Calibration of dental x-ray equipment can be performed by a dentist dental hygienist and dental assistant. T/F
false
Annual test for dental x-ray machines can be performed by a dentist dental hygienist and dental assistant
True
For quality control purposes each new box of unopened film should be tested for film freshness and fogging before it is used
True
After processing fresh film that has been properly stored and protected will appear
Clear with a slight blue tint
After performing the film screen contact test, a wire mesh image of uniform density appears. These results indicate what
Adequate film screen contact
When functioning properly, a view box should emit a uniform and brilliant light
False
One of the most critical areas of quality control that requires daily monitoring is what
Processing of films
The coin test is used to check
Proper safe lighting
The following must be closely monitored with manual processing techniques
Temperature of the water bath, levels of the processing solutions, accuracy of the timer, and accuracy of the thermometer
On the average, processing solutions must be changed
Every 3 to 4 weeks
On the average processing solutions should be replenished
Once each day
Fresh films and fresh chemicals must be used when preparing reference radiographs
True
Reference radiograph
A radiograph that is processed under ideal conditions and then used to compare the film densities of radiographs that are processed daily
The densities seen on the daily radiograph appear lighter than the densities seen on the reference radiograph these results indicates that
the developer soln is too cold
The clearing test is used to monitor
Fixer strength
Regardless of who actually exposes and processes radiographs, the dentist is ultimately responsible for the diagnostic quality of all dental radiographs
True