Given an adequate amount of photons being counted, how does the thickness of the Na(TI) crystal affect imaging spatial resolution?
Thinner the crystal, spatial resolution decreases
The ..... sorts out photons of different radionuclides with different photon energies and separates scattered photons from the useful ones
Pulse Height Analyzer
The dose calibrator control procedure performed to assess the device's ability to measure accurately a range of low to high activities is called...?
Linearity
What is the purpose of the pulse height analyzer?
Selects only pulses that correspond to an acceptable range of energies
The ...... of a peak is the physical characteristic of a radionuclide
Energy
At a minimum, how often should Chi-square testing be performed on a well counter?
Quarterly
The ...... the bell-shaped curve of an energy spectrum the poorer the energy resolution
wider
What type of gamma camera bar phantom only requires one acquisition per gamma camera head per week?
Hine-Duley bar phantom
.........is the quantitative expression of camera resolution
Point-spread function
Which of the following is caused when the next pulse is received before the last pulse has returned to baseline?
...
The dose calibrator operates in which section of the Voltage Response curve?
Ionization
The relative variations in count densities between adjacent areas in the image of an object are called?
Contrast
How often should dose calibrator accuracy be performed?
Annually
Which of the following quality control procedures are required for proper functioning of the survey meter?
Constancy and calibration
Which of the following is a common test used to assess medical image display devices?
SMPTE
A technologist needs to assess the gamma camera's uniformity with the collimator. Which of the following will the technologist perform?
Extrinsic flood field
Crystal
converts photon energy into light
Photocathode
converts light into electrical energy
Dynode
Amplifies/multiplies the photon
Pre-amplifier
amplifies pulses produced in PMT
Amplifier
Final amplification and shaping the bell curve
PHA
selects pulses for imaging
The dose calibrator quality control procedure testing a long-lived standard at each of the frequently used radionuclide settings is called....?
constancy
Low-energy all-purpose
Tc-99m
Medium-energy all-purpose
In-111
High-energy all-purpose
I-131
Thyroid(best collimator)
Pinhole
.... is defined as the largest variation in counts over the useful field of view
Integral uniformity
As hole diameter ....., photons at greater angles pass through the collimator without being absorbed
increases
You have one camera in your nuclear medicine department and the weight limit is 500. You have a patient who needs a STAT V/Q scan and he weighs 509 lbs. What is your next course of action?
Call the ordering physician and let them know you cannot scan the patient because they exceed the weight limit
Which of the following quality control procedures are required for proper functioning of the survey meter?
Constancy and calibration
Which filter removes 1/r blurring when reconstructing a SPECT data set using filtered backprojection?
Ramp
The effect of ..... is that some events are lost because they occur too soon after the preceding event, and therefore the count value we measure is not the true count value, but somewhat less than that
Dead time
Which of the following parameters must be specified for a SPECT acquisition?
Time per frame
Image matrix
Number of frames
How often should linearity of a dose calibrator be measured?
Quarterly
The dose calibrator quality control procedure testing a long-lived standard at each of the frequently used radionuclide settings is called...?
Constancy
........ is the most common SPECT acquisition technique in which the camera is stationary during acquisition and does not acquire while rotating between frames
Step-and-shoot acquisition
What type of crystal is used in a nuclear medicine gamma camera?
NaI(TI)
Wipe testing to detect removable contamination in each area of use must be performed...?
Weekly
Which of the following devices converts light into measurable electronic pulses?
Photomultiplier tube
Pre amplifier?
According to the NRC, how often should uniformity floods be performed on a gamma camera?
Daily
If a detector has a FWHM of 68 and is calibrated with Cesium-137, what would the detector efficiency be?
10.3% ((FWHM/photopeak) x 100)
According to the NRC, survey meters must be calibrated ..... and following repairs
Annually
Which of the following is caused when the next pulse is received before the last pulse has returned to baseline?
Pulse pile up?
What is the starting point for the linearity dose calibrator quality control test?
Highest activity used in the department