Beam-Hardening Artifact Manifestation
Petrous pyramids of skull
Chest and shoulder
Hips
Beam-Hardening Artifact Common Locations
Beam-Hardening
Cause
X-ray beams are composed of different energies
Remedy
Use appropriate filtration
Calibration and correction software
Increase kVp setting
Partial Volume Effect
Cause
More than one type of tissue in a voxel
Dense object at edge of FOV (off-center)
Remedy
Use thinner slices
Fine lines
Aliasing Effect Manifestation
Aliasing Effect
Cause
Too few samples
Remedy
If partial scan used â rescan using complete arc
Increase scan time
Reduce pitch
Straight line from high-contrast areas
barium adjacent to air
Edge Gradient Effect Manifestation
Edge Gradient Effect
Cause
Angle of x-ray beam varies between two similar views
Remedy
Largely unavoidable
Reduced by:
Thinner slices
Use low or neutral HU-value oral contrast instead of barium
Shading
Streaking
Blurring
Ghosting
Motion Artifact Manifestation
Motion Artifact
Cause
Voluntary or involuntary patient motion
Remedy
Give clear breathing instructions
Reinforce importance of staying still
Use positioning aids / immobilization devices
Consider sedation (especially pediatric)
Use shortest scan time
For cardiac â consider β-blockers
Streaks
Metallic Artifact (Star Artifact) Manifestation
Metallic Artifact (Star Artifact)
Cause
Objects beyond dynamic range of scanner
Object absorbs photons
Remedy
Remove metallic objects from SFOV
Angle gantry
Increase technique (kVp)
Use thin slices
Single ring or concentric rings
Ring Artifact Manifestation
Ring Artifact
Cause
Detector problem
Remedy
Recalibrate
If persistent â call service
No specific pattern:
Single streak â severe mottling
Tube Arcing Manifestation
Tube Arcing
Cause
Electrical surge in x-ray tube
Remedy
Call service
Tube replacement
Subtle inaccuracies in CT number
Spiral Interpolation Artifact Manifestation
Spiral Interpolation
Cause
Images created from views not in same plane
Remedy
Lower pitch
Lines in windmill formation
Cone Beam Effect Manifestation
Cone Beam Effect
Cause
Occurs in MDCT
Due to cone-shaped x-ray beam
Remedy
Use manufacturer-recommended pitch
Out-of-Field Artifact
Cause
Anatomy extends outside SFOV
Patient body obstructs detectors
Remedy
Select SFOV larger than patient
Bright and dark streaks
Photon Starvation (Noise) Manifestation
Photon Starvation (Noise)
Cause
Poor exposure technique
Photons pass through thicker anatomy
Remedy
Increase mA
Use filtering
Step-like appearance
Staircase Artifact Manifestation
Staircase Artifact
Cause
Improper:
Slice thickness
Slice incrementation
Occurs in MPR and 3D images
Remedy
Increase mA
Use filtering