1/21
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
|---|
No study sessions yet.
Ohm’s Law
voltage, current and resistance
energy and power
clinical relevance
strength duration curve
rheobase
chronaxie
clinical relevance
Ohm’s Law Equation
V= I*R
I = V/R
R= V/I
V(V), I (A), R (Ω)
Fundamentals of Electronics
voltage (V): force moving current- electromotive push
current (I): flow of electrons through a circuit, mA
resistance (R): Impedance- opposition to current flow: unit of measure (Ω)
System Impedance
cathode electrode impedance (tip)
anode electrode impedance (ring)
polarization impedance
tissue impedance
typical range for a pacing lead impedance: 300-1200+ ohms
Current Flow
low resistance: high current flow
high resistance: low current flow
Lead fracture results in what?
decrease in current drain
Equation
E= V2*t/R
doubling the voltage is 4x the energy and doubling the PW is 2x the energy
Charge
in pacing, the charge describes the amount of electricity delivered to the myocardium during an atrial or ventricular output
charge (Q) is the product of pulse duration (t) in ms and output current (I) in mA. In pacing it is usually expressed in microcoulombs
Power
the rate of doing work or transmitting energy
power (P) = I*V; where P is in watts, I is in amperes and V is in volts
Energy
the ability to do work; for batteris usually expressed in watt-hours
Energy Density
the energy content of a battery or capacitor based on volume or mass. For a battery it is expressed in units of watt-hrs per cubic cm (Wh/cm3) or in watt-hr per gram1 (Wh/g1)
Watt
international unit of electrical power, watts are equivalent to joules per second (J/s)
Current Drain
the average amount of current drawn from a battery by the external load; typically, pacemakers draw from 10-30 mA2
Quiescent Current
housekeeping-current utilized when not pacing
i.e., current used for ongoing tasks like monitoring the intrinsic rhythm and data collection
Typically, the 8-12 uA range for pacemakers
markers, intervals, diagnostics, bluetooth, RF
Factors Affecting Current Drain
lead/tissue interface
stimulation threshold and safety margin desired
quiescent current
impedance
electrode design and material
pacemaker programming
% pacing
Factors Affecting Pacemaker Longevity
battery capacity
rate of drain
circuit efficiency
pacing rate
parametier: programmed output and pulse width
pacing: % of the time
Factors You Can Control
implant
parameters
ethics
Implant CSIT(T)
current of injury
sensing
impedance
threshold
Parameters
output (V); pulse width (ms)
rate
diagnostics
Ethics
“right is always right”
Pacing Rate
highest consumption of battery