveform is ascending and descending with inflection points at 25 and 2 mmHg. Based on this information, the therapist should recommend

advancing the catheter

deflating the catheter balloon

rotating the catheter

monitoring the patient closely

advancing the catheter

The pulmonary artery catheter waveform that has a high inflection point of 25 and a low inflection point of 0-2 mmHg is an indication that the tip of the catheter is in the right ventricle of the heart. The proper placement of this catheter is in the pulmonary artery. Therefore, advancing the catheter is indicated. This is done by inflating the balloon and allowing the catheter to sail into a proper position in the pulmonary artery.

39
New cards

A respiratory therapist is initiating mechanical ventilation at the following settings and preferences:

Mode Assist/control

Mandatory rate 20

VT 500 mL

FIO2 0.5

PEEP 5 cm H2O

I:E 1:2

Which of following is the minimum inspiratory flow setting?

80 L/min

40 L/min

65 L/min

28 L/min

40 L/min

To calculate the minimum inspiratory flow rate for a given set of ventilator parameters the I:E ratio can be added together and then multiplied by the minute ventilation. In this case 1+2=3. 3 x 10.0L = 30L/min.

40
New cards

A febrile patient has the following blood gas results: (results are not corrected for temperature)

pH 7.35

PaCO2 44 mmHg

PaO2 28 mmHg

HCO3- 24 mEq/L

BE 0 mEq/L

O2 Sat 90%

Which of the following requires attention?

Clark electrode

Severinghaus electrode

Sanz electrode

spectrophotometer

Clark electrode

In this arterial blood gas a PaO2 of 28 mmHg is reported. That is too low for the patient to still be alive and therefore this data is probably faulty. The question is essentially asking which electrode would you suspect is not working properly. The Clark electrode is another name for the PO2 electrode. The Sanz electrode is another name for the pH electrode and the Severinghaus electrode is called the CO2 electrode.

41
New cards

A patient with diabetes is demonstrating a Kussmaul breathing pattern. The patient is receiving supplemental oxygen at FIO2 1.0 by a single large volume nebulizer device. The inspiratory flow produced by the patient is measured and found to 20 L/min. Pulse oximetry reveals a saturation of 89%. Which of the following would be most helpful in this situation?

Install a tandem large volume nebulizer in the current system

Increase the oxygen flow rate at the flowmeter to 22 L/min

Replace the large volume nebulizer with an ultrasonic nebulizer device

Administer a mild sedative

Install a tandem large volume nebulizer in the current system

When a large-volume nebulizer device is set to 100% oxygen, the maximum achievable flow rate is about 15 L/min. Because this patient has a minute ventilation of 20 L/min, his or her inspiratory demand is not being met. This can be overcome by installing a second, or tandem, large-volume nebulizer. Together, the nebulizers may produce up to 30 L/min, which will exceed the patient's inspiratory demand.

42
New cards

A pressure-volume ventilator graphic fails to return to the point of origin. Which of the following could be the cause?

excessive cuff pressure

leak in the ventilator circuit

pulmonary distension

inadequate PEEP

leak in the ventilator circuit

A pressure volume loop on a ventilator should end in the same place it starts. If a pressure volume loop fails to return to zero, or its point of origin, there is likely a leak in the circuit, reducing return volume.

43
New cards

Excessive bubbling is noted in water seal chamber of a chest tube drainage system. Which of the following could be the cause?

fluid level in the suction control chamber is too low

leak in the tubing between the patient and fluid collection chamber

water seal fluid level is too high

water seal fluid level is too low

leak in the tubing between the patient and fluid collection chamber

Excessive bubbling in the water-seal chamber of a chest tubes drainage system is caused from a leak somewhere between the water-seal compartment and the patient's lung. In this case a leak in the tubing between the patient and fluid collection chamber would cause the excessive bubbling.

44
New cards

Your patient is ordered to start breathing treatments with pentamidine isethionate (NebuPent). To minimize your risk of drug exposure, you should do which of the following?

place an HME between the nebulizer and the mouthpiece

use a metered dose inhaler+valved holding chamber

use a SVN with nonrebreathing circuit+expiratory filter

use a small-particle aerosol generator with a HEPA filter

use a SVN with nonrebreathing circuit+expiratory filter

When administering aerosolized drugs whose toxicity or side effects pose a threat to you or other caregivers or patients, you should make efforts to contain the aerosol by using a nonrebreathing, filtered nebulizer system (such as the RespiGuard) and applying environmental protection strategies, to include use of negative pressure chambers/booths, scavenging system and caregiver wearing of HEPA filtered masks. methods.

