Doppler shift is also called
Doppler frequency
Process of extracting the low doppler frequency from the transducers carrier frequency is called
Demodulation
Relative motion between ___ and ___ creates frequency change
Sound source and receiver
This physical principle of frequency change is used to measure ___ of blood in circulation
Velocity
If the reflected sound has a higher frequency than the transmitted sound the doppler shift is?
Positive
When doppler shift is positive blood is moving in which direction
Toward the transducer
The equation for doppler shift is
Reflected frequency - transmitted frequency
What is created when transmitted sound waves strike moving red blood cells?
Doppler shift
When blood cells move away from the transducer the doppler shift is
Negative
Doppler frequencies indicate ___, not speed
Velocity
Speed is purely a
Magnitude
Indicates the distance that a red blood cell moves in 1 second
Speed
Units of speed
Cm/s (distance over time)
Velocity is defined by a ___ and a ___
Magnitude and direction
Doppler shift equation (the longer equation)
2 x velocity of blood × transducer freq x cosine ÷ propagation speed
Doppler shift is ___ related to velocity of blood
Directly
Why is the there a 2 in the doppler equation?
2 doppler shifts
What are the 2 doppler shifts that occur
1. Sound wave strikes RBCS
2. Transducers reception of the sound wave
Modern ultrasound systems measure the ___ ___ between received and transmitted sound waves
Frequency difference
X axis of doppler spectrum represents
Time
Y axis of doppler spectrum represents
Doppler shift or velocity
Relationship between doppler shift and frequency
Direct
Doppler shift has a unit of
Hertz
Measurement of doppler frequency depends on the relationship between
Direction of blood flow and direction in which sound waves propogates
When blood cells are moving ___ to sound beam the entire velocity is measured
Parallel
When an angle exists between the direction of flow and the sound beam the measured velocity is
Less than true velocity
The percentage of true velocity that is measured depends on the ___ of the angle between sound beam and direction of flow
Cosine
What is the relationship between cosine and doppler shift
Direct
When the angle between direction of motion and sound is 0° or 180° flow is
Parallel
Cosine of 0°
1
Cosine 0° indicates what direction of flow
Toward the transducer
Cosine -180° is
-1
Cosine -180° indicates what direction of flow
Away from transducer
When the direction of flow is 90° the measured velocity is
0
The cosine of 60° is
0.5
The cosine of 30°
0.87
The cosine of 120°
-0.5
The cosine of 150°
-0.87
The most common clinic doppler devices are
Bidirectional
Commonly used signal processing technique for bidirectional doppler
Phase quadrature of quadrature detection
Are arterial velocities higher/lower than venous velocities
Higher
Continuous wave doppler requires __ crystals in the transducer
2
Greatest advantage of continuous wave doppler us it's ability to
Accurately measure very high velocities
Primary disadvantage of continuous wave doppler is that
The exact location of the moving blood cells cannot be determined
The limitation of continuous wave doppler is that the exact location of moving blood cells cannot be determined and this is called
Range ambiguity
Another disadvantage of continuous wave doppler is its lack of
TGCs
A dedicated continuous wave transducer is simple because it does not create
Anatomic images
Dedicated continuous wave transducers contain __ elements, each shaped like a __
2, semicircle
Simultaneous imaging and doppler imaging is called
Duplex imaging
Dedicated continuous wave transducers do not use
Backing material
Dedicated continuous wave transducers do not use backing material which results in what
Undampened signal, narrow bandwidth, high q factor and high sensitivity
In pulsed wave doppler how many pzt crystals are necessary?
1
Greatest advantage of pulsed doppler is
Being able to select the exact location where velocities are measured
The advantage of pulsed doppler is called
Range resolution, range specificity, or freedom from range ambiguity artifact
With a pulsed doppler the sonographer positions a small marker called the ___ on a 2d image
Sample volume or gate
During pulsed doppler the transducer emits a sound pulse and then waits the time of flight before listening for a reflection, if the rrasnducer relieves an echo the echo was produced at the
Gate
Primary disadvantage of pulsed doppler is
Inaccurate measurement of high velocity signals
When high velocity flow is incorrectly displayed as traveling in the opposite direction the inaccuracy is called
Aliasing
Pulse doppler transducers are imaging transducers that contain
Backing material
Since pulsed doppler transducers have backing material they have characteristics similar to those of imaging transducers which are -
Low quality factor, Low sensitivity, wide bandwidth
The most common error associated with doppler ultrasound
Aliasing
And alias is a
False identity
Aliasing occurs only with ___ doppler
Pulsed
Occurs when the doppler sampling rate is to low in comparison to the measured blood velocities
Aliasing
Nyquist limit is ___ of the PRF
1/2
Nyquist limit equation
PRF/2
Highest doppler frequency that can be measured without appearance of aliasing
Nyquist limit
2 ways to avoid aliasing are
Raise nyquist limit or reduce doppler shift
The sampling rate of pulsed doppler is the systems
PRF
When sample volume is deep the PRF Is ___ and the nyquist limit is ___
Low, low
Since deeper sample volumes have a lower PRF ___ happens
Aliasing
Higher/lower frequency transducers create more aliasing
Higher
5 techniques used to avoid aliasing artifact
1 adjust scale to max
2 select new ultrasonic view with shallower sample volume
3 select lower frequency transducers
4 use baseline shift
5 use continuous wave doppler
When the sonographer adjusts the scale to its max the ___ is also increased to its max
PRF
The advantage of maximizing PRF is
Raises the nyquist limit and aliasing is less likely to appear
Disadvantage of raising the PRF to max is
Reduces sensitivity to low velocities
What are the disadvantages to selecting a new ultrasonic view with a shallower sample volume
None
Advantage of selecting a lower frequency transducer
Lower shifts occur with lower frequency transducer
Disadvantages of selecting a lower frequency transducer
No significant disadvantages l
Advantage of zero baseline shift
High velocity flows are displayed in proper direction
Advantage of using continuous wave doppler
Aliasing never appears with continuous wave doppler
Eliminating aliasing improves the ability to measure the ___ ___ with doppler
Maximum velocity
Gray shades on a doppler spectrum are related to
Amplitude of reflected signal or number of blood cells creating reflection
Form of 2d doppler where velocity information is coded into colors and superimposed on a 2 dimensional gray scale anatomic image
Color flow doppler
Color doppler provides information on the location of flow and therefore
Is a pulsed ultrasound technique, has range resolution, is subject to aliasing
Color doppler provides information regarding direction of flow and is considered
Semi quantitative
Spectral doppler measures
Peak velocity
Color doppler measures
Mean velocities
Color doppler uses a dictionary or a ___ ___ to concert measured velocities into colors that appear on the image
Lookup table
The two most commonly used dictionaries are called
Velocity mode and variance mode
What do the colors in a velocity mode map mean
Flow direction and velocity
The black region in the middle of the color map indicates
No doppler shift
With color doppler multiple ultrasound pulses are used to accurately determine blood velocities, these multiple pulses are called a
Packet or ensemble
Packets composed of a larger number of Pulses have 2 advantages which are
More accurate velocity measurement and increased sensitivity to low flow
Packets with more pulses have 3 disadvantages which are
More time needed to acquire data, reduced frame rate, decreased temporal resolution
Power mode only identifies the presence of a
Doppler shift
Power mode is ___ color doppler
Non directional
Power doppler is also called
Energy mode or color angio
In power mode the ___ of the reflected signal is processed without regard to the bloods direction or speed
Strength (amplitude)
The amplitude of the reflection is ___ related to the number of moving blood cells
Directly