QAQC: Midterm

0.0(0)
studied byStudied by 0 people
0.0(0)
full-widthCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/61

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

62 Terms

1
New cards

FILTRATION

The most important patient protection characteristics of a radiographic imaging system

2
New cards

white x-ray

low energy x-rays is aka

3
New cards

reduce patient dose

primary function of filtration

4
New cards

filter out low energy x-rays

secondary function of filtration

5
New cards

inherent filtration

added filtration

total filtration

3 TYPES OF FILTRATION

6
New cards

inherent filtration

filtration that is a result of the composition of the tube and housing

7
New cards

added filtration

The filtration that is added to the port of the x-ray tube.

8
New cards

total filtration

the sum of inherent and added filtration

9
New cards

0.5

inherent filtration constant

10
New cards

half-value layer (HVL)

Filtration is Measured By:

11
New cards

COLLIMATION

X-ray field must coincide with the light field of the variable-aperture light-localizing collimator

12
New cards

tape measure

Use to measure collimation

13
New cards

marked on the x-ray tube housing

Location of Focal Spot:

14
New cards

Collimation

The restriction of the size and shape of the x-ray beam in order to reduce patient exposure.

15
New cards

Focal Spot Size (FSS)

It principally determines the spatial resolution of a radiographic imaging system

16
New cards

pinhole camera

star pattern

slit camera

FOCAL-SPOT SIZE is measured by:

17
New cards

Pinhole Camera

It is difficult to use

It requires excessive exposure time

18
New cards

star pattern

It is easy to use

19
New cards

slit camera

The standard for measurement of focal-spot size

20
New cards

3 mm

Star Pattern Limitation for Focal-Spot Sizes less than _____

21
New cards

kVp Calibration

Type of calibration that must be done whenever high-voltage generator components have changed significantly.

22
New cards

kVp Calibration

The x-ray generator must be properly calibrated

23
New cards

ion chamber

photodiode

Device used in kVp Calibration

24
New cards

photodiode

Converts light into electron

25
New cards

cutie pie

ion chamber is aka

26
New cards

4%

Variation in kVp necessary to affect image OD & radiographic contrast

27
New cards

Exposure Time

responsible for patient dose & OD

28
New cards

ion chamber

photodiode

Device used in Exposure timer accuracy

29
New cards

EXPOSURE LINEARITY

The ability of a radiographic unit to produce a constant radiation output for various combination of mA & exposure time

30
New cards

dosimeter

EXPOSURE LINEARITY is determined by:

31
New cards

EXPOSURE REPRODUCIBILITY

The ability of a radiographic unit to produce a precisely the same radiation exposure when all the techniques are changed & returned to previous value

32
New cards

1. Select a combination of technique factors & hold them constant for a series of 10 exposure

2. Make a series of at least three exposures at the same technique

factors, having changed technique controls between exposure

Two Methods of Evaluation for EXPOSURE REPRODUCIBILITY:

33
New cards

Film Illuminators

It should be analyzed photometrically

34
New cards

photometer

A device used to measure light intensity at areas of the illuminator

35
New cards

Screen-Film Contact Test

Cassette holders must be examined every month for worn closures, light leaks, and warping.

Screens should be cleaned monthly with commercially available cleaners

36
New cards

Radiographing a wiremesh pattern

Analyzing the image for areas of blur

Method of Evaluation for Screen-Film Contact

37
New cards

Replaced Felt or Foam Pressure Pad

if blur appears

38
New cards

Replaced Cassette

if replacement of foam or pressure pad does not correct the problem

39
New cards

Processor Cleaning

RT removes the transport and crossover racks & cleans them & the processing tanks with appropriate fluids

40
New cards

feed/entrance rack

developer rack

cross-over rack

fixer rack

water rack

dryer rack

RACKS IN AUTOMATIC PROCESSING:

41
New cards

cross-over rack

prevents contamination of developer from fixer

42
New cards

power of hydrogen

pH

43
New cards

90 seconds

Dry-To-Dry Time:

44
New cards

500 films/hr

Processing Rate:

45
New cards

Hydrometer

measures specific gravity

46
New cards

hygrometer

measures humidity

47
New cards

specific gravity

measures the activity of developer

48
New cards

35C / 95F

High development temperature

49
New cards

31C / 87F

Wash water temperature

50
New cards

SCHEDULED MAINTENANCE

Routine procedures that are performed usually weekly or monthly

51
New cards

PREVENTIVE MAINTENANCE

A planned program of parts replacement at regular intervals

It requires that a part be replace before it fails

52
New cards

Non-Scheduled Maintenance

The worst kind A failure in the system that necessitates processor repair

53
New cards

proper program of scheduled and preventive maintenance

Non-Scheduled Maintenance Minimize By:

54
New cards

1.000

specific gravity of water

55
New cards

1.030

specific gravity of developer

56
New cards

10-10.5 pH level

Normal pH level of solution

57
New cards

contaminated

if the pH level of solution decreased 0.4 the solution is _____

58
New cards

oxidized

if the pH level of solution decreased 0.8 the solution is _____

59
New cards

oxidation

loss of electron

60
New cards

replace solution

once ph level is too lo or too high:

61
New cards

over diluted

if the specific gravity is low =

62
New cards

too concentrated

if the specific gravity is high =