the ability of the system to detect weak echoes is called …
sensitivity
target used for measuring axial resolution are
closely spaced targets of varying distances
poor elevation resolution will cause
You will not be able to clear anechoic tubes oriented with their axis in the scan plane
what describes the dead zone
distance from transducer to the first identifiable echo
region of best lateral resolution
at the focal point
what is evaluated using a doppler flow phantom
accuracy of velocity estimation
accuracy of sample volume placement
ability to discriminate different flow directions
accuracy of color registration
all of the above
with using a phantom, which series of pins would be helpful in distinguishing radial resolution
axial resolution
set of curved pins
with using a phantom, what series of pins determine the dead zones
slanted line of pins closest to the top
with a phantom, which set of pins determine vertical depth calibration
pins going up and down in the middle
an abdominal ultrasound scan result reveals a pancreatic mass. CT (gold standard) report states a normal abdominal findings. the US exam is a …
FP, false positive
the ability of a test to correctly identify those without the disease
specificity
__ __ is routine, periodic evaluation of US system to guarantee optimal image quality
quality assurance
what are the four requirements of QA
component testing
repairs
preventive maintenance
record keeping
what are the goals of QA
proper equipment operation
detect gradual changes
minimize downtime
reduce number of repeat scans
reduce number of non-diagnostic exams
what does objective mean
facts, hard evidence that something is going on
what does subjective mean
opinions or feelings
quality assurance programs should be based on ___ standards
objective
what are three deivces used for quality assurance
tissue equivalent phantom
doppler phantom
beam profile / slice thickness phantom
the tissue equivalent phantom evaluates
gray scale
tissue texture
multi and adjustable focus
the tissue equivalent phantom is similar to soft tissue how
prop speed
attenuation
scattering characteristics
echogenicity
a __ __ is a circulation pump stimulate RBC reflections for doppler evaluation
flow phantom
a flow phantom evaluates
CW, PW, and doppler
what are other types of flow phantoms
vibrating string
moving belt
the __ __ phantom measures beam thickness
slice thickness
the greater the slice thickness the __ the spatial resolution is
poorer
what does slice thickness look like
pulls in weak echoes into other structures (thick plane)
for maximum sensitivity, what adjustments should be made
output power and amplification set at max
depth set at its deepest
what is the dead zone
region close to TRX where image is inaccurate
distance from TRX to first visible
with higher f transducers, there will be a __ __ dead zone
more shallow
how to help visualize superficial structures
use an acoustic standoff or gel pack
creates anechoic aerea
what is range accuracy or vertical depth calibration
accuracy in placing reflectors at correct depth parallel to sound beam
what is horizontal calibration
accuracy in placing reflectors in the correct positions perpendicular to sound beam
performance measurements are needed to determine the …
registration accuracy
what is distance measurement accuracy
check for accurate horizontal and vertical caliper measurements from phantom
the distance along the US beam where intensity is the greatest and beam width is narrowest
focal zone
the focal zone improves __ resolution
lateral
since our beam width varies with depth so our __ resolution will also vary at different depths
lateral
a cyst will be
anechoic
a solid structure / tumor will be
solid, filled in, hyperechoic
we must compare our laboratory test to a ….
gold standard
what is included in the gold standard
CT
MRI
arteriogram
surgery
ultrasound is __ accurate in determining gallstones
98%
the gold standard is __ __
always right
what is a true positive result
lab test and gold standard are both positive
what is a false positive
lab test was postive
gold standard was normal or negative
what is a false negative
lab test normal or negative
gold standard postitive
what is a true negative
lab test and gold standard both normal or negative
__ is the percent of correct laboratory tests compared to the gold standard
accuracy
__ is the incidence of disease in the total population
prevalence
the __ __ __ tells us the likelihood that the disease is REALLY present
positive predictive value
the __ __ __ tells us the likelihood that disease is NOT present
negative predictive value
what is the equation for sensitivity
TP / TP + FN
what is the equation for specificty
TN / TN + FP
equation for accuracy
TP + TN / TN + TP + FN + FP
equation for PPV
TP / TP + FP
equation for NPV
TN / TN + FN
what axis of the table are the lab tests presented on
Y axis - side
what axis is the gold standard presented on
x axis - top
what does the statistics table look like
…. +. _
+ . TP. FP
-. FN. TN