1/28
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced | Call with Kai |
|---|
No analytics yet
Send a link to your students to track their progress
Goal of CT QC
Ensure the best possible CT image quality with minimal patient dose
QA program in CT must meet accreditation standards of ________ along with vendor specifications
American College of Radiology (ACR)
Quality control testing must also be completed after major repairs and before the first clinical scan after the repair. Examples of repairs include:
Replacement or repair of x-ray tube, generator, collimator assembly or x-ray detectors
Acceptance testing
Performed on new equipment
Routine testing
Performed on existing equipment on an ongoing basis
3 basic principles of QC
Test performed on a regular basis
Test measurements should be documented on a designated form
Can use QA data form provided by the manufacturer but each facility can design their own forms
Indicate if results are within specified guidelines
Equipment used in CT QA tests
Phantom (along with test protocols to ensure optimal operation is within established tolerances)
LAP IsoMark (test the gantry laser and BB alignment)
Laser check device (check the room’s external lasers)
Tube conditioning
A process which allows the x-ray tube to come to its normal operating temperature after a pause in system operation of more than 10 hours
High-contrast (spatial) resolution
Minimum distance between two objects that allows them to be seen as separate and distinct
Low-contrast resolution
Subtle differences in tissue densities from one region of anatomy to another
Image uniformity
The ability of the CT scanner to yield the same CT number regardless of the location of the ROI within a homogenous object
Noise
Represents the portion of the CT image that contains no useful information
Produces a “salt-and-pepper” appearance or grainy quality in the image
What is the primary factor limiting low-contrast resolution in CT?
Image noise
Sources of noise in a CT image
Object size, reconstruction algorithms, detector efficiency, and artifacts
As noise increases on an image, the edge definition of anatomic borders and subtle differences in attenuation between tissues ________
Decreases
Methods to minimize noise
Tube warm-ups
Daily system calibrations
Pitch (spiral pitch ratio)
Describes the relationship between the patient couch movement through the CT scanner and the x-ray beam width. Changing pitch may cause overlap or gaps in beam width during table movement
CT numbers represent the ________ within the body according to their atomic number and physical density
Attenuation values of different structures
Daily tolerance limit of water
0±5 HU
HU of air
-1000
HU of adipose tissue
-100 (slight range of -50 to -100 HU)
HU of water
0
HU of muscle
50 (range of 10 to 50 HU)
HU of bone
1000
Alignment light accuracy test (lasers)
Set device up to lap zero
Load preset beams and physically confirm lasers adjust to the correct position
Are CT lasers movable or fixed?
Movable
Laser/image/treatment isocenter coincidence daily tolerance for non-SRS, non-SBRT, and SBRT
2mm
Laser/image/treatment isocenter coincidence daily tolerance for SRS
1mm
Who performs laser/image/treatment isocenter coincidence daily checks
Therapist