1/61
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
Automation
the process by which analytical instruments perform multiple tests with minimal involvement of an analyst
controlled operation of an appartus, process, or system by mechanical or electronic devices with minimal human intervention
total laboratory automation
refers to a comprehensive, automated system encompassing the preanalytic, analytic, and postanalytic phases of testing
Automation advantages
assists technologists in test performance
assesses results of the tests performed
reduces or eliminates monotonous and repetitive tasks
single channel analysis
each specimen is subjected to a single process so that only results ofr a single analyte are produced
ex. osmo
multiple-channel analysis
each specimen is subjected to multiple analytical processes so that a set of test results is obtained on a single specimen
ex. CMP
parallel analysis
all specimens are subjected to a series of analytical processes at the same time and in a parallel fashion
centrifugal analyzers use discreet pipetting to load aliquots of specimen and reagents sequentially in to discreet chambers
the specimens are analyzed in parallel
analyzer set up for parallel analysis
either
a multiple specimen/single chemistry mode (running the same test on several specimens)
single specimen/multiple chemistry mode (running multiple tests in parallel on one specimen)
sequential analysis
each specimen enters the analytical process one after another
each result or set of results emerges in th same order as specimens are entered
batch analysis
many specimens are grouped in the same analytical session
only one analyte assay is ran at a time
different assays may be ran on an analyzer by swithcing reagents
random access analysis
most common configuration of analyzers
any specimen, by comman to the processing system, is analyzed by any available process in or out of sequence with other specimens
each specimen may be analyzed for a different selection of tests
What is the error rate of human entry/misreading of specimen numbers?
1 in 300
What is the error rate of barcode misreading specimen labels?
1 in 1 million
Liquid level sensors
minimizes splatter
continuous flow
sample is aspirated though the smaple probe into a stream of flowing liquid, whereby it is transported to analytical stations in the instrument
peristaltic pumps
discete anlysis
samplee is aspirated by sample probe and delivered, often with reagent, through the same orifice into a reaction cup or another container
positie liquid displacement pipettes
fixed volume pipettes
used for delivery of samples and reagents in analyzers that perform a single assay or small variety of tests
selectable-volume pipettes
allow for the selection of a limited number of predetermined volumes
used in systems that allow many different applications
carry-over
transport of a quantity of analyte or reagent from one specimen reaction into a subsequent one
adequate flush-specimen ratio and wash stations reduce
in discrete systems, caused by pipetting system
dispoable sample probe tips eliminate:
contamination of one sample by another inside the probe
carryover of one specimen into the specimen in the next cup
closed analyzers
can only use reagents from a particular manufacturer
open analyzers
able to use in-house reagents or reagents purchased from suppliers different from the analyzer’s manufacturer
factors that affect the chemical reaction:
the vesssel in which the reaction occurs (RV)
the cuvet in which the reaction is monitored
the timing of the reaction
mixing and transport of reactants
thermal conditioning of fluids
bidirectional interface
analyzer can receive data from the patient sample barcode and also transmit results to the LIS in electronic formats
auto-verification
automatically releasing results that have no associated problems
A tray of clear sample cups sits on a lab bench top for an extended period of time. Which one of the following analytes might be affected?
bilirubin
manufacturer-recommended sequence of events
preventative maintenance
reagent loading
calibration
QC
operating instrument
evaluating diagnostic messages
immunoassay
an assay based on the reaction of an antibody that is specific for an antigen
most commmonly used IgG
immunoassay label
a substance with measurable property attached to an antigen, antibody, or binding substance
easiest way to identify an assay
radioactive labels
first assay developed
labels used are radioactive isotopes
no longer used due to safety
enzyme labels
Ex: ALP, horseradish peroxidase, Beta- galactosidase
used for antigen or antibody and react with ligand
an enzyme substrate is added to produce a detectable color change
conversion of substrate is proportional to the quantity of antigen
enzyme label requirements
high amount of stability
extreme specificity
abscense from the antibody or antigen
no alteration by inhibitor with the system
fluorescnet labels (fluorochromes/phores)
compouds that absorb radiant energy of one wavelength and emit radiant enery of a longer wavelength
generally emitted light is detected at an angle of 90* from the path of excitation light using a fluorometer or modified spectrometer
fluorescence polarization immunoassay ises polarized light to excite label
stokes shift
the difference between the excitation wavelength and emission wavelength
Luminescent labels
unitilizes label to emit a photon of light as the result of an electrical, biochemical, or chemical reaction
examples
luminol
acridium ester
competitive immunoassay
