QC of CT System

0.0(0)
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/60

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

61 Terms

1
New cards

spatial resolution (high resolution)

Can be calculated from the analysis of the spread of information within the system using the MTF. 

2
New cards

line pair phantoms

SPATIAL RESOLUTION (HIGH CONTRAST): Measuring the spatial resolution directly, using a _____________, is a simpler method and is often performed by a technologist. 

3
New cards

visible line pairs

The spatial resolution is given as the maximum number of _______________ (lead strip and space) per millimeter. 

4
New cards

monthly

SPATIAL RESOLUTION (HIGH CONTRAST): This test is performed ________ in most quality assurance programs.

5
New cards

10 to 20 lp/cm

SPATIAL RESOLUTION (HIGH CONTRAST): The spatial resolution of current scanners when images are reconstructed in a high-resolution algorithm is in the range of _______

6
New cards

0.5%

CONTRAST RESOLUTION (LOW CONTRAST): At the minimum, contrast resolution should be such that with a density difference of ____ a 5-mm object can be displayed. 

7
New cards

high

______ contrast objects are easier to resolve

8
New cards

laser lights

___________ located both inside and outside the gantry are used extensively for patient positioning and alignment. 

9
New cards

Accurate laser light performance

__________________ is critical and can be measured, most often using a specific phantom designed for the purpose and provided by the scanner manufacturer. 

10
New cards

2 mm

LASERLIGHT ACCURACY: The light field should coincide with the radiation field to within ______

11
New cards

semiannually

LASERLIGHT ACCURACY: This test is usually performed ___________.

12
New cards

Noise

_______ is measured by obtaining the standard deviation (SD) of the CT numbers within a region of interest (ROI). 

13
New cards

Uniformity

__________ refers to the ability of the scanner to yield the same CT number regardless of the location of an ROI within a homogeneous object. 

14
New cards

water phantom

NOISE AND UNIFORMITY: measured using a ______________

15
New cards

weekly

NOISE AND UNIFORMITY: These tests should be performed on a ________ basis

16
New cards

10

NOISE: A standard deviation of an ROI in a water phantom should not exceed ___

17
New cards

2

UNIFORMITY: there should be no more than a ±__ HU variation from an ROI placed at the center of the water phantom to those placed at the periphery. 

18
New cards

LINEARITY

This refers to the relationship between CT numbers and the linear attenuation values of the scanned object at a designated kVp value.

19
New cards

daily calibrations

LINEARITY: _______________ help to avoid fluctuations in linearity by compensating for these tiny changes. 

20
New cards

scanner malfunction

LINEARITY: Any deviation from the straight line indicates ______________

21
New cards

semiannually

Linearity is typically measured ____________

22
New cards

radiation

These measurements are performed by a medical physicist who follows very detailed specifications to report parameters such as effective dose and dose length product. 

23
New cards
  • Beam hardening

  • Partial volume artifacts

  • Aliasing

  • Edge gradient effect

  • Motion

  • Metallic artifacts

  • Ring artifacts

  • Out-of-field artifacts

  • Tube arcing artifacts

  • Helical and cone beam artifacts

Image artifatcs

24
New cards

Beam hardening

Individual rays are hardened to differing degrees, and this variation cannot be adjusted for by the reconstruction algorithm

25
New cards

Beam hardening is also known as cupping artifacts

26
New cards

dense objects

BEAM HARDENING: The beam is hardened more by ___________ (e.g., more by bone and less by fat).

27
New cards
  1. Filtration

  2. Calibration correction

  3. Beam hardening correction software

features to minimize beam hardening

28
New cards

Partial volume artifacts

occurs when more than one type of tissue is contained within a voxel.

29
New cards

Partial volume artifacts

occurs when a dense object lies to the edge of the FOV.

30
New cards

aliasing

Insufficient projection data

31
New cards

undersampling

aliasing is also known as

32
New cards

fine stripes

Undersampling causes inaccuracies related to reproducing sharp edges and small objects and results in an artifact known as aliasing, in which ____________ appear to be radiating from a dense structure.

33
New cards

slowing ; reducing

Aliasing can be combated by _______ gantry rotation speed (i.e., increasing scan time) or by ________ the helical pitch.

34
New cards

streak artifacts

Metal objects in the SFOV will create

35
New cards

Out-of-field artifacts

caused by anatomy that extends outside of the selected SFOV.

36
New cards

streaks and shading

Out-of-field artifacts appear as ____________ on the image 

37
New cards

third-generation scanners

Ring artifacts occur with ____________ and appear on the image as a ring or concentric rings centered on the rotational axis.

38
New cards

Ring artifacts

They are caused by imperfect detector elements—either faulty or simply out of calibration.

39
New cards

recalibrating

In some instances technologists may eliminate circular artifacts by _________ the scanner. 

40
New cards

Tube arcing artifacts

A common cause of equipment-induced artifact occurs when there is an undesired surge of electrical current (i.e., a short-circuit) within the x-ray tube. 


41
New cards

arcing

________ tends to occur whenever there is a large difference in electrical potential, such as the case between the anode and cathode in an x-ray tube. 

42
New cards

increases

As an x-ray tube ages, the tendency to arc often __________ owing to such factors as degradation of the vacuum within the tube, which results in increased gas pressure.

43
New cards

x-ray output

The arcing causes a momentary loss of __________, which contaminates the x-ray signal collected at the detectors, affecting proper image reconstruction and hence producing artifacts. 

44
New cards

Helical interpolation artifacts

result in subtle inaccuracies in CT number and can be easily misinterpreted as disease

45
New cards

low pitch

Helical interpolation artifatcs can best be avoided by using a _________ whenever possible

46
New cards

windmill artifacts

_____________ appear only on MDCT (Multi-detector CT) helical systems.

47
New cards

Cone beam artifacts

Appear as either streaks or as bright and dark shading near areas of large density differences (e.g., bone and muscle). 

48
New cards

Beam hardening

Cause: X-ray beams are composed of different energies

49
New cards

Aliasing effect

Cause: Too few samples

50
New cards

Edge gradient effect

Cause: Angle of x-ray beam varieties between two similar view

51
New cards

Metallic

Cause: Objects present that are beyond the dynamic range of the scanner

52
New cards

Ring

Cause: Detector problem

53
New cards

Tube arcing

Cause: Electrical surge within the x-ray tube

54
New cards

Spiral Interpolation artifacts

Cause: Images are created from view that are not all in the same plane

55
New cards

Cone beam effect

Cause: Only on MDCT, from the cone-shaped x-ray beam

56
New cards

recalibrate

Corrective step for ring artifact

57
New cards

lower pitch

corrective step for spiral interpolation artifacts

58
New cards

Edge Gradient effect

straight line radiating from high-contrast areas

59
New cards

Tube arcing

no specific pattern

60
New cards

Spiral interpolation artifacts

subtle inaccuracies in CT number

61
New cards

cone beam effect

lines appear in a windmill formation