1/108
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
A young girl comes to the radiology department for scoliosis evaluation. To produce the highest degree of radiation protection for her developing breast tissue, which of the following projections should the technologist use?
posterioanterior (PA)
Which one of the following is not a beam limitation device?
compensating filter
Gonad shielding may reduce female gonad dose by up to ____ and male gonad dose up to ____.
50% - 95%
Which one of the following sets of exposure factors would produce the least patient exposure?
5 mAs at 90 kVp
When using an automatic exposure control (AEC), regulations state that generators must terminate the exposure at 600 mAs or ____ seconds for exposures above 50 kVp.
6
kVp relates to the _____ of the x-ray beam.
quality/energy
The radiographic beam should be collimated so that it is:
no larger than the image receptor
Which field size would lower the patient’s dose the most if all technical factors remained the same?
8 × 10-inch field size
Which one of the following is the MOST versatile type of x-ray beam limitation device?
collimator
When using AEC with digital imaging systems, assuming all other factors are correct, selecting the center chamber on a PA chest image results in:
increased exposure in the lung area
All of the following demonstrate adherence to the As Low As Reasonably Achievable (ALARA) principle for the patient EXCEPT:
wearing occupational dosimeters
Intensity relates to the _____ of the x-ray beam.
quantity
Modern automatic exposure control (AEC) sensors are of the ____ type.
Ionization chamber
All of the following EXCEPT _____ will result in an increase in AEC response/ exposure time.
a. decreasing the kVp
b. increasing the focal spot size
c. increasing the part thickness
d. decreasing the mAs
b. increasing the focal spot size
The rationale behind the development of automatic exposure control (AEC) is to:
Reduce repeats and improve image quality
A backup timer for the automatic exposure control (AEC) serves to:
terminate the exposure in the event of equipment or operator error
AEC devices work by measuring:
radiation that exits the patient
If mAs doubles, what happens to patient dose?
it doubles
400 mA at 0.03 seconds would be equal to ____ mAs.
12
Gonad shielding should be considered if the:
shielding does not interfere with the objectives of the study
person has reasonable reproductive potential
gonads lie wihtin or close to the exposure field
1, 2, and 3
The AEC controls which of the exposure variables?
seconds
The primary function of the filtration is to reduce:
patient skin dose
To reduce patient exposure during pulse mode fluoroscopy, _____ should be chosen.
smallest pulse width
Which one of the following would result in the highest patient dose?
conventional/ continuous fluoroscopy
Automatic brightness control (ABC) in fluoroscopy adjusts which two factors for various part/patient thickness to provide adequate brightness and contrast of the image?
kVp and mA
Which one of the following radiation situations is potentially the most harmful?
a large dose to the whole body all at one time
Which image would produce the highest patient dose?
image B (the one on the right
To radiograph an infant for suspected free air within the abdominal cavity, which of the following projections of the abdomen will demonstrate the condition with the least patient exposure?
left lateral decubitus with grid
The _____ is the primary protective barrier for the fluoroscopic imaging system.
image intensifier / Flat panel
Radiographic grids should be used only when the body part examined is greater than _____ cm. in thickness.
10
Which image would produce the highest patient dose?
image B
The kerma (kinetic energy released in matter) rate, measured in air at the position where the center of the useful beam enters the patient for both fixed and mobile fluoroscopy units should NOT exceed the following entrance skin exposure rates (ESE) is there if no optional high-level control.
100 mGya/min
With pulsed digital fluoroscopy, which pulse system would reduce the patient exposure the most?
3.0 pulses/second
What is the minimum aluminum equivalency of total filtration required for diagnostic radiographic x-ray tubes operating above 70 kVp?
2.5 mm aluminum
During a fluoroscopic examination, which of the following would reduce the patient exposure?
intermittent fluoroscopy
last image hold
use of a grid
collimation
1, 2, and 4
Automatic exposure rate control _____ the quantity of radiation for large patients compared to smaller patients
increases
The source-to-skin distance must be no less than ____ on moble fluoroscopic equipment.
30 cm (12 in)
The source-to-skin distance must be no less than ____ on starionary fluoroscopic equipment.
38 cm (15 in.)
increasing filtration will:
increase the average energy of the x-ray beam
Intensity means the same thing as:
quantity of radiation
When using an automatic exposure control (AEC), regulations state that generators must terminate the exposure at 600 mAs or ____ seconds for exposures above 50 kVp.
6
This factor mainly controls the quantity of radiation produced.
mA/mAs
If mAs is doubled, what happens to patient dose?
patient dose doubles
Kilovoltage (kVp) relates to the:
penetrating power/ energy of the x-ray beam
How many detectors are typically found in an AEC system?
3
Modern automatic exposure control (AEC) sensors function by using ____.