45
New cards

When making routine equipment checks you hear the relief valve of a patient's bubble humidifier sounding. Which of the following is the most likely cause of this problem?

excessive oxygen input flow

clogged bubble/diffuser

high wall outlet pressure

decreased patient compliance

excessive oxygen input flow

The relief valve of a humidifier sounds when the pressure in the reservoir container exceeds the valve's threshold pressure (usually between 1-2 psig). The most common reasons for this to occur are 1) excessive oxygen input flow, and 2) downstream obstruction to outflow. In bubble humidifiers that use a diffuser plate to break up the gas stream, clogging of this element would prevent gas from entering and pressurizing the system. Flowmeter restriction prevents high wall outlet pressures (> 50 psig) from affecting pressure in the humidifier. And changes in patient mechanics (e.g., compliance or resistance) would have no effect on pressure in the humidifier.

46
New cards

A large volume jet nebulizer is operating at the 100% O2 setting with an input flow of 12 L/min. What effect will changing the entrainment port setting from 100 to 70% have on the system?

the input liter flow will increase by 30%

the total output liter flow will decrease

the aerosol weight density will increase

the total output liter flow will increase

the total output liter flow will increase

When the air-entrainment port of a large volume jet nebulizer is opened, room air is entrained, and total flow through the system increases (in direct proportion to the size of the opening). This increased output flow increases total aerosol output per minute, but actually lowers aerosol density.

47
New cards

Which of the following are potential complications associated with the use of heated humidification systems with mechanical ventilators?

pulmonary dehydration

thermal injury to the airway

increase of heat and water vapor content of inspired gas

atelectasis

thermal injury to the airway

Overheating of the airway may occur with the use of an electrically powered heater if not properly maintained. Because the device is electrically powered, it also may be associated with electric shock of the patient. These complications are not common, but possible.

48
New cards

An apneic post-op patient is receiving pressure-cycled ventilation with a set expiratory time in the recovery room. Which of the following would occur if the patient's artificial airway were to become partially obstructed?

  1. Decreased delivered volume

  2. Increased respiratory rate

  3. Decreased inspiratory time

1, 2 and 3

2 and 3

1 and 3

1 and 2

1, 2, and 3

When a true pressure-cycled ventilator encounters high resistance (e.g., partial airway obstruction), it will cycle off prematurely. As a result, inspiratory time and delivered volume will decrease. In addition, with a shorter inspiratory time and fixed expiratory time, the respiratory rate will increase.

49
New cards

A respiratory therapist is monitoring a continuous flow CPAP device set at 8 cm H2O. With each inspiration from the patient the monitored pressure drops from 8 cm H2O to 3 cm H2O but then returns to 10 cm H2O during expiration. Which of the following is the best explanation for this observation?

leak in the mask

gas flow to the patient is insufficient

normal operation

the patient is in need of Bi-level therapy

gas flow to the patient is insufficient

50
New cards

What would occur on a time-cycled ventilator with a fixed rate if the inspiratory flow rate were reduced?

decrease in inspiratory time

increase in tidal volume

decrease in tidal volume

increase in inspiratory time

decrease in tidal volume

51
New cards

A patient receiving volume-controlled ventilation is coughing uncontrollably. The high-pressure alarm is sounding continuously. The respiratory therapist should first

silence the ventilator alarm, calm the patient's anxiety

sedate the patient

provide the patient with manual ventilatory support

suction the patient

provide the patient with manual ventilatory support

The first response to any ventilator alarm is to disconnect the ventilator from the patient and begin manual ventilation. Once ventilation and oxygenation of the patient are assured, troubleshooting for the source of the problem may begin.

52
New cards

Blood pressure obtained by a sphygmomanometer reads higher than the indicated blood pressure from an arterial line on the same patient. To correct problem, the respiratory therapist should FIRST

flush the art line with sodium heparin

discontinue use of the sphygmomanometer

check for air bubbles in the transducer dome

advance the arterial line catheter

check for air bubbles in the transducer dome

When blood pressure by sphygmomanometer (cuff) differs from the blood pressure obtained from an indwelling arterial line, the blood pressure taken by cuff is considered more accurate and reliable. Problems that may occur in the ART line include clots in the line and bubbles in the transducer dome.

