Single phase heat unit factor
1
3 phase, 6 pulse heat unit factor
1.35
3 phase, 12 pulse heat unit factor
1.41
high frequency heat unit factor
1.45
Single phase generators
- Full-wave rectified
- 120 pulses per second
- Voltage ripple 100% (voltage drops to 0 between pulses)
- average kV produced is 33% of the set kVp. less efficient.
3 phase, 6 pulse generator
- 3 sources of AC current
- 360 pulse per second
- voltage ripple 13% to 18% (never drops to zero between pulses)
- average kV produced is 91% of the set kVp.
3 phase, 12 pulse generator
- 3 sources of AC current
- 720 pulse per second
- voltage ripple 4% (never drops to zero between pulses)
- average kV produced is 97% of the set kVp.
high frequency generator
- voltage ripple 1% (never drops to zero)
- average kV produced is 99% of the set kVp. The most efficient