what does the small m in TPm mean
Modular
how many sample probes on DXC
2
how are modular reagents delivered to reaction cup
a pump
how are the sample and reagent mixed in the modular chemistry reaction cups
by stir bars
what is the name of the creatinine calibrator
Aquacal
How many levels are required to calibrate creatinine
3
what is the purpose of the air displacement syringe
to draw sample into the probe
is it necessary to assign a rack and position number to a calibrator or control that has a barcode label ID?
no
how is the creatinine reagent prepared
reagent combined with picric acid and buffer, then degass overnight
what is the name of the quality control product used for testing creatinine, total protein, and albumin?
synchron 1, synchron 2, synchron 3
under what condition would you use the priority load button
for stat samples
how is the sample probe cleaned in between samples
the wash collar cleans the outside of the probe after the specimen/sample
what is the analysis time for modular chemistry testing
1 minute
what 4 pieces of information are provided on each reagent bottle/cartridge
test name, expiry date, lot number, bar code
which side of analyzer has a refrigerated reagent compartment
Cartridge chemistry
how many probes does the cuvette wash station have
4
what is the purpose of the hydropneumatics system
to provide components such as compressed air vacuum, diluted wash solution, and deionized water to the analyzer
how often does the total protein and albumin need to be calibrated
monthly
what is the purpose of no foam
to prevent bubbles in the waste line
are the cuvettes on the CC side resuable or disposable
reusable