MRI Registry
TR increases
SNR increases
Scan time increases
Effect on detail NO
SAR decreases
OTHER low T1, more slices
TE increases
SNR decreases
Scan time increases
Effect on detail NO
SAR NO
OTHER high T2, less slices
FOV increases
SNR increases
Scan time no (unless voxel size is constant)
Effect on detail decrease spatial resolution
SAR no
OTHER increase coverage
NSA/NEX increases
SNR increases
Scan time increases
Effect on detail NO
SAR increases
OTHER increase motion
ETL/TSE increases
SNR decreases
Scan time decreases
Effect on detail NO
SAR increases
OTHER increase blurring
TI increases
SNR decreases
Scan time increases (due to increased TR)
Effect on detail NO
SAR NO
OTHER increase tissue saturation
Phase Matrix increases
SNR decreases
Scan time increases
Effect on detail increase spatial resolution
SAR increases (indirect)
OTHER decrease truncation
flip angle increases
SNR increases
Scan time NO
Effect on detail NO
SAR increases
OTHER decrease susceptibility
rBw increases
SNR decreases
Scan time decreases
Effect on detail NO
SAR NO
OTHER decrease chemical shift and metal artifact
PI reduction factor increases
SNR decreases
Scan time decreases
Effect on detail NO
SAR decreases
OTHER decreases DWI susceptibility
DWI b-value increases
SNR decreases
Scan time increases (due to increase TR required)
Effect on detail NO
SAR NO
OTHER increased stroke/neoplasm detection
number of SAT bands increases
SNR decreases
Scan time increases (due to increased TR required)
Effect on detail NO
SAR increases
OTHER decreased reduction of motion/ flow artifacts
slice thickness increases
SNR increases
Scan time NO
Effect on detail decreased spatial resolution
SAR NO
OTHER increase partial volume artifacts
number of slices acquired increases
SNR NO
Scan time increases (due to increase in TR required)
Effect on detail NO
SAR increases
OTHER increase coverage