1/75
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
other terms for incident dose
dose to Dmax; given dose
ID
incident dose, aka dose to Dmax or given dose
incident dose (aka dose to Dmax or given dose) depends on ___ and ___
beam energy and field size
the primary photon beam also has some low energy ___ and ___ produced by the beam hitting the ___ which adds to dose
electrons and photons; collimators/filters/etc.
larger field sizes produce more or less scatter?
more
why does Dmax change a little bit with larger field sizes? how does the effect change at lower energies?
larger fields produce more scatter. scatter consists of lower energy photons and electrons that deposit their energy on the skin, which skews the ratio and effectively decreases the depth of Dmax.
less of a problem at lower energies
SAD
source to axis distance
D
dose at depth
CR
central ray
F.s.
field size
in higher energy beams, what is greater, the surface dose or the exist dose?
exit dose
a dose increment (usually daily)
fraction
number of fractions and time sequence in which the total dose is given
fractionation schedule
when 2 smaller fractions are given twice per day over a normal time frame, creating a higher total dose (each dose is a little more than just dividing the daily dose in half)
hyperfractionation
when 2 normal fractions are given twice per day, for the same total dose but protraction cut in half
accelerated fractionation
when the total dose is given over a shorter time frame (ex. Canadian breast study)
hypofractionation - i.e. lots of breast is now 5 or 16 fractions
BSF
backscatter factor
OF
output factor
DD or PDD
depth dose, or percentage depth dose
TAR
tissue-air ratio
TMR
tissue-maximum ratio
examples of equipment attenuation factors
block trays, table, pads
examples of patient attenuation factors
DD or PDD
TAR
TMR
radiation that is scattered back by the patient (or phantom)
backscatter factor (BSF)
BSF =
BSF = dose at Dmax in phantom / dose without backscatter (dose in air)
top number is always higher, because dose to Dmax (aka incident dose) is higher because of scatter, meaning BSF is always over 1
backscatter factor (BSF) is always over ___, up to about ___ for larger field sizes
1.0; 1.5
why is BSF not used in linac treatment calculations?
it’s highest for orthovoltage x-rays. less important for linacs with higher energy.
BSF varies with ___
beam energy
at lower energies, ___ and ___ don’t contribute much to surface dose (types of photon interactions)
coherent scatter and photoelectric effect
as beam energy increases, ____ (type of photon interaction) predominates at the skin surface. therefore, backscatter increases with energy up to a point.
Compton effect
scatter becomes more ___ as energy increases, and more ___ as field size increases
forward; prevalent
above 2 MeV, virtually all scatter is in the ___ direction and BSF would now be called ___ with a value of ___
forward; peak scatter factor (PSF); 1.0
PSF
peak scatter factor
replaces BSF above 2 MeV
output factor (OF) is measured in ___ for low energy photons and Co-60, and in ___ for high energy machines
R/min; cGy/MU
the larger the field size, the ___ the output factor (OF)
greater
OF =
dose at Dmax for f.s. (the one you’re using) / dose at Dmax for a standard f.s. (10 × 10)
the O.F for a 10 Ă— 10 field will always be ___, for less than 10 Ă— 10 field will be ____ than 1, and for greater than 10 Ă— 10 field will be ___ than 1
1.0; less than 1; greater than 1
the reason the output factor changes is because of ___
scatter
ESF
equivalent square fields
three methods to find equivalent square fields
chart
ESF = 4 x A/P
2 (WxL)/W + L
how is the equipment attenuation factor written in the MU calculation formula?
Cattn
Cattn
Cattn = dose with device in beam / dose without device in beam
what is the only thing that affects the equipment attenuation factor (Cattn)
beam energy
the equipment attenuation factor, Cattn, is always ___
below 1.0
if the factor is .97, that means the device attenuates 3% of the beam
patient attenuation factors depend on these 4 things
beam energy
field size
treatment depth
sometimes distance (SSD)
all patient attenuation factors (TAR, TMR, PDD) will ___ with increase in energy and field size (scatter) and ___ with increase in depth (attenuation)
increase; decrease
PDD =
PDD = dose at treatment depth / dose at Dmax
how does PDD change with depth, field size, energy, and SSD?
depth = indirect relationship (increase/decrease)
field size, energy, and SSD = direct relationship (increase/increase)
PDD is only used for what type of calculations?
