In many cases, digital fluoroscopy eliminates which of the following?
Postprocedure "overhead" images
What is the smallest particle of an element that retains the characteristics of the element?
Atom
In digital fluoroscopy, the image must be turned into digital form by what device?
Analog-to-digital converter
X-rays travel as bundles of energy called:
Photons
What is the atomic mass?
The number of protons plus the number of neutrons
What device may be used to ensure consistency of radiographic quality from one exposure to the next?
Automatic exposure control
When a predetermined level of ionization is reached in the ionization chamber, what does the machine do?
The exposure is terminated
What type of x-ray machine uses a continually decreasing mA for the shortest times possible?
Falling load generator
What type of current is required for proper operation of the x-ray tube?
Direct
The law stating that the outer shell of an atom can contain no more than eight electrons is called
Octet rule
Examples of particulate radiation are
Helium nuclei and beta particles
Electromagnetic radiation travels
In waves along a straight path
The latest digital fluoroscopic units use
flat panel plates similar to DR
At what speed do x-rays travel?
The speed of light-186,00 miles per second
Waves of radiation are called?
Sine waves
Wavelength is defined as the distance from
Peak to peak of the wave
Frequency is defined as
The number of waves passing a point per unit time
The speed of x-rays is based on
The fact that they are a form of electromagnetic radiation
Wavelength and frequency are
Inversely proportional to each other
The x-ray beam changes as it travels through the patient by a process called
Attenuation
The intensity of radiation is inversely proportional to the square of the distance between the source of radiation and the person receiving it. This describes the
Inverse square law
As radiation strikes matter
The energy is transferred to the atoms according to the law of conservation of energy
Choose all of the following statements that are accurate regarding electrostatic charges:
A) Electrostatics is the study of electric charges at rest
B) The movement of electrons from one object to another is called ionization
C) Like charges attract, and unlike charges repel
D) Electrostatic charges concentrate on a conductor in the area of greatest curvature
E) Friction, contact, and induction are methods of ionization
A & D
Choose all of the following that are accurate:
A) A magnetic field always surrounds an electrical charge in motion
B) Current flows back and forth in AC
C) Current flows in one direction in DC
D) The volt is the unit of electrical current
E) A conductor allows the flow of electrons
F) The ampere is the unit of electromotive force
G) The volt is the unit of potential difference
H) The path of electrical current is called the circuit
A, B, C, E, G, & H
Electromagnetic induction is the process of causing an electrical current to flow in a conductor:
When it is placed in the magnetic field of another conductor
The two types of electromagnetic induction are
Self-induction and mutual induction
Self-induction is used in the operation of what device?
Autotransformer
The strength of the magnetic fields in a transformer is increased by
Coiling the wires and letting their magnetic fields overlap
Electricity is supplied to the imaging department by a
Generator
The electricity provided to the radiology department is
60 Hz AC
The electricity provided to the radiology department operates at
120 pulses per second
High-frequency power
Has almost no ripple
The primary advantage of three-phase power is that
Voltage never drops to zero
A variable transformer that is used to select kVp for the x-ray circuit is the
Autotransformer
A transformer that has more turns in the secondary coil than in the primary coil is called a
Step-up transformer
What is the transformer used to boost voltage to kilo-voltage levels called?
Step-up transformer
Voltage coming to the x-ray machine is kept constant through the use of a(n)
Line voltage compensator
A step-down transformer
Steps down voltage
Where does thermionic emission occur?
Cathode
Which of the following devices is prereading?
kVp meter
Which device reduces voltage and provides current to produce an electron cloud or space charge at the filament?
Step-down transformer
Which device is electronic, with a setting as low as 0.001?
Timer
What changes AC to DC?
Rectifier
Which of the following is surrounded by a negatively charged focusing cup?
Filament
Which of the following is composed of solid-state silicon-based diodes?
Rectifier
What regulates the duration of x-ray production?
Timer
What is located in the x-ray circuit between the high-voltage transformer and the x-ray tube?
Rectifier
Which of the following measures tube current?
mA meter
What device spins at 10,000-12,000 rpm?
Anode
What device uses maximum heat storage ability of the tube to deliver mAs?
Falling load generator
What is the source of bremsstrahlung and characteristic rays?
Anode
What device increases voltage approximately 500 times?
