CSSD Water Treatment & Sterilization Quality Tests

0.0(0)
studied byStudied by 0 people
0.0(0)
full-widthCall with Kai
GameKnowt Play
New
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/29

flashcard set

Earn XP

Description and Tags

Fill-in-the-blank flashcards covering water treatment, quality tests, instrument inspection, biological & chemical indicators, and PPE in the CSSD.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

30 Terms

1
New cards

_ (RO) is the water purification method used to supply all water in the CSSD.

Reverse osmosis

2
New cards

Treated, distilled water prevents __ of salts and residues inside sterilizers.

deposition / precipitation

3
New cards

During a water hardness test, if the sample turns blue it is considered .

soft

4
New cards

During a water hardness test, a red or brown colour indicates the water is __.

non-soft / hard

5
New cards

After detecting hard water, the technician must immediately inform the _ responsible for repair.

engineer

6
New cards

Water treatment in the CSSD passes through _ sequential filters.

three (3)

7
New cards

The first filter removes large particles by using small _ inside the tank.

stones / pebbles

8
New cards

The second filter contains resin that binds magnesium and potassium ions.

ion-exchange

9
New cards

The third filter removes excess ions before osmosis.

sodium

10
New cards

An acceptable salt (TDS) level of 0–1 requires testing on a basis.

weekly

11
New cards

If the salt level exceeds 3, the department’s work must be .

stopped

12
New cards

The TDS meter must show turbidity of no more than ____.

50 (ppm/NTU)

13
New cards

Chlorine in treated water must be between ppm.

1–1.5

14
New cards

Water pH should remain between to ensure safe use.

7.2–7.6

15
New cards

The test performed with an empty autoclave to check chamber integrity is called the test.

leak

16
New cards

A failed leak test usually indicates a problem with the autoclave’s or temperature retention.

pressure / vacuum

17
New cards

The Bowie-Dick test checks for adequate removal and steam penetration.

air

18
New cards

A correct Bowie-Dick result shows the indicator sheet uniformly _.

black

19
New cards

Internal chemical indicators are placed every pack or tray before sterilization.

inside

20
New cards

External chemical indicator _ are applied to the outside of packages.

tapes

21
New cards

The washer-disinfector performance is verified with the _ test strip.

TOSI

22
New cards

After a washer load, the TOSI strip must be completely free of simulated .

blood / soil

23
New cards

Visual inspection of instruments is aided by a glass to detect residue.

magnifying

24
New cards

Functional testing of scissors is done by cutting a piece of .

cotton / glove

25
New cards

Ratchet instruments are checked by locking them on the first tooth and striking them times.

three

26
New cards

Protein residue on manually cleaned instruments is detected using a test.

protein

27
New cards

Steam biological indicators contain spores of stearothermophilus.

Geobacillus

28
New cards

Ethylene oxide biological indicators use spores of atrophaeus.

Bacillus

29
New cards

Personal protective equipment must be donned starting with the gown and mask, and removed beginning with the covers.

shoe

30
New cards

Utility gloves used during EtO handling are made of heavy material such as to resist irritation.

cotton