1/23
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
quality control (QC)
(set of) procedure(s) intended to ensure product/service adheres to defined set of criteria or meets requirements of client/customer
quality assurance (QA)
program for systematic monitoring and evaluation of aspects of project/service/facility to ensure standards are met
quality improvement (QI)
systematic and continuous actions that lead to measurable improvement in health services and health of patients
elements of quality assurance
methodology
type of equipment, manufacturer’s recommendations
elements of quality assurance
equipment maintenance
type and complexity of tests (short-/long-term)
daily maintenance (replace disposable items)
preventative maintenance (scheduled)
corrective maintenance (unscheduled)
elements of quality assurance
control methods
known test sample used to determine accuracy and precision
needed software and hardware
biologic controls (healthy people)
elements of quality assurance
testing techniques
methods, guidelines, schedules, and protocols
accuracy
how true something is
precision
reproducibility
bias
systematic inaccuracy
linearity
resemblance to a straight line
QA for spirometers
equipment calibration
can measure ≥ 8 L
exhalation maneuvers for ≥ 15 seconds
volume accuracy: ±2.5% or ±0.05 L, whichever is greater
flow range: 0-14 L/sec
3-L syringe accuracy: ±0.5%
flow accuracy: ±5% (-14 to +14 L/sec)
sensitivity: 0.25 L/sec
flow and volume linearity: syringe used 3 times for each
QA for spirometers
calibration schedule
TEST | MINIMUM INTERVAL | ACTION |
volume | daily | calibration check with 3-L syringe |
leak | daily | 3 cmH2O constant pressure for 1 minute |
volume linearity | quarterly | 1-L increments with calibrating syringe |
flow linearity | weekly | test ≥ 3 different flow ranges |
time | quarterly | mechanical recorder check with stopwatch |
QA for spirometers
technique validation (FVC and FEV1)
ATS criteria
repeatability criteria
convert values to BTPS
QA for body plethysmograph
equipment calibration (TLC)
performed daily
mouth pressure via manometer
> ±50 cmH2O at ≥ 8 Hz
mouth flow and volume via 3-L syringe
change in box pressure via small pump
±0.2 cmH2O
biological controls tested regularly (bimonthly)
Levey-Jennings plot
QA for body plethysmograph
technique validation (TLC)
allow sufficient time for temperature stabilization
patient wears nose clips and holds cheeks while panting
repeat panting until ≥ 3 FRCs within ±5%
inspiratory VC after panting
QA for DLCO
equipment calibration
volume measurements (daily)
volume ±2% over 8-L range
He and CO analyzer linearity
max error ≤ ±0.5%
computer timing
±1% in 10 seconds
dead space of circuit <200 mL
QC done weekly using calibrated syringe
QA for DLCO
technique validation
inhaled volume ≥ 90% of best VC
inhale test gas within 4 seconds
alveolar gas sample within 4 seconds of exhalation
minimum of 2 tests (4 seconds between)
tests within 2 mL/min/mmHg of each other
QA for gas analyzers
calibrated to match physiologic range of measurements
2-point calibration
O2, He, CO, NO
calibrated before each patient
ABGs
CLIA
Clinical Laboratory Improvement Amendment
ABGs
pre-analytical
air bubbles
venous admixture
continued metabolism (ice if >15 minutes)
infection control
ABGs
analytical (including hemoximeters)
analyzer calibration/verification
values
pH: 6.840 + 7.384
PO2: 80 + 150
PCO2: 40 + 80
sample testing
review QC and proficiency testing
infection control
closed-circuit with volumetric spirometer
bellows flushed with room-air 5 times between tests
change non-disposable circuits after each patient
infection control
open-circuit
follow department’s guidelines on how to clean/dispose equipment