53
New cards

A therapist is unable to obtain a reading with a galvanic-type oxygen analyzer when attempting to measure oxygen percentage inside an infant oxygen hood while heated humidity is also applied. The therapist should?

change the electrode

discontinue heated humidity

change the electrolyte solution

change the batteries

change the electrode

When a galvanic-type oxygen analyzer fails to produce a reading, it is likely a problem with the power source. These type of analyzers do not have batteries. The electrode is in essence a battery. So, the solution is to change the electrode. The electrode is also called the fuel cell.

54
New cards

Which of the following are associated with a fenestrated tracheostomy tube?

silver

cap

button

foam cuff

cap

A fenestrated tracheostomy tube has an inner cannula with a hole to facilitate talking and weaning. It also has an inner cannula for resuscitation, or mechanical ventilation. The cuff is like that of an endotracheal tube, low-pressure, high-volume. And finally, a cap is used to close the tube for speech therapy. When the cap is used, the inner cannula should be removed and the cuff should be deflated.

55
New cards

When selecting a suction catheter to be used in an oral endotracheal tube, the respiratory therapist should select a catheter whose diameter should not exceed what fraction of the internal diameter of the endotracheal tube?

1/4

1/2

2/3

3/4

1/2

56
New cards

A 68-kg (150-lb) post-operative female patient is receiving mechanical ventilation in the SIMV mode with a set mechanical rate of 6/min and an actual rate of 10/min. Her set tidal volume is 500 mL with a spontaneous tidal volume of 400 mL. She is comfortable with stable vital signs. What should the respiratory therapist do?

Increase the set tidal volume to 650 mL.

Increase the mechanical rate to 16/min.

Set the low VE alarm to 4.0 L/Min

Start pressure support at 15 cm H2O

Set the low VE alarm to 4.0 L/Min

In this case the patient is comfortable with stable vital signs during weaning. There is no indication that any increase in ventilation is necessary. A respiratory therapist may seek a response that is more aggressive in weaning, but no such choice is offered. Therefore, the best response is to set the alarms and monitor the patient closely to assure that she will continue her spontaneous breaths.

57
New cards

Which of the following oral airways would you select for a newborn infant?

Yankauer

Guedel

Miller

Murphy

Guedel

When using an oral airway in infants, a Guedel type (with a central passageway) is probably the best choice, since the infant tongue may easily occlude the lateral slots of other designs (like the Berman), thereby worsening the obstruction.

58
New cards

You observe that a pressure-cycled ventilator fails to cycle off when the patient exhales. The most likely cause of this problem is:

a improper sensitivity setting

an obstructed endotracheal tube

an airway or patient circuit leak

patient-ventilator dyssynchrony

an airway or patient circuit leak

59
New cards

While providing manual ventilation during a transport of a patient with a standard tracheostomy tube, the therapist notices it is difficult to squeeze the resuscitator bag. The therapist should NEXT

add air to the airway cuff.

palpate the neck and clavicle for subcutaneous emphysema.

deflate the cuff and reattempt manual ventilation.

provide manual ventilation with a bag and mask.

deflate the cuff and reattempt manual ventilation.

When it is difficult to provide manual resuscitation through a tracheostomy tube, the first action should be to check the patency of the tube by inserting a suction catheter. The next step would be to deflate the cuff to ensure that it has not herniated over the end of the tracheostomy tube.

60
New cards

You attach a disposable jet nebulizer to a flowmeter at 12 L/min. If no mist is generated, you should:

increase the pop-off pressure to 5 psi

decrease the FIO2

replace the nebulizer

set the flowmeter to 8 L/min

replace the nebulizer

61
New cards

Which of the following are power sources can be used with most portable O2 concentrators?

