Linac QA Procedures

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/30

encourage image

There's no tags or description

Looks like no tags are added yet.

Last updated 10:59 PM on 3/28/26
Name
Mastery
Learn
Test
Matching
Spaced
Call with Kai

No analytics yet

Send a link to your students to track their progress

31 Terms

1
New cards

Morning QA procedures

Turn machine key switch “On”

12-minute time delay (TDLY)

Rotate gantry to 0 degrees

Turn on MLCs – using HyperTerminal to initialize MLCs (3-5 minutes)

Check water (temp, pressure, and level) and gas pressure

Perform daily safety checks (door interlock, door closing safety, beam on/off, beam on indicator/sign, audiovisual monitors, safety collision interlocks, table motion function)

Check gantry, collimator, and couch position indicators (digital readouts vs. manual protractor*)

Laser and ODI alignment/radiation field light coincidence/imaging checks

Beam output checks

2
New cards

Laser alignment and ODI accuracy steps

Set QA device on couch – use ODI to set surface to 100 SSD

Verify that side lasers align vertically with the top of the QA device

If a discrepancy is found, it may not be immediately obvious whether the problem lies with the ODI or lasers (further tests might be necessary)

Check laser coincidence with crosshairs

3
New cards

Laser alignment and ODI accuracy purpose

Ensure readings are accurate

Incorrect SSD/SAD readings result in the delivered dose differing from the prescribed dose

4
New cards

ODI daily tolerance

2mm

5
New cards

Laser alignment daily tolerance non-IMRT

non-IMRT 2mm

IMRT 1.5mm

SRS/SBRT 1mm

6
New cards

Collimator size indicator steps

Also known as: Field size digital accuracy/light field check

With QA device, align center of device with cross hairs and set a 20x20 cm field. Verify that measured light field is within 1 or 2mm, depending on institution’s acceptable tolerance

Check to determine if a given field size setting, using the collimator size indicator, results in the production of light and radiation fields of that size at the specified treatment distance

7
New cards

Collimator size indicator purpose

Determines if the light and radiation fields are congruent and on the CAX

8
New cards

MLC checks

Perform picket fence image weekly at a minimum

Two types of QA concerns: attenuation properties and positional precision

Test deviations in interleaf transmission

Verify that MLCs are aligned as expected

9
New cards

Collimator angle check purpose

Mechanical or digital readouts indicate the rotational position of the collimator and the beam orientation

Accurate readouts are necessary for the translation of the set-up from treatment planning to the treatment unit to be reproducible and follow the treatment prescription

10
New cards

Collimator angle check steps

Use of a spirit level

Rotate the gantry so that the collimator axis of the rotation is horizontal

Open the jaws so that there is space to rest the small spirit level on the lower jaw (be

careful not to disturb the cross wires)

Rotate collimator until the bubble is exactly centered

If this is meant to be the 180-degree position, the pointer should be at 180 degrees on the scale as well as the digital display

11
New cards

Gantry angle check purpose

For daily reproducibility of setup and the angular position of the various beams used in treatment plans

12
New cards

Gantry angle check steps

Gantry angle may be indicated by means of a protractor and pointer, which move relative to one another with the gantry rotation, by a digital display on the treatment machine, and by a monitor in treatment room and on the treatment console. Can test gantry rotation readout with the use of a spirit level as well

13
New cards

Who performs daily dosimetric checks?

Therapists

14
New cards

Who performs monthly dosimetric checks?

Physicists

15
New cards

How often are output constancy checks and dynamic wedge checks?

Daily basis

16
New cards

How often are checks of flatness, symmetry, and energy?

Monthly basis

17
New cards

Output constancy check steps

Place QA device on treatment couch. Set SSD to 100 and align to field light.

Deliver 100MU at clinically used dose rate (600MU/min) and compare the measured dose with the expected dose of 100cGy for each photon energy and electron beam

The software compares the results with the baseline values (determined during physics calibration)

18
New cards

Output constancy daily tolerance

Must be within 3% (between 97-103cGy if 100cGy is delivered)

19
New cards

Output constancy monthly tolerance

Within 2% (between 98-102cGy if 100cGy is delivered)

20
New cards

Treatment couch position indicators

Therapist again should check manual protractor (if applicable) and/or digital readouts daily

21
New cards

Treatment couch position non-IMRT and IMRT tolerance

2mm/1 degree

22
New cards

Treatment couch position SRS and SBRT tolerance

1mm/0.5 degree

23
New cards

Treatment couch annual tolerances

Tabletop sag – 2mm from baseline

Table angle – 1 degree

Table travels max range movement in all directions 2mm

24
New cards

Light/radiation field coincidence monthly tolerance

2mm for symmetric fields, 1mm for each asymmetric field edge

25
New cards

Collimator, gantry, and couch digital accuracy monthly tolerance

1 degree

26
New cards

Laser positions monthly tolerance IMRT/SRS/SBRT

1mm

27
New cards

Laser positions monthly tolerance non-IMRT

2mm

28
New cards

Annual output check

Uses ionization chamber with an actual liquid water phantom

This large tank water phantom is used to take beam profile scans, depth dose scans, and make other point dose measurements that are related to MU calc parameters

29
New cards

Annual output check tolerance

Within 1%

30
New cards

EDW checks

Produce wedged isodose distributions by varying the amount of time that the beam is on in different portions of the field

The y jaw closes (IN or OUT) so that more intensity is delivered to the part of the field that remains open and less to the part that has been closed TG-142 requires a daily check that EDW are functional

31
New cards

EDW tolerance

CAX output must be within 2% of the commissioning value

Explore top notes

note
Chapter 19- Ammonia
Updated 1283d ago
0.0(0)
note
SAT Vocabulary
Updated 702d ago
0.0(0)
note
unit 5 vocab
Updated 381d ago
0.0(0)
note
DNA Damage
Updated 1323d ago
0.0(0)
note
Arthritis Pain of the Elbow
Updated 1142d ago
0.0(0)
note
Chapter 19- Ammonia
Updated 1283d ago
0.0(0)
note
SAT Vocabulary
Updated 702d ago
0.0(0)
note
unit 5 vocab
Updated 381d ago
0.0(0)
note
DNA Damage
Updated 1323d ago
0.0(0)
note
Arthritis Pain of the Elbow
Updated 1142d ago
0.0(0)