main priority when planning for xray images
to ensure that persons in the vicinity of the facility are not exposed to levels of radiations which surpass the regulatory exposure limits
adequate shielding that must be present to meet the following requirements
controlled areas for MRTs = 20 mSv per year
uncontrolled areas for all = 1mSv per year
three steps to ensure adequate shielding
preparation of facility plans
considerations for room design and layout
determination of parameters governing shielding requirements
radiation workers’ dose limits in controlled areas
20mSv per year
non-radiation workers’ dose limits in uncontrolled areas
1mSv per year
pediatric ward uncontrolled area dose limit
0.30 mSv per year
X-ray Workload
a measure of the operational time or the amount of use of the X-ray equipment
occupancy factor (T)
the fraction of time that an area is occupied by an individual who spends the most time there while the X-ray equipment is operating
use factor (U)
the fraction of the workload during which the X-ray beam is pointed in the direction under consideration
Primary Protective Barriers
provide radiation shielding from the direct x-ray beam
eg: gonadal shielding, wall behind the upright bucky or the floor under the table bucky
Secondary Protective Barriers
provide shielding only from scattered and leakage x-rays
eg: ceiling, doors
control booth protective barrier type
a secondary barrier and the xray tube should NEVER be directed towards the control booth
Primary and secondary barriers must be provided when
where the x-ray tube can be moved
Angiographic and radioscopic equipment requires what type of protective barrier
secondary shielding throughout the room
All radiographic, radioscopic, and computed tomography (CT) equipment requirements
Indicators of radiation danger that include warning signs
Specific markings
Mechanical stability
Indicator lights
Indication of loading factors
Irradiation control
Controlling timers
X-ray tube shield
X-ray beam filter
in CR and DR systems the electronic display devices must be assessed how often
daily
Tests to measure geometric distortion, luminance and resolutions should be performed how often
monthly
lens of the eyes annual dose limit
MRT
public
radiation workers: 20mSv
public: 15mSv
skin annual dose limit
MRT
public
radiation workers: 500mSv
public: 50mSv
hands and feet annual dose limit
radiation workers: 500mSv
minimum of lead for certain kV
0.25mm lead → <100kV
0.35mm lead → 100kV<kV<150kV
0.50mm lead → >150kV