  1. Household AC current

  2. 12 volt DC car outlet

  3. Portable battery pack

  4. 2 and 3

1 and 2

1 and 3

1, 2 and 3

1, 2, and 3

62
New cards

While performing a test discharge of a metered dose inhaler (MDI) prior to administering to a patient, the respiratory therapist is unable to visually confirm the discharge of any particles or aerosol. The therapist should

Shake the canister and administer the dose to the patient

Administer the dose to the patien

Discard the canister, obtain a new MDI

Utilize a chamber and administer the dose to the patient

Discard the canister, obtain a new MDI

63
New cards

Just prior to providing a patient with manual ventilation, following oral intubation, the respiratory therapist assesses the function of a self-inflating resuscitator bag. The therapist notices the bag is easy to squeeze when attached to the patient. It is also evident that there is no chest rise with each squeeze. Which of the following could be the cause of the problem?

insufficient flow

detached reservoir

bag is too small

faulty inlet valve

faulty inlet valve

On a self-inflating resuscitator bag, there is most likely a faulty inlet valve if the bag is easy to squeeze once it is connected to the patient. In normal circumstances, when the bag is connected to the patient, pulmonary resistance should be felt through the resuscitator device. If no resistance is felt, it is likely that the gas is being ejected through a one-way valve that is responsible for allowing gases to enter the bag. This would suggest a faulty inlet valve that is allowing gases to move both directions.

64
New cards

A large volume jet nebulizer is operating on 70% O2 at 12 L/min. You observe that the aerosol being delivered in short rapid puffs and note a gurgling sound in the system. Which of the following actions would be most appropriate?

increase the nebulizer input flow to 15 L/min

switch the nebulizer entrainment port to 40% O2

drain any accumulated water from the delivery tubing

reduce the water volume in the nebulizer reservoir

drain any accumulated water from the delivery tubing

Aerosols produced by large volume jet nebulizers contain a wide range of particle sizes. Many of the larger unstable particles will impact or deposit by sedimentation in the tubing, causing water to accumulate at low points in the circuit, and intermittently blocking flow (as evidence by the gurgling sound). To ensure proper flow and the set FIO2, this water should be regularly drained AWAY from the patient and discarded (never returned to the reservoir).

65
New cards

A patient with CHF requires 5 L/min humidified oxygen by nasal cannula continuously. The patient is being discharged home. Which of the following modalities is not indicated?

molecular sieve device

portable oxygen tanks

simple mask

pulse-dose conservation device

simple mask

A simple mask requires at least 6 L/min flow and is not appropriate for delivery of 5 L/min

66
New cards

Excessive bubbling is noted in the suction control bottle of a chest tube drainage system. All connections are secure. The respiratory therapist should

reduce the wall suction pressure.

advance the tube further into the chamber

increase the wall suction pressure.

discontinue the chest tubes

reduce the wall suction pressure.

Excessive bubbling in the water-seal compartment of a chest tube drainage system is caused by either a leak in the system, a perforation in the patient's lung tissue, or excessive suction pressure. Of the options offered only a reduction in wall suction pressure will most likely remedy the problem.

67
New cards

A patient is receiving heliox therapy with a mixture of 70% / 30% by nonrebreathing mask to reduce airway resistance. The therapist also observes the reservoir bag is not collapsing at all with each inspiration. After ensuring the face mask is sealed around the face properly, the therapist should

Obtain a partial rebreathing mask

Reduce the total flow of gas mixture to the patient

Obtain an arterial blood gas

Discard the valve between the mask and the reservoir

Reduce the total flow of gas mixture to the patient

When a reservoir bag on and on nonrebreathing mask does not collapse at all with each breath, the cause may be one of two problems: It may be due to an improper seal between the mask and the patient's face. It may also be due to excessive total gas flow going to the reservoir bag.

68
New cards

Immediately after insertion of a #4 laryngeal mask airway (LMA) in a 70 kg adult, you should inflate the cuff to:

60 mL volume

60 cm H2O pressure

10 mL volume

30 cm H2O pressure

60 cm H2O pressure

In general, regardless of the size of the LMA, you should inflate the cuff to 60 cm H2O, while at the same time keeping the maximum inflation volume within that specified by the manufacturer. For a #4 LMA, the maximum recommended cuff inflation volume is 30 mL.

69
New cards

A 40-week gestational age infant is to receive oxygen therapy via a blender set at 35% and humidification through a jet nebulizer. The respiratory therapist should properly set the nebulizer by setting it to

21%

anything over 35%

the same FIO2 as the blender

100%

100%

When using an air-oxygen blender, oxygen percentage is predetermined prior to entering the large volume nebulizer device. When this happens, the large volume nebulizer should be set at 100% in order to avoid entraining additional room air and lowering FIO2.

70
New cards

During a disaster, an H-cylinder with 1000 psi is being used to power a Bird Mark 7 pressure ventilator that is set to deliver 100% oxygen at a tidal volume of 500 mL, respiratory rate of 12/min. How many hours will the tank last?

