1/34
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Who does Daily QA checks
Therapist
Who does monthly and annual QA checks
Physics
CT number for accuracy with water is checked on a ______ basis
Daily
Beam Symmetry and Flatness are defined over ____% of the field
80
Interlock systems ensure the safety for who?
Patient
Staff
Machine
How often do you do QA for Gantry collision
Monthly
How often do you do QA for electron Cone collision
Annually
How often do you do QA for Imager collision
Daily
How often do you do QA for Safety Lights
Daily
How often do you do QA for Emergency Off
Monthly
How often do you do QA for AV Systems
Daily
How often do you do QA for Radiation Output
Daily
Tolerance for Radiation Output
3%
Tolerance for Beam Flatness and Symmetry
2%
How often do you do QA for Light Field Congruence
Monthly and After Bulb Change
Tolerance for Light Field Congruence: Asymmetric Jaws
1mm
Tolerance for Light Field Congruence: Symmetric Jaws
2mm
How often do you do QA for Lasers
Daily
Tolerance for Lasers: Non-IMRT
2mm
Tolerance for Lasers: IMRT
1.5mm
Tolerance for Lasers: SRS/SBRT
1mm
Tolerance for Lasers: CT Sim
2mm
How often do you do QA for ODI
Daily
Tolerance for ODI
2mm
How often do you do QA for Collimator Indicator
Daily
Tolerance for Collimator Indicator: Non-IMRT/IMRT
2mm
Tolerance for Collimator Indicator: SRS/SBRT
1mm
How often do you do QA for MLCs
Weekly
How often do you do QA for Gantry and Collimator Mechanical Rotation
Annually
How often do you do QA for Gantry and Collimator Radiation Iso Rotation
Annually
How often do you do QA for Gantry and Collimator Angles
Monthly
Tolerance for Gantry and Collimator Angles
1°
Level 1 Action
What to do?
Out of tolerance?
Inspection
Not out of tolerance
Level 2 Action
What to do?
Out of tolerance?
Schedule Maintenance ASAP
Close to Tolerance
Level 3 Action
What to do?
Out of tolerance?
No treatment, Immediate action
Out of tolerance