SSD
(because with SAD/isocentric, the whole point is that dose is 100% at the depth we’re treating)
on a PDD chart (relating depth and beam energy to find PDD), the PDD for each energy is always 100% at
Dmax
MU calculation formula for SSD treatment
Nmu =Â ______TD_______
           Ccal x Cfs x Cattn x PDD
TD = tumor dose (divided by number of fields)
Ccal = calibration factor of machine; 1.0 unless told otherwise
Cfs = OF field size correction factor (look up on table)
Cattn = equipment attenuation factors (not always used)
PDD = look up on table
what would the MU calculation equation look like for a single field, 180 cGy, with blocks (TF = .97), Cfs = 1.03, PDD = 88%
Nmu = _______180_______
               1.03 x .97 x .88 Â
for an extended SSD MU calculation, what factors would you have to add to the equation?
ISL correction factor (in denominator) and Mayneord factor (multiply MF by PDD to correct the PDD for the problem)
ISL cf = (SSD1 + Dmax)2 / (SSD2 + Dmax)2
ID =
ID = TD / DD
(incident dose = total dose / depth dose or PDD)
so if total dose is 204.5 cGy, but the tumor is getting 88% of the dose, it gets 180 cGy
ratio of dose in phantom to dose in air
TAR = dose at depth in phantom at SAD / dose without phantom at SAD (dose in air)
tissue-air ratio (TAR) is NOT dependent on ___, and is used for ___ calculations at low energy
SSD; SAD
TAR = BSF at ___
Dmax
backscatter factor (BSF) can only be used for the depth of ___, whereas TAR can be used for any depth
Dmax
TAR was invented for ___ calculations, so we won’t be using them
TAR
the most commonly used patient attenuation factor in dose calculations
tissue maximum ratio (TMR)
because it’s used for SAD/isocentric technique, which is most modern treatments
TMR
tissue maximum ratio
TMR is only used for what kind of calculations?
SAD/isocentric technique
is TMR dependent on SSD?
no
TMR =
TMR = dose at depth / dose at Dmax (ratio of dose at tumor to maximum dose)
equation that relates BSF, TAR, and TMR
BSF = TAR/TMR
equation for MU calculations for SAD/isocentric technique
MU =Â ______TD___________
         Ccal  x Cfs  x Catt  x TMR x SADcfÂ
TD = tumor dose (needs to be divided by number of fields)
Ccal = calibration factor, 1 unless told otherwise
Cfs = OF field size correction factor (look up on table)
Catt = equipment attenuation factors
TMR = will be told
SADcf = SAD factor, a constant for each energy. (SAD = Dmax)2 / (SAD)2
is SADcf a constant?
yes, a constant for each energy
6 MV = (100+1.5)2 / 1002 = 1.03
10 MV = 1.05
what would the MU calculation look like for an SAD treatment, single field, 200 cGy, 10 Ă— 10 cm fs, no blocks, 6 MV (=1.03 SADcf), TMR = 0.8 (look up in appendix 5)
200 / 0.8 Ă— 1.03
isodose curves are lines that represent areas of equal ___
depth dose, or percent depth dose
when looking at isodose curves, lower energies have greater ___ and the size of each isodose area is ___
penumbra; larger
constancy of intensity across the beam (80% of the beam width, because of penumbra)
flatness
intensity difference between opposite sides of the beam
symmetry
the field width/beam edge is defined at the ___% intensity line
50%
the smaller the field, the ___ the flatness
poorer (because of less scatter)
on a dose profile, higher intensity regions beneath the thin edge of the flattening filter
horns / ears
graph that shows flatness and symmetry of the beam
dose profile