8

4

6

10

8

To determine the answer to this question, the liter flow per minute must be determined. At a rate of 12 per minute and a tidal volume of 500 mL, the minute ventilation is 6.0 L. Tank duration = 3.14 x 1000 psi = 3,140 L. 3,140 L / 6 L/min = 523 minutes. 523 min / 60 = 8.7 hours. The closest answer is eight hours.

71
New cards

A patient is receiving volume ventilation through a tracheostomy tube. In response to a high pressure alarm, the respiratory therapist attempts suctioning but notices the suction catheter will not pass beyond the end of the tracheostomy tube. Which of the following would be the first action the therapist should take?

Resume volume ventilation

Bronchoscopy

Deflate the cuff

Increase catheter size and suction pressure

Deflate the cuff

Because the catheter will not pass beyond the end of the tracheostomy tube, the most obvious suspicion is a herniated tracheostomy tube cuff. To quickly determine if this is the cause of the high-pressure and the inability to pass the catheter, the cuff should be deflated and the catheter should be inserted again to determine if the blockage remains. If the cather is able to pass, then a herniated cuff is present, and the tube should be replaced

72
New cards

Which of the following is TRUE regarding the FIO2 provided by an IPPB device that uses air-entrainment to enhance flow?

the shorter the inspiratory time, the higher the FIO2

the entrainment ratio is fixed at about 3:1 or 40% O2

the higher the inspiratory pressure, the higher the FIO2

air entrainment increases as inspiratory pressure increases

the higher the inspiratory pressure, the higher the FIO2

73
New cards

You are using a portable pressure-cycled ventilator with a heated humidifier and IPPB circuit to temporarily ventilate an intubated patient who is regaining consciousness in the recovery room. When you check the patient and ventilator, you notice that the inspiratory time is prolonged and the machine does not cycle off without active patient effort. How should you correct the problem?

decrease the sensitivity setting

increase the control pressure

switch the unit to 100% source gas

check/adjust ET tube cuff pressure

check/adjust ET tube cuff pressure

A prolonged inspiratory time and delay in off-cycling with a pressure-cycled ventilator indicate a leak. Leakage may be occurring around the ET tube cuff, which can be rectified by adjusting cuff pressure. Alternatively, leakage may be occurring in the delivery circuit. The two most common sources of circuit leakage would be loose connections at the humidifier or (for IPPB circuits) at the nebulizer manifold.

74
New cards

A respiratory therapist working in a 250-bed medical center has a mechanically ventilated patient who has been receiving volume-controlled ventilation for several days due to ARDS. The patient's compliance has improved significantly in the past 12 hours. Which high alarm parameter should be adjusted?

pressure

respiratory rate

volume

minute ventilation

pressure

With improved lung compliance, peak pressures will diminish with constant tidal volumes. Therefore, the therapist should adjust the peak pressures downward in order to remain 10-15 cm H20 above the basline peak pressures.

75
New cards

Which of the following would cause the FIO2 delivered by an air entrainment mask to be different than that specified by the manufacturer?

an increase in resistance downstream from the jet

a decrease in the patient's peak inspiratory flow

an increase in the flow of oxygen into the jet

obstruction of the patient's nasal passages

an increase in resistance downstream from the jet

An increase in flow resistance downstream from the jet (such as would occur were bedding to cover the mask's outlet ports) decreases the amount of air entrained, thereby decreasing total output flow. A similar effect is produced by obstruction of the air entrainment ports. Under these conditions, the delivered O2 concentration rises, but the actual FIO2 may decrease as room air is inhaled around the mask parts. Changes in O2 input flow do not affect the delivered O2 concentration because for a given size jet and entrainment ports, the mixing ratio is fixed. Patient changes also do not affect the O2 concentration delivered by these devices, but (as described above) may alter the actual FIO2. This occurs most commonly when the patient's peak inspiratory flow exceeds that provided by the device, a problem often seen with systems delivering over 35 to 40% O2. Mouth vs. nose breathing has no impact on full face mask systems.