Step-up transformer
What is the most commonly used AEC?
Ionization chamber
What device always delivers the shortest exposure time possible?
Falling load generator
What device is turned by a rotor?
Anode
Which of the following is located between the patient and the image receptor?
Ionization chamber
The filament is kept warm by
A standby current form the time the x-ray machine is turned on
Activating the rotor
Reduces tube life
When making an exposure, the radiographer should
Activate the rotor and exposure switch in one continuous motion
The process of thermionic emission causes
Electrons to boil off the filament
The force with which the electron stream passes from cathode to anode is a result of___ passing through the x-ray tube.
Kilovoltage
Heat is produced in the x-ray tube as
Electrons interact with the target material
Most of the energy conversion in the x-ray tube produces
Heat
X-rays are produced as incident electrons are slowed going past the nuclei of target atoms by a process called
Bremsstrahlung
X-rays are produced as incident electrons collide with inner shell electrons in target atoms by a process called
Characteristic
What percentage of energy in the x-ray tube is converted to x-rays?
1%
Choose all that are properties of x-rays
-Affect film emulsion
-Scatter and produce secondary radiation
-Invisible to the human eye
-Travel in bundles of energy called photons
-Can ionize matter and gases
-Cause phosphors to fluoresce
The x-ray beam is
Heterogeneous or polyenergetic-consisting of many different energies (wavelengths)
The x-ray emission spectrum consists of
Discrete spectrum (produced by characteristic rays) and continuous spectrum (produced by brems rays)
The primary purpose of filtration is
Radiation protection
The amount of material needed to reduce the intensity of the beam by 1/10 is called
Tenth-value layer
Which of the following statements regarding filtration is true?
Total filtration (not less than 2.5-mm aluminum equivalent) = inherent filtration (glass envelope, tube housing, oil) + added filtration (aluminum)
Calculating heat units for three-phase, 12-pulse equipment requires the use of ____ as a constant; calculating heat units for single-phase equipment requires the use of ___ as a constant; calculating heat units for three-phase, 6-pulse equipment requires the use of ___ as a constant.
1.41, 1.0, 1.35
When a quality control test is performed to ensure that the penetrating ability of the x-ray beam is accurate, the result must be within what amount of the control panel setting?
50%
The primary type of grid used in diagnostic imaging is
Focused
The portion of the image-intensifier tube that converts electron energy to visible light is the
Output phosphor
The portion of the image-intensifier tube that converts visible light to electrons is the
Photocathode
The input phosphor of the image-intensifier tube converts
X-ray energy to visible light
Total brightness gain achieved using an image intensifier equals
Flux gain times minification gain
Single-phase, full-wave rectification produces
Pulsating, direct current with 120 pulses per second and 100% ripple
Three phase, 6 pulse full wave rectification produces
Direct current with 13% ripple
Three phase, 12-pulse full-wave rectification produces
Direct current with 4% ripple
The increase in average photon energy when using three-phase, 6-pulse equipment compared with single-phase equipment is
35%
The increase in average photon energy when using three-phase, 12-pulse equipment compared with single-phase equipment is
41%
Programs that deal with the safe and reliable operation of equipment and programs that address all aspects of the delivery of radiology services are called, respectively
Quality control and quality assurance
An image intensifier's conversion factor is expressed as the ratio of
The luminance of the output phosphor to the exposure rate at the input phosphor
The collimator must be accurate to a level of
+/-2% of SID
kVp must be accurate to within
10% of the control panel setting
Exposure linearity must be accurate to within
10%
Exposure reproducibility must be accurate to within
5%
Fluoroscopy exposure rate may not exceed
100 mGya per minute
The feature that automatically adjusts kVp and mA during fluoroscopy is called
ABC
The test that measures the accuracy of adjacent mA stations is
Exposure linearity
The test that measures the accuracy of successive exposures is
Exposure reproducibility
The accuracy of the timer must be within
5% of the time chosen for exposures over 10 ms
Resolution of the television system may be measured using the following tool(s)
Resolution test pattern
AECs may be tested using
Phantoms
The number of mAs used for analog fluoroscopy is
3 to 5
Marks on the focal track of the anode resulting from bombardment of electrons are called
Pitting
Effective quality control and quality assurance programs are required for accreditation by
The Joint Commission