If the radiographer is unable to achieve a short object-image distance due to structure of the body part or patient condition, which one of the following adjustments may be made to minimize magnification?
longer source-image distance should be used
What is the single most important factor controlling size distortion?
OID
Spatial resolution is directly related to:
SID
Types of shape distortion include:
foreshortening
magnification
elongation
1 and 3
Which one of the following sets of factors will produce the most magnification?
40 inch SID and a 4 inch OID
The line-focus principle expresses the relationship between:
the actual focal and effective focal spot
To get a magnification factor of 2, where should the object be placed?
halfway between the image receptor and x-ray tube
The effective focal spot is always larger than the actual focal spot.
false
If the SID is 40 inches and the OID is 6 inches, what is the magnification factor?
1.17
If an examination of the skull is to be taken using 400 mA, 1/8 second, 80 kVp, 36 inche SID, 4 inch OID, and if the object size is 8 inches, what will the image size be?
9 inches
As focal spot size increases, distortion would:
focal spot does not affect magnification
An object is radiographed utilizing a 40 inch SID and a 4 inch OID. What would be the resultant effect if the object is radiographed utilizing a 48 inch SID and an 8 inch OID?
increased magnification
Focal spots are represented in millimeters. Example 0.6, 0.9, 1.0, 1.2. Which focal do these numbers represent?
effective focal spot
An object is radiographed utilizing a 40 inch SID and a 4 inch OID. What would be the resultant effect if the object is radiographed utilizing a 44 inch SID and a 2 inch OID?
increased spatial resolution
Which of the following could be utilized to reduce geometric blur resulting from theuse of a large focal spot?
increase kVP
Increase OID
Increase SID
3 only
Use of a perpendicular beam technique should be used whenever possible:
decrease long scae contrast
minimize elongation
minimize shape distortion
2 and 3
A PA projection of the chest was produced with an SID of 72 inches and an OID of 8 inches. What is the MF?
1.12
On a PA chest image taken with an SID of 72 inches and an OID of 6 inches, the size of a round lesion in the left lung measures 2.6 inches in diameter on the radiographic image. What is the object size of this lesion?
2.38
When a focal spot is evaluated, the measurment for size (0.3, 0.6, 1.2 mm); which focal spot is actually being measured?
effective focal spot
Which one of the following would produce the best recorded detail?
72 inch SID, 4 inch OID, small focal spot
What does a line pair mean?
one black line and one white line
The area on the anode where the electrons hit is called the:
actual focal spot
The higher the spatial resolution, the better the recorded detail.
true
A radiograph is performed using 50 mAs, 80 kVp, 40 inch SID, and 16 inch OID and a 2 mm focal spot. The magnification factor for this image is approximately:
1.6
An object measuring 18 cm in length is radiographed at a 100 cm SID and a 10 cm OID. How long will the resulting radiographic image be?
20 cm
A condition caused by angling the xray tube against the main axis of the part is:
foreshortening
Of the following, which one of the following is MOST responsible for radiographic spatial resolution?
focal-spot size
An undistorted radiographic image will result when the object plane and the image receptor plane are:
parallel