119 Terms

1
New cards

Doppler shift is also called

Doppler frequency

2
New cards

Process of extracting the low doppler frequency from the transducers carrier frequency is called

Demodulation

3
New cards

Relative motion between ___ and ___ creates frequency change

Sound source and receiver

4
New cards

This physical principle of frequency change is used to measure ___ of blood in circulation

Velocity

5
New cards

If the reflected sound has a higher frequency than the transmitted sound the doppler shift is?

Positive

6
New cards

When doppler shift is positive blood is moving in which direction

Toward the transducer

7
New cards

The equation for doppler shift is

Reflected frequency - transmitted frequency

8
New cards

What is created when transmitted sound waves strike moving red blood cells?

Doppler shift

9
New cards

When blood cells move away from the transducer the doppler shift is

Negative

10
New cards

Doppler frequencies indicate ___, not speed

Velocity

11
New cards

Speed is purely a

Magnitude

12
New cards

Indicates the distance that a red blood cell moves in 1 second

Speed

13
New cards

Units of speed

Cm/s (distance over time)

14
New cards

Velocity is defined by a ___ and a ___

Magnitude and direction

15
New cards

Doppler shift equation (the longer equation)

2 x velocity of blood × transducer freq x cosine ÷ propagation speed

16
New cards

Doppler shift is ___ related to velocity of blood

Directly

17
New cards

Why is the there a 2 in the doppler equation?

2 doppler shifts

18
New cards

What are the 2 doppler shifts that occur

1. Sound wave strikes RBCS

2. Transducers reception of the sound wave

19
New cards

Modern ultrasound systems measure the ___ ___ between received and transmitted sound waves

Frequency difference

20
New cards

X axis of doppler spectrum represents

Time

21
New cards

Y axis of doppler spectrum represents

Doppler shift or velocity

22
New cards

Relationship between doppler shift and frequency

Direct

23
New cards

Doppler shift has a unit of

Hertz

24
New cards

Measurement of doppler frequency depends on the relationship between

Direction of blood flow and direction in which sound waves propogates

25
New cards

When blood cells are moving ___ to sound beam the entire velocity is measured

Parallel

26
New cards

When an angle exists between the direction of flow and the sound beam the measured velocity is

Less than true velocity

27
New cards

The percentage of true velocity that is measured depends on the ___ of the angle between sound beam and direction of flow

Cosine

28
New cards

What is the relationship between cosine and doppler shift

Direct

29
New cards

When the angle between direction of motion and sound is 0° or 180° flow is

Parallel

30
New cards

Cosine of 0°

1

31
New cards

Cosine 0° indicates what direction of flow

Toward the transducer

32
New cards

Cosine -180° is

-1

33
New cards

Cosine -180° indicates what direction of flow

Away from transducer

34
New cards

When the direction of flow is 90° the measured velocity is

0

35
New cards

The cosine of 60° is

0.5

36
New cards

The cosine of 30°

0.87

37
New cards

The cosine of 120°

-0.5

38
New cards

The cosine of 150°

-0.87

39
New cards

The most common clinic doppler devices are

Bidirectional

40
New cards

Commonly used signal processing technique for bidirectional doppler

Phase quadrature of quadrature detection

41
New cards

Are arterial velocities higher/lower than venous velocities

Higher

42
New cards

Continuous wave doppler requires __ crystals in the transducer

2

43
New cards

Greatest advantage of continuous wave doppler us it's ability to

Accurately measure very high velocities

44
New cards

Primary disadvantage of continuous wave doppler is that

The exact location of the moving blood cells cannot be determined

45
New cards

The limitation of continuous wave doppler is that the exact location of moving blood cells cannot be determined and this is called

Range ambiguity

46
New cards

Another disadvantage of continuous wave doppler is its lack of

TGCs

47
New cards

A dedicated continuous wave transducer is simple because it does not create

Anatomic images

48
New cards

Dedicated continuous wave transducers contain __ elements, each shaped like a __

2, semicircle

49
New cards

Simultaneous imaging and doppler imaging is called

Duplex imaging

50
New cards

Dedicated continuous wave transducers do not use

Backing material

51
New cards

Dedicated continuous wave transducers do not use backing material which results in what

Undampened signal, narrow bandwidth, high q factor and high sensitivity

52
New cards

In pulsed wave doppler how many pzt crystals are necessary?