76
New cards

A venturi or air-entrainment mask will deliver more oxygen than intended if:

the flow is set 5 L/min too high

aerosol is being added through the air-entrainment ports

the holes in the mask are too large

the air-entrainment ports have been blocked

the air-entrainment ports have been blocked

77
New cards

A volume ventilator is providing ventilation to a 32-year-old patient with head trauma and possible neurological impairment. The patient is manually ventilated while the respiratory therapist prepares to transfer the ventilator from ER to ICU. With the ventilator running, the therapist disconnects the compressed air supply hose and hears a sudden new thumping type sound emanating from the ventilator. The therapist should

transfer and set up the ventilator as ordered

inform the receiving nurse and therapist that the ventilator may be faulty

exchange the ventilator for a new one during the transfer

change the air inlet filter prior to use in the ICU

transfer and set up the ventilator as ordered

Most ventilators come equipped with a built-in air compressor. If the ventilator detects a lack of 50 PSI pressure from the air hose, there is an automatic trigger to turn on the internal compressor to provide the appropriate air supply so that gases may be mixed and appropriate oxygen provided. When the compressor activates, it often causes a loud thumping sound that can be heard. Without being familiar with the sound, the practitioner may suspect something is wrong. In fact, it is evidence that the machine is operating properly.

78
New cards

How does CO2 measured from expired gas compare to that from arterial CO2?

up to 10 mm Hg lower

higher immediately following CPR associated with cardiac arrest

about the same

up to 10 mm Hg higher

up to 10 mm Hg lower

79
New cards

At the higher end of its flow range, a high flow nasal cannula provides performance comparable to what other O2 delivery device?

simple mask

nonrebreathing mask

28% air-entrainment mask

transtracheal catheter

nonrebreathing mask

80
New cards

Which of the following is used by the ultrasonic nebulizer to produce droplets from the water reservoir?

heat

steam

baffling

vibration

vibration

81
New cards

In preparation for a helium dilution study, a respiratory therapist is calibrating the helium analyzer. While exposing the analyzer to ambient room air, what will the analyzer read for helium concentration?

2%

79%

0%

21%

0%

To calibrate a helium analyzer, sometimes called a Wheatstone Bridge, the device must be calibrated to room air for the low calibration and to a known level of helium for the high calibration. Because room air has no significant level of helium, helium analyzers should read 0% when exposed to ambient room air conditions.

82
New cards

Which of the following is most accurately descriptive of VC ventilation?

flow is constant until a preset volume is delivered

pressure is constant for a specified period of inspiratory time

inspiration ends when a preset pressure setting is reached

the inspiratory phase terminates after delivery of a preset volume

the inspiratory phase terminates after delivery of a preset volume

83
New cards

A pressure-volume ventilator graphic fails to return to the point of origin. Which of the following could be the cause?

excessive cuff pressure

pulmonary distension

inadequate PEEP

leak in the ventilator circuit

leak in the ventilator circuit

A pressure volume loop on a ventilator should end in the same place it starts. If a pressure volume loop fails to return to zero, or its point of origin, there is likely a leak in the circuit, reducing return volume.

84
New cards

Which of the following factors effect the oxygen concentration delivered by an air entrainment system?

  1. Oxygen input to the jet

  2. Device air to oxygen ratio

  3. Downstream flow resistance

1 and 3

1 and 2

1, 2 and 3

2 and 3

2 and 3

The O2 concentration delivered by an air entrainment system is primarily a function of the air:O2 ratio of the device. The delivered O2 concentration will however rise if resistance to flow increases downstream from the jet (which will also lower the total output flow). Total output flow also varies with both the air:O2 ratio and the O2 input flow to the jet. Changes in O2 input flow to the jet do not affect O2 concentrations because for a given jet and entrainment port size the air:O2 mixing ratio is constant.

85
New cards

You know that an O2 concentrator is working properly if:

it provides 100% O2 at maximum flow

the inlet filter is clean and dry

the flow meter reads 5 L/min when wide open

it provides at least 85-90% O2 at maximum flow

it provides at least 85-90% O2 at maximum flow

Oxygen concentrator FIO2s should be checked and confirmed as part of a routine monthly maintenance visit. Routine maintenance of these devices should include cleaning and replacing filters, checking the alarm system, and confirming the FIO2s using either the unit's oxygen sensor or a separate calibrated O2 analyzer. In general, a properly functioning concentrator should be able to provide at least 85 to 90% O2 at its maximum flow setting.

86
New cards

A pressure cycled ventilator with a heated humidifier and modified IPPB circuit is being used to deliver an aerosolized medication and ventilate a patient in the Recovery Room. When the respiratory therapist checks the patient and ventilator, it is noticed that the inspiratory time is longer than when first charted and the machine does not cycle off until the patient blows out. What should be done to correct the problem?