Ionization chambers
When performing a lateral knee examination and only the center chamber is activated, the image receptor (IR) would be _______________ if the center chamber was centered over the patella.
underexposed
AEC devices work by measuring:
radiation that exits the patient
Accurate operation of the AEC device depends on:
The thickness and density of the object
positioning of the object with respect to the photocell
bea, restriction (coning)
1, 2, and 3
The function/purpose of the backup timer when using an AEC is to:
protect the radioprapher from overexposure
protect the patient from overexposure
protect the x-ray tube
2 and 3
When using AEC, the radiographer ___ set the mAs.
does not
When using AEC with digital imaging systems, assuming all other factors are correct, selecting the center chamber on a PA chest image results in:
increased exposure in the lung area
If the center photocell was selected for a lateral projection of the lumbar spine, and it was positioned over the spinous processes (posterior) instead of the certebral bodies, the resulting effect would be:
the IR would be underexposed
All of the following EXCEPT _____ will result in an increase in AEC response time.
increasing the focal spot size
As kVp increases, patient dose:
decreases
Which one of the following sets of exposure factors would produce the least patient exposure?
10 mAs at 80 kVp
The AEC controls which of the follwoing exposure variables?
seconds/time
Total filtration is the sum of _____ filtration plus added filtration.
inherent
The radiographic beam should be collimated so that it is:
No larger than the image receptor
Filters used in radiographic x-ray tubes are generally composed of:
aluminum
Which one of the following is the most versatile type of x-ray beam limitation device?
collimator
Which one of the following is used to limit the area of the patient being irradiated?
collimator
If the filtration for a radiographic unit were inadvertently omitted for a series of chest exposures, the greatest increase in exposures would occur in the:
skin of the chest
Gonad shielding may reduce male gonad dose by up to:
95%
Which field size would lower the patient dose the most if all technical factors remained the same?
8 X 10
Which type of gonadal shield would NOT be suitable for use during a fluoroscopic examination?
shadow shield
Which of the following radiologic examinations would deliver the greatest entrance skin exposure (ESE)?
abdomen
Which one of the following types of gonadal shielding should be used for a patient having an examination requiring a sterile field?
shadow shield
Which one of the following is not a beam limitation device?
filter
What is the minimum aluminum equivalency of total filtration required for diagnostic x-ray tubes operating above 70 kVp?
2.5 mm
The function of a filter is to:
reduce patient skin dose
An x-ray beam is:
heterogeneous
_____ is defined as the thickness of an absorbing material which will reduce the intensity of the primary beam by half of its initial/original value.
half-value layer
What is the minimal HVL for three-phase equipment operating at 90 kVp?
3.1 mm Al
A filter is placed in the path of the x-ray beam to remove:
long-wavelength x-rays
A variable aperture is also known as a:
collimator
Collimators that automatically restrict the beam to the size of the cassette have a feature called automatic collimation or:
positive beam limitation
The most commonly used gonadal shield is the:
contact shield
A flat piece of lead with a circle or square opening in the middle that is used to restrict the x-ray beam is called a:
aperture diaphragm
Gonadal shielding is indicated if the:
shielding does not interfere with the objectives of the study
Person has a reasonable reproductive potential
Gonads lie within or close (5 cm) to the exposure field
1, 2, and 3
Increasing filtration will:
increase the average energy of the x-ray beam
The advantages of beam restriction include which of the following?
less scattered radiation is produced
less biologic tissue is rradiated
less total filtration will be necessary
1 and 2
_____ provides the technologist with an indication of the accuracy of the exposure settings for that given image.
S number or exposure indicator
The largest dose to the male gonads is most likely to result from which of the following exposures?
cross-table lateral hip
Which one of the follwoing grids would produce the least patient exposure?
5:1
12:1
8:1
16:1
5:1
Overall, children are up to ____ times more radiosensitive than adults.
10
Which of the following would help in reducing the radiation exposure to a child during a radiographic study?
the use of gonadal shielding
the use of high-speed imaging system
the use of a tightly collimated beam
1, 2, and 3
Issues to consider when imaging bariatric patients include diameter of fluoroscopic imaging equipment and:
table weight limit
Which one of the following is used as part of an effort to observe the ALARA concept?
Collimation
Which combination of exposure factors would best help eliminate motion for a child having a radiographic procedure?
400 mA, 0.0025 second (10 mAs)
A radiographic image was performed without a grid using 6 mAs. The radiologist wants the image repeated with an 8:1 grid due to excessive scattered radiation on the radiographic image. What new mAs should be used to maintain receptor exposure?
24 mAs
Which one of the following procedures would most likely contribute a higher gonadal dose to a female than a male?
AP abdomen
All of the following can be used to reduce the repeat rate for radiographic images except:
effective communication skills
careful selection of exposure factors
proper immobilization techniques
high ratio radiographic grids
high ratio radiographic grids
Which one of the following positions or projections would provide the greatest radiation protection to the lens of the patient’s eyes?
left lateral skull
right lateral skull
AP skull
PA skull
PA skull
Which of the following radiation situations is potentially the most harmful?
A large dose to the whole body all at one time
All of the following devices are generally used to help reduce patient dose, except:
grid
collimator
gonadal shield
grid
The magnification mode in fluoroscopy will produce:
1. Less patient dose
2. Higher/better contrast resolution
3. Higher/better spatial resolution
2 and 3
According to 21 CFR (Code of Federal Regulations Title 21), how often shall the air kerma rate (AKR) be displayed on fluoroscopic equipment?
once every second
Automatic brightness control during fluoroscopy adjusts which two factors for various part/patient thickness to provide adequate brightness and contrast of the image?
kVp and mA
To reduce patient exposure during pulse mode fluoroscopy, ______ should be chosen.
smallest pulse width