1/100
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Quality Assurance (QA)
Planned and systematic actions to ensure a product or service meets quality requirements.
Quality Control (QC)
Part of QA focused on monitoring and maintenance of equipment performance.
Quality Assurance (QA)
Focuses on the people and process to prevent defects.
Quality Control (QC)
Focuses on identifying and correcting defects.
Quality Assurance (QA)
Proactive quality activity.
Quality Control (QC)
Reactive quality activity.
Quality Assurance (QA)
Usually performed by management and administrators.
Quality Control (QC)
Usually performed by technologists, physicists, engineers.
Acceptance testing, routine performance monitoring, and maintenance.
QC Program Three Steps
Acceptance testing
Initial QC test of new or repaired radiologic equipment to verify performance within manufacturer specifications before clinical use.
Medical physicist
Professional primarily responsible for establishing and overseeing the QC program in imaging facilities.
Positive Beam-Limiting (PBL) Collimator
Automatic collimator that matches the x-ray beam to the size of the image receptor to limit unnecessary radiation.
Should be 31 °C (87 °F) to ensure proper removal of residual chemicals from film.
Wash water temperature
Film analysis
Monthly review categorizing radiographs as good, poor, or reject to identify imaging quality problems.
Processor sludge
Deposits of chemical residue and debris on processor rollers caused by high-throughput, high-temperature operation.
Main goal of QA in radiology
Ensures consistent production of high-quality diagnostic images with minimum exposure to patients and personnel.
Reduce radiation exposure to patients and staff.
Reduce costs by reducing repeat examinations.
Improve department efficiency.
Ensure compliance with regulations and accreditation standards.
Objectives of QA
Food and Drug Administration (FDA)
US federal agency that regulates radiation-emitting products.
Nuclear Regulatory Commission (NRC)
Sets standards for safe use of radiation-producing equipment.
International Commission on Radiological Protection (ICRP)
Publishes recommendations on radiation protection.
International Electrotechnical Commission (IEC)
Publishes performance standards for radiologic equipment.
American College of Radiology (ACR)
Publishes guidelines on QA in diagnostic radiology.
QA Program
Systematic approach to quality involving: organization, data collection, evaluation, and corrective action.
Organization phase of QA
Step involving definition of responsibilities, staff training, and assignment of QA coordinator.
Data collection phase of QA
Step involving gathering data on equipment performance and image quality.
Evaluation phase of QA
Step involving comparison of data to established standards.
Corrective action phase of QA
Step involving implementing measures to address deficiencies.
Level I QC
Checks and adjustments performed by technologists on a daily basis.
Level II QC
More extensive tests performed by medical physicists or engineers quarterly or semiannually.
Level III QC
Specialized tests performed by service engineers after repairs or major changes.
Scheduled maintenance
Refers to routine procedures that are performed usually weekly or monthly.
Preventive maintenance
A planned program of parts replacement at regular intervals.
Non-scheduled maintenance
Unplanned repair work performed when equipment fails unexpectedly.
Reproducibility of exposure
QA test ensuring consistency of x-ray output at same exposure settings.
±5%
Acceptable variance for reproducibility of exposure.
Annually
Frequency of reproducibility of exposure test.
Linearity of exposure
QA test ensuring consistent output for different combinations of mA and time producing same mAs.
±10%
Acceptable variance for linearity of exposure.
Annually
Frequency of linearity of exposure test.
kVp accuracy test
QA test ensuring kVp accuracy compared to indicated kVp.
±5%
Acceptable variance for kVp accuracy.
Annually
Frequency of kVp accuracy test.
Beam uniformity test
QA test ensuring uniformity of x-ray beam intensity across the field.
Annually
Frequency of beam uniformity test.
Light fieldāradiation field congruence
QA test ensuring field size at the image receptor matches the collimator setting.
±2% of SID
Maximum allowed misalignment for light fieldāradiation field congruence.
Annually
Frequency of light fieldāradiation field congruence test.
Perpendicularity test
QA test ensuring the x-ray beam is aligned perpendicular to the image receptor.
Annually
Frequency of perpendicularity test.
X-ray beamāBucky tray alignment
QA test ensuring proper alignment of the x-ray tube, beam, and image receptor.
Annually
Frequency of x-ray beamāBucky tray alignment test.
Filtration test
QA test ensuring total filtration meets minimum standards.
2.5 mm Al equivalent
Minimum total filtration for equipment above 70 kVp.
Annually
Frequency of filtration test.
AEC performance test
Test ensuring exposure is terminated at proper time in AEC systems.
±10%
Acceptable variance for AEC performance.
Annually
Frequency of AEC performance test.
Backup timer test
Test ensuring AEC produces correct exposure regardless of patient thickness.
Annually
Frequency of backup timer test.
Grid alignment test
Test ensuring grid lines are not visible in image.
Annually
Frequency of grid alignment test.
CT QC tests
Tests ensuring CT number accuracy, slice thickness accuracy, and spatial resolution.
Annually
Frequency of CT QC tests.
Mammography QC test
Test ensuring mammography units meet spatial resolution and contrast standards.
Annually
Frequency of mammography QC test.
Fluoroscopy QC test
Test ensuring fluoroscopic units meet maximum exposure rate limits.
Annually
Frequency of fluoroscopy QC test.
Safelight test (fog test)
Darkroom test using film with half covered to check for light leaks.
0.05 OD
Maximum acceptable optical density increase from safelight fog.
Semiannually
Frequency of safelight test.
Coin test
Test using coin on film exposed to safelight to check for fogging.
Semiannually
Frequency of coin test.
Processor QC
Test to ensure processor maintains consistent film density, contrast, and base plus fog.
Daily
Frequency of processor QC testing.
Processor temperature test
Test to check processor temperature stability.
±0.5°C
Acceptable developer temperature variation.
Daily
Frequency of processor temperature test.
Sensitometric control test
Test to monitor replenishment rates and chemical activity.
Daily
Frequency of sensitometric control test.
Clearing test
Test ensuring fixer removes undeveloped silver halide crystals.
Weekly
Frequency of clearing test.
Digital imaging QC
Program ensuring image quality in digital systems by checking monitor resolution, luminance, and contrast.
Monthly
Frequency of digital imaging QC.
SMPTE pattern
Test pattern used for QC of monitors.
Printer QC
Test ensuring printer reproduces images accurately.
Monthly
Frequency of printer QC.
Pinhole camera
Focal spot test that is difficult to use and requires excessive exposure time
Star pattern
It is easy to use but has significant limitations for focal- spot sizes less than 0.3 mm.
Slit camera
The standard for measurement of effective focal-spot size.
±50% for focal spots <0.8 mm, ±40% for 0.8ā1.5 mm, ±30% for >1.5 mm
Acceptable focal spot size variance.
Phototimer
AEC device that uses a light-sensitive detector to terminate the exposure at proper image receptor dose.
6 seconds or 600 mAs
Backup timer should terminate at ___________, whichever occurs first
Intensifying screen
Convert x-ray photons to visible light to expose film, reducing patient dose.
Cleaned at least every other month with manufacturer-recommended cleaner and soft lint-free cloth.
Intensifying screen cleaning schedule
Screen-film contact test
Wire-mesh radiograph used to check for areas of blur indicating poor contact.
Once or twice a year
Frequency for screen-film contact test.
Annually
Frequency for protective apparel QC.
View box intensity should be at least 1500 cd/m² with no more than ±10% non-uniformity.
Illuminator QC standard
Annually
Frequency for illuminator QC.