Which reconstruction algorithm(s) used for PET reconstruction
a. filtered back projection
b. iterative reconstruction
c. attenuation correction
d. a and b
a and b
True or False: Absolute energy calibration allows the conversion of a patient scan into a representation of percent of injected dose per volume or gram of tissue
true
True or False: In PET imaging the FOV along the axis of the patient is approximately 50-100cm
false
a true count in PET is triggered by
the two gamma rays being detected within a certain time of one another
A 3d scanner
has increased random events
CT serves all of the following purposes for PET/CT EXCEPT
iodine contrast enhancement of metastatic tumors
This allows the conversion of a patient scan into a representative of percent injected dose per volume or gram of tissue (SUVās)
absolute activity calibration
True or False: The top row of images represents the images in which attenuation correction has been applied (AC). The bottom row of images is the non-attenuated corrected (NOAC)
false
The two gamma photons generated form the annihilation reaction leave the point with the
same energy and in opposite directions
True or False: Technologists have a lower radiation exposure when working in PET
false
Attenuation correction in a PET/CT scanner is accomplished by
CT scan used for transmission
True or False: SPECT scanners have better spatial resolution than PET scanners
false
Which of the following PET QC procedures is necessary for computation of attenuation correction
blank scan
True or False: syringe shields used in PET/CT are made of lead
false
ā is most effective in occupational protection form PET radionuclides
distance
When a positron radionuclide decays it leads to a āā being emitted
positron
The change detection of photons form unrelated annihilation events within the coincidence timing window is called
random coincidence
What is the best description of the collimation used with dedicated PET scanners
electronic
Thin lead or tungsten septa can be used to block angled photons. If septa are in place, the scanner is referred to as a
2D scanner
Correction are applied during PET reconstruction for which of the following
random events
true events
scatter
a and b
a and c
all of the above
a and c
A path of an annihilation photon pair determined by the electronic collimation and coincidence detection is known as what
line of response
True or False: Coincidence timing calibration adjusts amplifier gains from PMTās
false
True or False: All position emitters have a total energy of 1.02 meV and when annihilation occurs emit two 511 gamma photons
true
Letter C in the image below represents what
scatter
Identify B in the image below
true coincidence detection