1/50
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
|---|
No study sessions yet.
Quality Control program
are designed to ensure that the CT system is producing the best possible image quality using the minimal radiation dose to the patient.
CT technologists and medical physicists
Responsibility for performing and documenting quality control tests is often shared between
CT technologist
typically perform and record routine quality control tests
medical physicists
testing done by ___ is typically annual or semiannual
medical physicis
is required to obtain necessary dosimetric data
Alignment of table to gantry,
Gantry tilt,
Positioning of the patient support (longitudinal),
Axial patient positioning accuracy,
Collimation test (slice width)
Mechanical tests: (5)
Gantry Tilt
The accuracy of displayed gantry tilt can be assessed by supporting envelope-wrapped x-ray film at the gantry end of the patient table
Gantry Tilt.
The film must be held vertically (e.g. by taping to a Perspex block), so that it is parallel to the sagittal plane and intersects scan and coronal planes at right angles
maximum superior gantry tilt, maximum inferior gantry tilt, 0º gantry tilt
three axial exposures are made using the same film
ruler or tape measure
A ___ placed alongside the table, can be used to check that the degree of couch movement indicated on the gantry agrees with the actual distance moved.
70- 80 kg
A load of approximately ____ should be placed on the table in order to simulate the weight of a patient
table top
gantry
The test should be performed twice: by driving the ___ both away (forward) from and towards (backlash)the ___.
simple mechanical test
In helical scanning, it is not sufficient to use a ___ because the distance imaged depends on couch speed and scanner software
Perspex test object
One method of assessing imaged distance accuracy is to use a ___ containing two small radio-opaque markers, separated by a fixed distance
Scan Projection Radiography (SPR)
20 cm
The test object is scanned in ____ mode and a helical run is planned to start at the first marker and to end at a distance of ____ from the first marker If couch travel is accurate during the helical scan, the markers should be clearly seen on the first and final images of the series
markers
If couch travel is accurate during the helical scan, the ___ should be clearly seen on the first and final images of the series.
Axial Patient Positioning Accuracy
Axial patient positioning accuracy/Agreement between internal and external scan plane lights
Axial Patient Positioning Accuracy
Use the envelope-wrapped film recommended for that measurement, however, a piece of paper or card can also be use
external scan plane ligh
Axial Patient Positioning Accuracy
The wrapped film is placed flat on the table and illuminated by the
Axial Patient Positioning Accuracy
The position of the light is marked on the film envelope and the table is moved automatically to the scan plane
internal light
Axial Patient Positioning Accuracy
If the distance between the internal and external lights is correct, the ___ should now coincide with the mark on the film envelope
Coronal And Sagittal Plane Lights
A long, thin object, with a high CT number relative to air, such as the ‘lead’ in a pencil or a straightened paper clip, can be used as a marker to perform this test
Coronal And Sagittal Plane Lights.
The marker is supported above the patient table and aligned, using the indicating lights, so that it is positioned at the isocentre, parallel to the z-axis and perpendicular to the scan plane
x = 0, y = 0
Coronal And Sagittal Plane Lights
If indicating lights are accurately aligned to the coronal and sagittal planes, the marker should appear as a dot at exactly ___ on the axial image
Slice Width (Tomographic Section Thickness)
Axial mode measurements
Catphan
contains an insert suitable for this test
Dose Profiles (Irradiated Slice Width).
Expose envelope-wrapped x-ray film, supported in air at the isocentre, at each of the slice width settings
Dose Profiles (Irradiated Slice Width)
Once developed, optical density profiles may be plotted using a scanning microdensitometer or the width measured using a ruler
Dose Profiles (Irradiated Slice Width)
To provide an accurate dose profile, a calibration curve can be applied to convert optical density profiles into dose profiles, from which irradiated slice widths may be derived
Accuracy of X-Ray Tube Voltage (kV),
Linearity of X-Ray Tube Current (mA),
Accuracy of Irradiation Time (t),
Consistency of Radiation Output,
Radiation Quality (Total filtration – HVL)
Tests For High Frequency Generators (5)
minimum, maximum, mid range
When more than 3 selections of a CT condition of operation are available then measurement should be carried out at least for ___, ___ and one ___ value of the CT condition of operation.
Radiation Dose Test
Measurement of Computed Tomography Dose Index
CTDI in Perspex Phantoms
The body phantom placed on the patient table and the head phantom is supported in the head rest
CTDI in Perspex Phantoms
Phantoms are aligned centered at the scan isocenter
CTDI in Perspex Phantoms
The ion chamber is inserted into either the central or one of the peripheral cavities of the phantom (all other cavities being filled with Perspex rods
central CTDI
Dose measurements at the center are used to calculate the
Peripheral CTDI
is measured in at least four positions around the phantom, so as to achieve a true average
Test phantoms
are used for the purposes of calibration and evaluation of the performances of CT scanners (supplied by manufacturer)
CT number Uniformity, CT number Linearity, Noise, Low contrast resolution, Spatial resolution
Image Quality Parameters (5)
CT Number Accuracy
depends on tube voltage, filtration, object thickness/attenuation
CT Number Uniformity
It relates to the fact that a CT number of each pixel in the image of an homogeneous object should be the same over various regions
Physical Parameters Of CT Image
Image quality. May be expressed in terms of physical parameters such as uniformity, linearity
CT Number Linearity
It concerns the linear relationship between the calculated CT number and the linear attenuation coefficient of each element of the object
Noise
Is the local statistical fluctuation (standard deviation) of CT numbers of an homogeneous Region Of Interest (ROI)
Noise.
It strongly affects the low contrast resolution
Noise
_____ is dependent on the radiation dose
high contrast resolution
determines the minimum size of detail visualized in the plane of the slice with a contrast >10%
reconstruction algorithm,
detector width,
effective slice thickness,
object to detector distance,
X-ray tube focal spot size,
matrix size
high contrast resolution is affected by (6)
Ionization chamber/ Semiconductor based type survey meter
Water Phantom
Radiation Leakage Levels And Radiological Protection Survey Instruments and Accessories required:
Ionization Chamber/Semiconductor based survey meter
is used to measure the radiation levels
Near control console (operators position),
Outside patient entrance door,
Behind window (if any),
Patient waiting area
Provide The Values Of Maximum Radiation Level (mR/hr) at following places: (4)