X-ray output constancy (all energies)
3% for Non-IMRT
Electron Output Constancy (if available QD, if otherwise weekly)
3% Non-IMRT
Laser Localization for Non-IMRT
2mm
Laser Localization for IMRT
1.5mm
Laser Localization for SRS/SBRT
1mm
Distance Indicator
2mm for all
Collimator size indicator
2mm for Non-IMRT and IMRT
1mm for SRS/SBRT
Door Interlock
IMRT Functional
Door closing safety
IMRT Functional
AV Monitors
IMRT Functional
Stereotactic Interlocks
SRS/SBRT Functional
Radiation area monitor
IMRT Functional
Beam-on indicator
IMRT Functional
Monthly and Annual QA is based on TG ____
142