1/60
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
spatial resolution (high resolution)
Can be calculated from the analysis of the spread of information within the system using the MTF.
line pair phantoms
SPATIAL RESOLUTION (HIGH CONTRAST): Measuring the spatial resolution directly, using a _____________, is a simpler method and is often performed by a technologist.
visible line pairs
The spatial resolution is given as the maximum number of _______________ (lead strip and space) per millimeter.
monthly
SPATIAL RESOLUTION (HIGH CONTRAST): This test is performed ________ in most quality assurance programs.
10 to 20 lp/cm
SPATIAL RESOLUTION (HIGH CONTRAST): The spatial resolution of current scanners when images are reconstructed in a high-resolution algorithm is in the range of _______
0.5%
CONTRAST RESOLUTION (LOW CONTRAST): At the minimum, contrast resolution should be such that with a density difference of ____ a 5-mm object can be displayed.
high
______ contrast objects are easier to resolve
laser lights
___________ located both inside and outside the gantry are used extensively for patient positioning and alignment.
Accurate laser light performance
__________________ is critical and can be measured, most often using a specific phantom designed for the purpose and provided by the scanner manufacturer.
2 mm
LASERLIGHT ACCURACY: The light field should coincide with the radiation field to within ______
semiannually
LASERLIGHT ACCURACY: This test is usually performed ___________.
Noise
_______ is measured by obtaining the standard deviation (SD) of the CT numbers within a region of interest (ROI).
Uniformity
__________ refers to the ability of the scanner to yield the same CT number regardless of the location of an ROI within a homogeneous object.
water phantom
NOISE AND UNIFORMITY: measured using a ______________
weekly
NOISE AND UNIFORMITY: These tests should be performed on a ________ basis
10
NOISE: A standard deviation of an ROI in a water phantom should not exceed ___
2
UNIFORMITY: there should be no more than a ±__ HU variation from an ROI placed at the center of the water phantom to those placed at the periphery.
LINEARITY
This refers to the relationship between CT numbers and the linear attenuation values of the scanned object at a designated kVp value.
daily calibrations
LINEARITY: _______________ help to avoid fluctuations in linearity by compensating for these tiny changes.
scanner malfunction
LINEARITY: Any deviation from the straight line indicates ______________
semiannually
Linearity is typically measured ____________
radiation
These measurements are performed by a medical physicist who follows very detailed specifications to report parameters such as effective dose and dose length product.
Beam hardening
Partial volume artifacts
Aliasing
Edge gradient effect
Motion
Metallic artifacts
Ring artifacts
Out-of-field artifacts
Tube arcing artifacts
Helical and cone beam artifacts
Image artifatcs
Beam hardening
Individual rays are hardened to differing degrees, and this variation cannot be adjusted for by the reconstruction algorithm
Beam hardening is also known as cupping artifacts
dense objects
BEAM HARDENING: The beam is hardened more by ___________ (e.g., more by bone and less by fat).
Filtration
Calibration correction
Beam hardening correction software
features to minimize beam hardening
Partial volume artifacts
occurs when more than one type of tissue is contained within a voxel.
Partial volume artifacts
occurs when a dense object lies to the edge of the FOV.
aliasing
Insufficient projection data
undersampling
aliasing is also known as
fine stripes
Undersampling causes inaccuracies related to reproducing sharp edges and small objects and results in an artifact known as aliasing, in which ____________ appear to be radiating from a dense structure.
slowing ; reducing
Aliasing can be combated by _______ gantry rotation speed (i.e., increasing scan time) or by ________ the helical pitch.
streak artifacts
Metal objects in the SFOV will create
Out-of-field artifacts
caused by anatomy that extends outside of the selected SFOV.
streaks and shading
Out-of-field artifacts appear as ____________ on the image
third-generation scanners
Ring artifacts occur with ____________ and appear on the image as a ring or concentric rings centered on the rotational axis.
Ring artifacts
They are caused by imperfect detector elements—either faulty or simply out of calibration.
recalibrating
In some instances technologists may eliminate circular artifacts by _________ the scanner.
Tube arcing artifacts
A common cause of equipment-induced artifact occurs when there is an undesired surge of electrical current (i.e., a short-circuit) within the x-ray tube.
arcing
________ tends to occur whenever there is a large difference in electrical potential, such as the case between the anode and cathode in an x-ray tube.
increases
As an x-ray tube ages, the tendency to arc often __________ owing to such factors as degradation of the vacuum within the tube, which results in increased gas pressure.
x-ray output
The arcing causes a momentary loss of __________, which contaminates the x-ray signal collected at the detectors, affecting proper image reconstruction and hence producing artifacts.
Helical interpolation artifacts
result in subtle inaccuracies in CT number and can be easily misinterpreted as disease
low pitch
Helical interpolation artifatcs can best be avoided by using a _________ whenever possible
windmill artifacts
_____________ appear only on MDCT (Multi-detector CT) helical systems.
Cone beam artifacts
Appear as either streaks or as bright and dark shading near areas of large density differences (e.g., bone and muscle).
Beam hardening
Cause: X-ray beams are composed of different energies
Aliasing effect
Cause: Too few samples
Edge gradient effect
Cause: Angle of x-ray beam varieties between two similar view
Metallic
Cause: Objects present that are beyond the dynamic range of the scanner
Ring
Cause: Detector problem
Tube arcing
Cause: Electrical surge within the x-ray tube
Spiral Interpolation artifacts
Cause: Images are created from view that are not all in the same plane
Cone beam effect
Cause: Only on MDCT, from the cone-shaped x-ray beam
recalibrate
Corrective step for ring artifact
lower pitch
corrective step for spiral interpolation artifacts
Edge Gradient effect
straight line radiating from high-contrast areas
Tube arcing
no specific pattern
Spiral interpolation artifacts
subtle inaccuracies in CT number
cone beam effect
lines appear in a windmill formation