Looks like no one added any tags here yet for you.
What amendment is responsible for QA and QC laws
CLIA'88
QC refers mainly to what phase
analytical
Quality Assurance is a continuous improvement process that main 2 purposes is to decrease what?
decrease active errors (obvious). decrease latent errors
Latent errors are errors of what type
organization and design of the lab, overall system errors
quality assurance refers to which phases of the lab
pre-analytical, analytical , post-analytical
What is important in both QA and QC
documentation
QC documents
precision within a run or batch
QC detects
variations in precision (analytic errors)
QC prevents
reporting or erroneous patient values (ensures accuracy)
QC helps with
predicting the future and correcting it in advance
Normal distribution
Gausian (Bell) Curve. The distribution of values of the normal population. What reference range is based on.
Frequency
number of samples with a particular value
average/mean
add all the samples and divide by number of samples
median
middle value
mode
occurs most frequent
Range
interval with upper and lower limit.
Reference range/Normal range
range that the normal population falls within+/_2SD.
QC range
range that QC falls within +/_2SD.
Standard deviation
How much the numbers deviate from the mean
confidence limits
range +/-2SD for reference or QC
QC is in
within confidence limits
QC is out
not in confidence limits.
wider bell curve
greater SD, deviates more, less precise
Skinner bell curve
decrease SD, deviates less
When does a new QC range need to be established
new lot #, change methodology or instrument
How many QC level for a particular test and what are they?
at least 3=low, normal, high
How often is QC run
at least run 2 levels per shift (8hrs.) (CLIA'88) (mostly for chemistry)
How often is QC evaluated
every time you run QC done by MLT, MT, MLS. weekly and monthly done by MT, MLS.
What is the minimum number of data points to determine QC range
20
Who determines QC range
MLT, MT, MLS
Standard deviation measures the degree of...?
imprecision
+/-1SD includes what % of values
68%
+/-2SD includes what % of values
95%
+/-3SD includes what % of values
99.7%
How many days do you run QC serum alone
20-30 days
Levey-Jennings chart has how many charts?
One chart for each level of QC sera for each test (3 at least)
Random errors
no trend or means of predicting
Random errors examples
pipetting inconsistencies, improper mixing of sample, voltage or temperature fluctuations
What to do in case of random errors
re-assay (run again)- results will usually fall within limits
How often will QC be out from inherent variability
5% (document)
Systematic Error will be seen as
a trend (drifting) or shift (abrupt change) in the data
Systematic error affects
accuracy-values consistently too high or too low.
Examples of Systematic errors
poor calibration, reagent deterioration, bad reagent lot #
Westgard Rules applies to?
when 2 levels of control are used
Runs
any time you run QC (may run 2 levels or more within a run)
Across runs
comparing more than one day
Within run
values within same run. compare multiple levels
across levels
comparing different levels within the same run
Which Westgard Rules are Random errors
1-2s, 1-3s, R-4s
Which Westgard Rules are Systematic errors
2-2s, 4-1s,10-x
Which Westgard Rules get Warning
1-2s (possibly 10-x depending on the lab)
Which Westgard Rules get Reject
1-3s, 2-2s, R-4s, 4-1s, (possibly 10-x depending on the lab)
1-2s
one control value exceeds the +/- 2SD (other levels are ok).=random error, warning
1-3s
one control value exceeds the +/-3SD (other levels are ok). = random error, reject
2-2s
2 consecutive values exceed the same + or - 2SD 1)across runs same QC level outside the same 2 SD. OR 2) Within run- each QC level is outside same +or - 2 SD (one each).=systematic error, reject.
R-4s
"Across the range" 1)range between levels within a run exceeds 4 SD (one level +2SD and the other -2SD). Only to be used within a run, NOT across runs. =random error, reject
4-1s
1)four consecutive values exceed the same + or -1SD. OR 2) Same side exceeds + or - 1 SD across control levels 2 of each for a total of 4). =systematic error, reject
10-x
ten consecutive values on the same side of the mean. 1)same control OR 2)across control levels (5 each). = systematic error, warning or reject
Delta check
result for patient today compared to result for patient's past (within certain time frame) is very different. Instrument will notify you with this symbol. Happens everyday.
Proficiency testing is done by
tech with testing reagents sent by CAP
Evaluation of new methods includes
[precision studies, linearity studies =done by tech] [specificity and sensitivity studies=done by manufacturer]
Accuracy studies done by the manufacturer include
interference studies and recovery studies (add known amount of pure substance to serum pool and calculate how much is detected by a particular method)
Who does comparison of method, new vs. old studies
tech
coefficient of variation (CV)
SD expressed as a percentage. Allows comparison of 2 or more methods/instruments that have dissimilar SD ranges. CV = (SD/mean)*100. More reliable when comparing precision at different concentration levels (or units).