CRT Monitors
almost obsolete due to bulkiness and lower image quality compared to modern technologies.
Liquid Crystal Display
most common type of monitor used in medical imaging, good balance of image quality and energy efficiency.
Organic-Light Emitting Diode
the new standard, provides superior contrast and color accuracy, making them ideal for detailed medical imaging.
Pixel
smallest unit of digital image.
Dot Triads
three pixel of red blue and green to combine full spectrum of colors.
Resolution
no. of pixel in a display
3 megapixel
resolution used for radiologist workstation monitor5
5 megapixel
typically used resolution in mammography
10 megapixel
for highest end monitors.
Dot pitch
distance between dots/pixels.
Finer
smaller dot pitch _________ resolution
Refresh Rate
how fast the monitor show the next screen for every second.
motion blur
low refresh rate results to ____.
Aspect ratio
monitor with to the height
candela
measure of luminous density
Luminance
photon energy that reaches the eye.
Contrast/Brightness
range of luminance
93 percent
medical-grade 3-mp LCD monochrome monitor should be able to produce a contrast modulation of approximately ___ between off vs. on pixel.
Ambient light
light surrounding the area of the monitor/
can affect monitor performance, hence necessitating careful control in medical environments.
contrast sensitivity
sensitivity of “average“ human observer to luminance on a standard target.
Contrast Threshold
luminance differential required for the observer to detect a difference.
15-30 minutes
make sure eyes are adapted before making any primary diagnosis.
Spatial Resolution
capability to produce separable images of different points of an object with high fidelity.
Color Temperature
net appearance detected by the eye of the backlights on a monochrome display.
7200 degree kelvin
pure white appearance
10000-12000 degree kelvin
bluish tint appearance.
absence of gross artifacts
overall SMPTE image appearance should be inspected to assure _______.
atleast monthly
monitors used for interpretation should be tested.
linearity
spatial resolution refers to the (linearity vs distortion)
distortion
aliasing refers to the (linearity vs. distortion)
BWH pattern
test of continuous gray scale from the center to outside of the pattern.
no concentric ring like features should be present in the image.
Brigham Women’s Hospital Radiology Department
BWH pattern was developed in this institution.