  1. Tighten the humidifier jar and lid.

  2. Increase the peak pressure.

  3. Tighten the medication nebulizer jar and manifold.

  4. Give the patient 100% oxygen.

1, 2 and 3 only

1 and 2 only

1 and 3 only

2 and 4 only

1 and 3 only

There are clear indications of a leak in the system that would prevent a pressure- cycled IPPB-type machine from cycling off. Tightening the humidifier and medication nebulizer jar should help to seal the leaks. If leaks are not the source of the problem, the machine will likely have to be replaced.

87
New cards

A patient is receiving positive pressure ventilation treatments with Albuterol and normal saline. Before the patient is able to complete exhalation, the machine cycles into inhalation. The respiratory therapist should

decrease flow.

adjust sensitivity.

increase flow.

decrease pressure limit.

adjust sensitivity.

When a ventilator cycles into inspiration prematurely either the rate must be decreased or the sensitivity must be adjusted. Of the options offered adjusting sensitivity is the most appropriate response.

88
New cards

Blood pressure obtained by a sphygmomanometer reads higher than the indicated blood pressure from an arterial line on the same patient. To correct problem, the respiratory therapist should FIRST

advance the arterial line catheter

flush the art line with sodium heparin

check for air bubbles in the transducer dome

discontinue use of the sphygmomanometer

check for air bubbles in the transducer dome

89
New cards

A patient is being paralyzed and intubated in preparation to receive positive pressure ventilation with a volume-cycled ventilator in the control mode. Which of the following alarm settings is most important?

low return-volume alarm

low minute ventilation alarm

low PEEP alarm

high-pressure alarm

low PEEP alarm

This is a difficult question. When a patient is intubated after being paralyzed, they do not have the ability to signal or perform any kind of physical manifestation should they accidentally become disconnected from the ventilator. Thus, a ventilator alarm is paramount. The most sensitive alarm that would signify an accidental disconnection is the low PEEP alarm. On some ventilators this is called the disconnect alarm. Keep in mind, this alarm is more sensitive than the low pressure or low-volume alarms. In other words, the low PEEP alarm responds faster than any other alarm when a disconnection has occured.

90
New cards

A 6-year-old patient has been intubated in the emergency room for respiratory failure at a small rural hospital. The hospital has a volume-cycled ventilator with pediatric mode, but has no pediatric ventilator circuits. What should the respiratory therapist do?

Use an uncuffed endotracheal tube

Use an adult circuit following a pre-operational test

Use an infant designed time-cycled, pressure-limited ventilator

Provide IPPB with normal saline QID

Use an adult circuit following a pre-operational test

A pre-operational test on most microprocessor ventilators will result in a correction to the delivered tidal volume after measuring tubing compliance of the circuit. Even without this correction, the compliance factor differences between adult and pediatric circuits is not considered significant enough to matter. Therefore, adult circuits may be used with pediatric patients if necessary.

91
New cards

Increasing the amount of tubing between the "wye" connector of a dual limb ventilator breathing circuit and the patient's airway will have which of the following effects?

decrease in circuit compliance

increase in inspired PCO2

increase in delivered volume

decrease in rebreathed volume

increase in inspired PCO2

Any extra tubing between the "wye" connector of a dual limb ventilator breathing circuit and the patient's airway will increase mechanical deadspace and rebreathed volume, thereby increasing the inspired PCO2, and potentially raising the PaCO2 (the purpose of adding deadspace).The extra tubing will also increase the overall volume of the circuit. The larger the circuit volume, the greater the circuit compliance and volume lost to gas compression/tubing expansion. The greater this volume loss to the circuit, the less the actual volume delivered into the lungs.

92
New cards

In a dual-limb or 'Y' circuit, you will increase mechanical deadspace if you place any additional tubing between:

the heated humidifier and the 'Y' connector

the 'Y' connector and expiratory valve

the ventilator outlet and expiratory valve

the 'Y' connector and the patient's airway

the 'Y' connector and the patient's airway

93
New cards

A patient is in the intensive care unit receiving volume-controlled ventilation and is hemodynamically unstable. A new balloon-tipped, flow-directed, pulmonary artery catheter has been placed. Immediately following insertion, a chest radiograph shows the tip of the catheter is positioned over the right mid lung near the hilum. Which of the following can the respiratory therapist conclude concerning the catheter?

it should be withdrawn several centimeters

it has punctured the pericardial wall

it should be rotated away from the hilum

the balloon should be inflated and the catheter advanced

the balloon should be inflated and the catheter advanced

The chest radiograph reveals improper placement of the pulmonary artery catheter. To advance the catheter, the balloon should be inflated and the catheter should be sailed into place.

