Energy range for Grenz
5-15 kV
Superficial Machines energy range
50-150 kV
Superficial therapy machines were also known as ______ _______ _______ ______
Low energy Crookes tubes
Orthovoltage energy range
150-500 kV
Orthovoltage SSD
On the skin surface
What are the 3 main components to the Linac
Gantry
Drive Stand
Couch
What are the 5 main components to the Gantry?
Klystron/Magnetron
Circulator
Accelerating Waveguide
Electron Gun
Cooling System
Where is the source of electrons in the linac
Electron gun
Final e- energy is _________ (directly/indirectly) proportional to the length of the waveguide
Directly
Klystron is used for ____ (low/high) energy linacs and generated _____ (low/high) frequency microwaves
High; high
Magnetron is used for ____ (low/high) energy linacs and generated _____ (low/high) frequency microwaves
Low; High
Which energy source can amplify microwaves?
Klystron
High energy machines have _________ (vertical/horizontal) accelerator tubes
Horizontal
What directs the e- down and out towards the target?
Bending magnet
What are the two types of bending magnets used
90°
270°
Which type of bending magnet minimizes spatial dispersion of the beam?
270°
Intensity of an e- beam is higher in the _______
Center
A raw e- beam is referred to as a ______ beam
Pencil
What is used to spread the electron beam and achieve uniform fluence across the collimated area?
Scattering foil
To make photons/x-rays in a Linac, what happens?
e- beam is accelerated towards the target
E- hit tungsten tsarget
E- decelerate
Energy created is converted to x-ray photons
T/F: flattening filter is used for photons
False
What's the purpose of a flattening filter
To evenly distribute the energy of the beam
The initial exit of the beam is collimated by the _______ (primary/secondary) collimator
Primary
After hitting the primary collimator, what does the beam hit
Ion chamber
Ion chamber purpose
Keeps track of dose rate and beam symmetry
What happens to the beam after the ion chamber?
It hits the secondary collimators (MLCs)
What is the max field size for MLCs
40 × 40
Where is the field size of a linac defined at?
The isocenter
Values regarding machine output and performance are taken during _________ testing
Acceptance
What values are compared during acceptance testing?
Output values to what manufacturer promises