1/81
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
A tube housing is required to protect the patient and imaging
personnel from ?
Leakage radiation (off-focus)
Any radiation that doesn’t exit glass window
Leakage radiation
A tube housing is designed so that leakage radiation is measured at what distance from the x-ray source?
1 meter
Tube housing is designed so that leakage radiation does not exceed how much intensity?
100 mR/hr
What are the types of radiation produced in an x-ray room
Primary
Leakage
Scatter
A control panel is located where?
behind protective barrier
What is mandatory for a control panel to have when x-ray tube is energized?
positive indication (can be a sound or light to indicate exposure was taken)
The technical factors in a control panel should or should not be adjustable?
should be
what type of exposure button should control panels have?
Affixed behind the booth (protects techs by keeping them behind booth for exposure)
The x-ray table is made out of what material?
carbon fiber
Why must x-ray tables be made out of carbon fiber?
so minimal x-rays absorbed and reduces PT dose
What film screen is best to reduce PT dose
Higher screen speed
What the common speed for a screen film in general x-ray
400 speed system
Provides a way to measure the distance from the anode focal spot to the image receptor (IR) to ensure that the correct SID is maintained
Source to image receptor distance
what is also used to measure SID
lasers
Tape measure attached to the tube housing
Source-to-Image Receptor Distance Indicator should be within how much of the indicated SID
2%
Used to minimize skin exposure to elections produced by photon interaction with the collimator
Skin Sparring
The patients skin surface should be how much below the collimator for fixed radiographic equipment
15 cm
The patients skin surface should be how much of source to skin distance for mobile radiographic units
30 cm
Most popular is a light-localizing variable-aperture rectangular
collimator (predominant x-ray beam limitation device)
Purpose of a collimator
reduce scatter
The x-ray beam & the light beam must coincide how much w/ the SID
2%
A Positive Beam Limitation (PBL) is also known as what
automatic collimation
For PBL a tech must insure what ?
that collimation is adequate to the IR being used
PBL must be within how much of the SID
2%
Refers to brightness of a surface over a particular field of view
Luminance
Quantifies the intensity of a light source
luminance
what’s the unit for Luminance
•candela per square meter (nit)
Luminance is determined by measuring the concentration of light
over a particular field of view
Total filtration is how much in a tube?
2.5 mm/Al operated above 70 kVp
How much total filtration when operated between 50-70 kVp
1.5 mm
How much total filtration when operated below 50?
0.5 mm
What material is most common for filtration
Aluminum
When should HVL be measured?
once a year, or when x-ray tube is replaced or repair
What is the purpose of radiographic beam filtration?
To increase beam hardness, thereby reducing patient skin dose and the dose to superficial tissues
Used to accomplish dose reduction and uniform imaging of body parts that vary considerably in thickness or tissue composition
Compensating Filters
Consistency in output in radiation intensity for identical generator settings from one individual exposure to subsequent exposures. (Same kVp and mAs)
Exposure Reproducibility
For Exposure Reproducibility what variances is acceptable?
5% or less
Consistency in output radiation intensity at selected kVp settings when settings are changed from one mA and time combination to another
Exposure Linearity
The ratio of the difference in mR/mAs values between two successive generator settings to the sum of those mR/mAs values. It must be less than 0.1
Linearity
When settings are changed from one mA to a neighboring mA station, the most that linearity can vary is
10%
Minimum SSD of at least how much for mobile x-rays
30cm
Exposure switch for mobile is how long?
2m or 6ft
Where should the tech be standing when exposing a PT in a mobile study?
90 degrees from source
What radiographic procedures produce the largest PT radiation exposure rate?
Fluoroscopy
a radiation-reduction technique where the X-ray beam is produced in short, rapid bursts rather than a continuous stream
Intermittent (Pulsed) Fluoroscopy
Because radiation is produced only during pulses, the total exposure time is reduced and reduces PT dose
The tube has cooling time between pulses, reducing heat buildup. Less continuous exposure means less stress on the tube components.
Benefits of pulsing fluoro
Limiting the field size means restricting the x-ray beam to only the area of clinical interest using collimation some benefits are
Reduces patient radiation dose
Decreases scatter radiation
Improves image contrast
Reduces radiation exposure to staff
Regulations require that the x-ray beam cannot be larger than the image receptor or monitor display area
Fluoroscopic beam length and width limitation
Primary protective barrier in fluoro?
Fluoroscopy IR assembly (2mm of lead equivalent required)
A features that increase patient dose in fluro
Magnification mode
Why does magnification mode increase PT dose
An Increase in mA is needed to maintain a image quality as FOV decreases, thus resulting in increase patient dose
Selection of technical exposure factors for adult patients for fluro
75-100 kVp
SSD should not be less than what for stationary fluoro?
38cm (15in)
SSD should not be less than what for mobile fluoro?
30cm (12in)
Designed to ensure that the radiologist is aware of the relative beam-on time during each procedure
Cumulative Timer
When fluro time has exceeded the cumulative timer makes an audible sound after how many minutes
5 min
Position of the input phosphor surface of the image intensifier in relation to the patient should be maintained as close as is practical for what reason?
to reduce PT entrance exposure rate
How much technical exposure factors (kVp) for children be decrease in comparison to adults
25%
Depending on the area of the body being examined with a fluoroscopic image intensification system, for adult patients a range of … kVp is generally used
75-110
Portable x-ray unit that is C-shaped Has an x-ray-tube attached to one end of its arm and an image intensifier attached to the other end. Frequently used in the operating room for orthopedic procedures, cardiac imaging, interventional procedures, and other potentially lengthy tasks
C-arm
the last image formed remains on the monitor so that no further radiation exposure is needed to regenerate it
Last image hold
For Mobile C-Arm Fluoroscopy the Source-to-end of collimator assembly distance is required to be a minimum
30cm (12in)
Patient-image intensifier distance should be kept how?
as short as possible
To reduce PT dose how should the C-arm be position?
preferably under PT
With the C-arm x-ray tube positioned under the patient, scatter radiation is what
less intense
When the C-arm x-ray tube is positioned over the patient, scatter radiation becomes more intense, and radiation exposure of personnel increase or decreases?
Increases
is a quantity that reflects not only the patient radiation dose but also the amount of tissue irradiated
Dose Area Product (DAP)
DAP increases w/ increase field size and decreases w/ decrease field size while dose remains unchanged
DAP detectors positioned just after the collimators before the beam enters the patient
Unit of DAP
mGya-cm2
A DAP meter registers 625 mGy-cm^2 at 70 SID. The SID is then changed to 50SID. Now what is the patient DAP?
SID doesn’t affect DAP
Any wall to which the useful beam can be directed
Primary barrier
The intensity of scatter radiation 1 m from the patient is approximately … of the intensity of the useful beam at the patient
0.1%
Designed to shield areas from secondary radiation. Less thick than primary radiation
Secondary protective barrier
an area that can be occupied by anyone
uncontrolled area
The uncontrolled area has a limit of how much per year and week
week→2mrem/week
year→100mrem/year
It should ensure that no individual will receive more than 2.5 mrem/hr
protective barrier
During fluoroscopy the exposure to the collar is … times greater than the exposure to the trunk of the body under a protective apron
20
Every radiology department should provide a radiation safety program that includes…
training new employees
periodic in-services for all employees
counseling during pregnancy
The preferred way to hold patients who are unable to support themselves is to use a…
restraining device
A reasonable safeguard against accidental irradiation of a pregnant patient would be…
to post warning signs in the department
For diagnostic x-ray personnel, the highest occupational exposure occurs during… exams
Fluoroscopy
mobile radiography