94
New cards

An intubated adult patient with severe expiratory airway obstruction requires ventilatory support. Which of the following capabilities would be most important in selecting a ventilator for this patient?

ability to compensate for airway interface leaks

ability to run on 12 volt DC (battery) power

certification for use during MRI procedures

variable flow control and adjustable I:E ratios

variable flow control and adjustable I:E ratios

Of the functions listed, the most important capability when selecting a ventilator for an intubated adult patient with severe expiratory airway obstruction would be variable flow control and adjustable I:E ratios. This will allow clinicians to make sure the expiratory time is sufficiently long to prevent air-trapping/auto-PEEP

95
New cards

A 10-year-old child does not seem to be performing incentive spirometry well with a flow-oriented device. The patient says the device is boring and difficult to get all three balls floating. The respiratory therapist should

obtain a new incentive spirometer with lighter balls

replace the device with a volume-oriented incentive spirometer

provide encouragement and instruct the patient to continue efforts

test the incentive spirometer with a clean mouthpiece

replace the device with a volume-oriented incentive spirometer

96
New cards

A respiratory therapist is preparing to measure SVC on a patient in the emergency department. Which of the following equipment is required?

respirometer

Collins water-seal spirometer

Geissler tube

Wheatstone bridge

respirometer

97
New cards

While monitoring a newborn, the respiratory therapist notices the TcO2 tracing on the monitor suddenly rises sharply. Which of the following is the mostly likely explanation for this rise?

the transcutaneous electrode is below required temperature

the patient's arterial oxygenation has improved

the transcutaneous electrode requires calibration

transcutaneous electrode has become detached from the skin

transcutaneous electrode has become detached from the skin

98
New cards

A patient is receiving NPPV on the general floor to mobilize secretions by portable CPAP connected to medical wall gas. The threshold resistor is not maintaining consistent positive airway pressure when the patient attempts to sigh. What should the respiratory therapist do?

Increase CPAP level

Instruct the patient to avoid sighing

Increase the flowrate

Replace the CPAP unit

Replace the CPAP unit

In this example we have improperly functioning equipment. If given an option to replace such equipment, the respiratory therapist should always choose to do so. One might be tempted to increase the flow rate in this scenario, however, it will not make a difference with this type of resistor.

99
New cards
<p>After a patient performs a forced vital capacity (FVC) maneuver, you observe the following volume vs. time plot on the screen of a portable spirometer. What validity error does this plot indicate?</p><p>coughing during the maneuver</p><p>too slow a start to forced exhalation </p><p>breathing during the maneuver</p><p>premature end to expiration</p>

After a patient performs a forced vital capacity (FVC) maneuver, you observe the following volume vs. time plot on the screen of a portable spirometer. What validity error does this plot indicate?

coughing during the maneuver

too slow a start to forced exhalation

breathing during the maneuver

premature end to expiration

coughing during the maneuver

This FVC graph indicates coughing during the maneuver. An abrupt pause occurs about a third of the way through the effort, followed by a short inspiratory effort preceding the cough and then an irregular pattern of exhalation. A normal FVC trace would be smooth throughout.

100
New cards

The package insert quality-control material for a blood gas analyzer defines expected values and upper and lower tolerance levels as follows:

Value Tolerance

pH 7.40 0.5%

PCO2 40 mm Hg 5.0%

PO2 80 mm Hg 3%

Which of the following runs should be closely reviewed?

pH 7.36 PCO2 42 mmHg PO2 79 mmHg

pH 7.44 PCO2 38 mmHg PO2 82 mmHg

pH 7.40 PCO2 40 mmHg PO2 78 mmHg

pH 7.45 PCO2 37 mmHg PO2 81 mmHg

pH 7.45 PCO2 37 mmHg PO2 81 mmHg

The answer that shows a pH of 7.45 and a CO2 of 37 should be examined closely as these values appear to be outside the range of tolerance. If these results are accurate the arterial blood gas machine requires calibration and should not be used to report patient results until maintenance has occurred.