reactants are either simultaneously or sequentially mized together
simultaneous method immunoassay
labeled antigen and unlabeled antigen simultaneously compete with the antibody
the avidity of the antibody for labeled and unlabeled antigen must be the same
bound label is inversely proportional to the concentration of unlabeled antigen
sequential method immunoassay
step 1: unlabeled antigen from the patient is mixed with excess antibody; binding is allowed to reach equillibrium
step 2: labeled antigen is the added sequentially and allowed to equilibriate; after separation the bound label is measured and is used to calculate the unlabeled antigen concentration
noncompetitive immunoassay (excess reagent, two-site, or sandwich assay)
immobilized, unlabeled antibody captures the antigen of interest
a washing removes unreacted molecules
the labeled detector antibody (conjugate) is added
another washing removes free labeled detector antibody
the signal from the bound labeled antibody is proportional to the antigen captured
homogenous assays
does not require a physical separation of the free and bound labeled analyte
the activity of expression of the label depends on whether the labeled reactant is free or bound
it does not require a physical separation step such as washing
heterogenous assay
uses a physical separations step to separate free and bound labeled analyte
the labeled, unbound analyte is separated or washed away and the remaining labeled bound analyte is measured
solid phase
in heterogenous assays physical separation is achieved by interaction with this
solid particles, beads, trays, or tubes onto which antibody or antigen is absorbed
used to immobilize reagent antibody or antigen in order to separate free from bound labeled reactant after washing
adsorption
most common separation technique is the use of paramagnetic particles that can be quickly immobilized to a solid phase by application of a magnetic field
separation is accomplished by wash steps that occur while the magnetic particles are immobilized by a magnet
Adsorption technique
unbound constituents and labels are removed by aspiration
the magnet is removed
additional reagents are added to produce an alaytical signal
Enzyme linked immunosorbent assay (ELISA)
heterogenous immunoassay with enzyme label
use solid phase as separation technique
ELISA formats
competitive assay using labeled Ag
competitive assay using labeled Ab
noncompetitive assay using labeled Ag
noncompetitive assay using labeled Ab
Enzyme Multiplied Immunoassay Technique (EMIT)
homogenous assay
unlabeled Ag competes with labeled Ag for Ab binding sites
as the concentration of unlabeled Ag increases, less enzyme labeled Ag can bind to Ab
enzyme is catabolically active when labled Ag is free
calibration
links the analytical signal with the concentration of analyte
purpose of curve is to establish a relationship between concentration of analyte and magnitude of signal given by measuring device
interpolation
connecting the points on the calibration plot to form the line or curve of best fit
establishes an expected signal range of concentrations of analyte that fall between the lowest and highest on the calibrator
analyte measurement range (AMR)
defines the lowest to highest measurable quantities
if the signal is below the AMR…
the result is typically reported as less than the lower end of the AMR or less than the lowes calibrator used in the clinical laboratory
if the signal is above the AMR…
the result may be reported as greater than the upper limit of the AMR or greater than the highest calibration used in the lab
sample may be diluted to bring analyte to concentration within AMR and reanalyzed (the multiplied by dilution factor)
hook effect
excess antigen binds to free labeled antibody, prohibiting the labeled antibody to bind to the capture anitbody
human anti-mouse antibody
are found in individuals who are exposed to mouse antigens and produce FP
heterophile antibodies
are formed in patients who have autoimmune diseases and other disorders and produce FP
point of care testing
analytical patient testing activities provided within the institution but performed outside the physical facilities of the clinical lab
point of care testing locations
emergency departments, operating rooms, ICUs
outpatient clinics, physician offices
urgent care, nursing homes, pharmacies
counseling centers, ambulances
main advantages of POCT
smaller sample volumes
more efficient work flow
recued TAT
reduced cost of overall patient care
improved patient outcomes with immediate result access
disadvantages of POCT
more costly
less accuracy
central lab and POCT results not always comparable
result integration to EMR challenging
numerous operators to train
difficulties with documentation of test results, billing, regulatory compliance
maintenace of QC and quality assurance of test results difficult
lot numbers, reagent and control expiration can be problematic
CLIA categories of testing
waived tests
moderate complexity tests
high complexity tests
non-medical lab tests
CLIA certificate types
certificate of waiver(waived tests)
certificate of registration(moderate/high tests)
certificate of PPMPs(waived tests outside lab+microscopic)
certificate of compliance(inspection)
certificate of accreditation(billing)
key components of POCT design
operator interface
bar code identification system
sample and reagent delivery mechanisms
reaction cell
sensor
control and communications system
data management and storage
manufacturing requirements
connectivity
provides the ability to transmit patient results to the medical record