1

53
New cards

Greatest advantage of pulsed doppler is

Being able to select the exact location where velocities are measured

54
New cards

The advantage of pulsed doppler is called

Range resolution, range specificity, or freedom from range ambiguity artifact

55
New cards

With a pulsed doppler the sonographer positions a small marker called the ___ on a 2d image

Sample volume or gate

56
New cards

During pulsed doppler the transducer emits a sound pulse and then waits the time of flight before listening for a reflection, if the rrasnducer relieves an echo the echo was produced at the

Gate

57
New cards

Primary disadvantage of pulsed doppler is

Inaccurate measurement of high velocity signals

58
New cards

When high velocity flow is incorrectly displayed as traveling in the opposite direction the inaccuracy is called

Aliasing

59
New cards

Pulse doppler transducers are imaging transducers that contain

Backing material

60
New cards

Since pulsed doppler transducers have backing material they have characteristics similar to those of imaging transducers which are -

Low quality factor, Low sensitivity, wide bandwidth

61
New cards

The most common error associated with doppler ultrasound

Aliasing

62
New cards

And alias is a

False identity

63
New cards

Aliasing occurs only with ___ doppler

Pulsed

64
New cards

Occurs when the doppler sampling rate is to low in comparison to the measured blood velocities

Aliasing

65
New cards

Nyquist limit is ___ of the PRF

1/2

66
New cards

Nyquist limit equation

PRF/2

67
New cards

Highest doppler frequency that can be measured without appearance of aliasing

Nyquist limit

68
New cards

2 ways to avoid aliasing are

Raise nyquist limit or reduce doppler shift

69
New cards

The sampling rate of pulsed doppler is the systems

PRF

70
New cards

When sample volume is deep the PRF Is ___ and the nyquist limit is ___

Low, low

71
New cards

Since deeper sample volumes have a lower PRF ___ happens

Aliasing

72
New cards

Higher/lower frequency transducers create more aliasing

Higher

73
New cards

5 techniques used to avoid aliasing artifact

1 adjust scale to max

2 select new ultrasonic view with shallower sample volume

3 select lower frequency transducers

4 use baseline shift

5 use continuous wave doppler

74
New cards

When the sonographer adjusts the scale to its max the ___ is also increased to its max

PRF

75
New cards

The advantage of maximizing PRF is

Raises the nyquist limit and aliasing is less likely to appear

76
New cards

Disadvantage of raising the PRF to max is

Reduces sensitivity to low velocities

77
New cards

What are the disadvantages to selecting a new ultrasonic view with a shallower sample volume

None

78
New cards

Advantage of selecting a lower frequency transducer

Lower shifts occur with lower frequency transducer

79
New cards

Disadvantages of selecting a lower frequency transducer

No significant disadvantages l

80
New cards

Advantage of zero baseline shift

High velocity flows are displayed in proper direction

81
New cards

Advantage of using continuous wave doppler

Aliasing never appears with continuous wave doppler

82
New cards

Eliminating aliasing improves the ability to measure the ___ ___ with doppler

Maximum velocity

83
New cards

Gray shades on a doppler spectrum are related to

Amplitude of reflected signal or number of blood cells creating reflection

84
New cards

Form of 2d doppler where velocity information is coded into colors and superimposed on a 2 dimensional gray scale anatomic image

Color flow doppler

85
New cards

Color doppler provides information on the location of flow and therefore

Is a pulsed ultrasound technique, has range resolution, is subject to aliasing

86
New cards

Color doppler provides information regarding direction of flow and is considered

Semi quantitative

87
New cards

Spectral doppler measures

Peak velocity

88
New cards

Color doppler measures

Mean velocities

89
New cards

Color doppler uses a dictionary or a ___ ___ to concert measured velocities into colors that appear on the image

Lookup table

90
New cards

The two most commonly used dictionaries are called

Velocity mode and variance mode

91
New cards

What do the colors in a velocity mode map mean

Flow direction and velocity

92
New cards

The black region in the middle of the color map indicates

No doppler shift

93
New cards

With color doppler multiple ultrasound pulses are used to accurately determine blood velocities, these multiple pulses are called a

Packet or ensemble

94
New cards

Packets composed of a larger number of Pulses have 2 advantages which are

More accurate velocity measurement and increased sensitivity to low flow

95
New cards

Packets with more pulses have 3 disadvantages which are

More time needed to acquire data, reduced frame rate, decreased temporal resolution

96
New cards

Power mode only identifies the presence of a

Doppler shift

97
New cards

Power mode is ___ color doppler

Non directional

98
New cards

Power doppler is also called

Energy mode or color angio

99
New cards

In power mode the ___ of the reflected signal is processed without regard to the bloods direction or speed

Strength (amplitude)

100
New cards

The amplitude of the reflection is ___ related to the number of moving blood cells

Directly