1/30
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
Morning QA procedures
Turn machine key switch “On”
12-minute time delay (TDLY)
Rotate gantry to 0 degrees
Turn on MLCs – using HyperTerminal to initialize MLCs (3-5 minutes)
Check water (temp, pressure, and level) and gas pressure
Perform daily safety checks (door interlock, door closing safety, beam on/off, beam on indicator/sign, audiovisual monitors, safety collision interlocks, table motion function)
Check gantry, collimator, and couch position indicators (digital readouts vs. manual protractor*)
Laser and ODI alignment/radiation field light coincidence/imaging checks
Beam output checks
Laser alignment and ODI accuracy steps
Set QA device on couch – use ODI to set surface to 100 SSD
Verify that side lasers align vertically with the top of the QA device
If a discrepancy is found, it may not be immediately obvious whether the problem lies with the ODI or lasers (further tests might be necessary)
Check laser coincidence with crosshairs
Laser alignment and ODI accuracy purpose
Ensure readings are accurate
Incorrect SSD/SAD readings result in the delivered dose differing from the prescribed dose
ODI daily tolerance
2mm
Laser alignment daily tolerance non-IMRT
non-IMRT 2mm
IMRT 1.5mm
SRS/SBRT 1mm
Collimator size indicator steps
Also known as: Field size digital accuracy/light field check
With QA device, align center of device with cross hairs and set a 20x20 cm field. Verify that measured light field is within 1 or 2mm, depending on institution’s acceptable tolerance
Check to determine if a given field size setting, using the collimator size indicator, results in the production of light and radiation fields of that size at the specified treatment distance
Collimator size indicator purpose
Determines if the light and radiation fields are congruent and on the CAX
MLC checks
Perform picket fence image weekly at a minimum
Two types of QA concerns: attenuation properties and positional precision
Test deviations in interleaf transmission
Verify that MLCs are aligned as expected
Collimator angle check purpose
Mechanical or digital readouts indicate the rotational position of the collimator and the beam orientation
Accurate readouts are necessary for the translation of the set-up from treatment planning to the treatment unit to be reproducible and follow the treatment prescription
Collimator angle check steps
Use of a spirit level
Rotate the gantry so that the collimator axis of the rotation is horizontal
Open the jaws so that there is space to rest the small spirit level on the lower jaw (be
careful not to disturb the cross wires)
Rotate collimator until the bubble is exactly centered
If this is meant to be the 180-degree position, the pointer should be at 180 degrees on the scale as well as the digital display
Gantry angle check purpose
For daily reproducibility of setup and the angular position of the various beams used in treatment plans
Gantry angle check steps
Gantry angle may be indicated by means of a protractor and pointer, which move relative to one another with the gantry rotation, by a digital display on the treatment machine, and by a monitor in treatment room and on the treatment console. Can test gantry rotation readout with the use of a spirit level as well
Who performs daily dosimetric checks?
Therapists
Who performs monthly dosimetric checks?
Physicists
How often are output constancy checks and dynamic wedge checks?
Daily basis
How often are checks of flatness, symmetry, and energy?
Monthly basis
Output constancy check steps
Place QA device on treatment couch. Set SSD to 100 and align to field light.
Deliver 100MU at clinically used dose rate (600MU/min) and compare the measured dose with the expected dose of 100cGy for each photon energy and electron beam
The software compares the results with the baseline values (determined during physics calibration)
Output constancy daily tolerance
Must be within 3% (between 97-103cGy if 100cGy is delivered)
Output constancy monthly tolerance
Within 2% (between 98-102cGy if 100cGy is delivered)
Treatment couch position indicators
Therapist again should check manual protractor (if applicable) and/or digital readouts daily
Treatment couch position non-IMRT and IMRT tolerance
2mm/1 degree
Treatment couch position SRS and SBRT tolerance
1mm/0.5 degree
Treatment couch annual tolerances
Tabletop sag – 2mm from baseline
Table angle – 1 degree
Table travels max range movement in all directions 2mm
Light/radiation field coincidence monthly tolerance
2mm for symmetric fields, 1mm for each asymmetric field edge
Collimator, gantry, and couch digital accuracy monthly tolerance
1 degree
Laser positions monthly tolerance IMRT/SRS/SBRT
1mm
Laser positions monthly tolerance non-IMRT
2mm
Annual output check
Uses ionization chamber with an actual liquid water phantom
This large tank water phantom is used to take beam profile scans, depth dose scans, and make other point dose measurements that are related to MU calc parameters
Annual output check tolerance
Within 1%
EDW checks
Produce wedged isodose distributions by varying the amount of time that the beam is on in different portions of the field
The y jaw closes (IN or OUT) so that more intensity is delivered to the part of the field that remains open and less to the part that has been closed TG-142 requires a daily check that EDW are functional
EDW tolerance
CAX output must be within 2% of the commissioning value