1/74
Looks like no tags are added yet.
Name  | Mastery  | Learn  | Test  | Matching  | Spaced  | 
|---|
No study sessions yet.
Mammography purpose
To detect and diagnose breast diseases, especially breast cancer
Typical x-ray energy for mammography
Low energy x-rays (20–35 kVp)
Target material in mammography tube
Molybdenum (Mo) or Rhodium (Rh)
Reason for using Mo and Rh targets
Provide characteristic x-rays with optimal contrast for soft tissue
Focal spot size for mammography
Small focal spot (0.1–0.3 mm) for high resolution
Compression purpose in mammography
Reduce tissue thickness, improve image quality, and reduce dose
Automatic exposure control (AEC)
Ensures consistent image density by adjusting exposure
Anode angle in mammography
Small (16–24 degrees) to maintain uniform exposure
Heel effect in mammography
X-ray intensity higher on cathode side, used to match thicker chest wall
Grid use in mammography
Reduces scatter radiation, improves contrast
Typical grid ratio in mammography
4:1 or 5:1
Grid movement type
Reciprocating (moving) grid to avoid grid lines
Filtration purpose
Removes low-energy photons that increase dose without improving image
Common filters used
Molybdenum (Mo), Rhodium (Rh), or Aluminum (Al)
Half-value layer (HVL) range for mammography
0.3–0.4 mm Al equivalent
Focal spot blooming
Occurs when high mA causes focal spot enlargement, reducing sharpness
Spatial resolution in mammography
Typically 15 line pairs per mm
Breast composition
Glandular, fibrous, and fatty tissues
Radiographic density in mammography
Determined by tissue composition and thickness
Breast positioning for CC view
Breast compressed horizontally, x-ray beam perpendicular to detector
Breast positioning for MLO view
Breast compressed obliquely, includes pectoralis muscle
Purpose of MLO projection
Visualize upper outer quadrant and axillary tail
Cranio-caudal (CC) view landmarks
Nipple in profile, medial tissue visualized
Mediolateral oblique (MLO) view landmarks
Pectoralis muscle down to nipple line
Axillary tail of Spence
Extension of breast tissue into axilla, best seen in MLO
Magnification mammography
Uses small focal spot and increased OID for detail visualization
Magnification factor
Typically 1.5x to 2.0x
Compression paddle
Transparent device that flattens breast for imaging
AEC detector placement
Under densest portion of breast (usually posterior)
Mammography detector types
Screen-film, CR, or digital flat panel detectors
Advantages of digital mammography
Wide dynamic range, post-processing, lower repeat rates
Phantom used for QC
Simulates breast tissue to evaluate image quality
Phantom image evaluation
Checks for fibers, specks, and masses visibility
MQSA purpose
Mammography Quality Standards Act ensures consistent image quality and safety
Daily QC test
Processor or detector performance check
Weekly QC test
Phantom image quality test
Monthly QC test
Repeat analysis and visual inspection
Semiannual QC test
Compression force and beam quality (kVp and HVL) measurement
Annual QC test
Comprehensive system performance evaluation
Mean glandular dose limit per exposure
Less than 3 mGy for a standard breast
Typical mammography kVp range
25–32 kVp
Typical mAs range
50–120 mAs depending on breast thickness
Focal spot for magnification imaging
0.1 mm small focal spot
Image receptor sizes
18x24 cm and 24x30 cm
Anode heel orientation
Cathode toward chest wall
Artifact causes in mammography
Dust, improper processing, or detector defects
Motion artifact prevention
Proper compression and short exposure time
Digital breast tomosynthesis (DBT)
3D imaging using multiple low-dose projections
DBT advantage
Improves lesion visibility by reducing tissue overlap
DBT projection angle range
15–50 degrees arc around the breast
DBT reconstruction technique
Reconstructs thin slices (1 mm) through the breast
QC for digital detectors
Flat-field test and SNR check
QC for compression system
Measures applied force (should be 25–45 lb or 111–200 N)
QC for AEC system
Checks consistency of optical density across varying thickness
QC for alignment
Ensures x-ray field and detector field coincide
QC for viewbox or monitor
Luminance and uniformity test
QC for darkroom (film systems)
Safe light test and processor temperature control
Breast implants imaging technique
Eklund (implant displacement) technique
Eklund technique purpose
Pushes implant back to image more breast tissue
Spot compression view
Used to evaluate focal lesions under higher compression
Cleavage view
Used to visualize medial breast tissue
Tangential view
Used to show skin lesions or calcifications
Rolled view
Used to separate overlapping tissue
Lateral view (ML or LM)
Used to localize lesions seen on MLO
Implant-displaced CC and MLO
Used for patients with breast implants
Calcification appearance
Tiny high-density spots; can be benign or malignant
Clustered microcalcifications
Suspicious for malignancy
Architectural distortion
Disruption of normal breast pattern; may indicate cancer
Asymmetry definition
Area seen on one view but not corresponding area on another
Fibroadenoma appearance
Well-defined, oval, homogeneously dense mass
Cyst appearance
Round, well-circumscribed, radiolucent center
Malignant mass characteristics
Irregular margins, high density, spiculated appearance
Skin line artifact
From improper positioning or tight compression
Nipple marker use
To identify nipple location or retroareolar lesions
Lead marker use
Indicates lesion, scar, mole, or nipple