Tomography
A picture of a slice, imaging technique that creates detailed 3D, sectional, or "slice-by-slice" views of an object
Tomography Images
2D (Planar/Projection) --> Semiqualitative
3D (Tomogram) --> Qualitative
Sinogram
Transform of projection data, 2D representation of tomographic X-ray or PET data
SPECT Projection
Image @60-90 degrees for 20-40 seconds
PET/CT sensitivity vs SPECT
PET/CT Sensitivity = 100-1000x SPECT
Scintillation Detectors
Convert absorbed radiation into pulses of light
Ionization Chambers
Gas filled chambers that detect ion pairs
ROC Curve
Assesses the overall diagnostic performance of a test
Inverse Square Law
Intensity if inversely proportional to the square of the distance
Deadtime
Time for a detector to process an event
Dose Calibrator QC
Accuracy (Daily), Constancy (Daily), Geometry (At installation), Linearity (Quaterly)
How Scintillator works
Uses a PMT to amplify light signal using dynodes
Photopeak Energy Used
20%
Intrinsic
Without collimation
Extrinsic
With collimation
Useful FOV
Crystal area
CFOV
Inner 75% of crystal area (Clinical imaging area)
Gamma Camera Collimation types
Parallel hole, Diverging, Converging, Pinhole
Anger logic
used to determine the x,y position of an event
Gamma Camera QC
Energy resolution, Count rate performance, Uniformity, Sensitivity, Spatial resolution, Constancy, Multipeak registration
Flood field nonuniformity
Mistuning, uncoupled PMT, cracked crystal, crystal hydration, outdated correction tables
Gamma camera collimator does
directionalizes radiation by only letting perpendicular radiation to reach the crystal
________ resolution dominates __________ resolution
collimator, system
SPECT reconstruction algorithms
Filtered back projection (Analytic, fast), Iterative reconstruction (serial guesses, slow, better quality)
Fourier transform
expression of an image as a sum of sin waves
Nyquist frequency
Maximum spatial frequency
PET Positron negatron annhilation
To antiparallel 511 kev gamma photons are created
Coincidence detection
2 events received within the same timing window
2D PET perk
minimize scatter
3D PET perk
Maximum sensitivity
Time of flight PET
determine the location between two detectors at which annihilation occurs based on delta t
PET Corrections
Decay, Deadtime, randoms, normalization, scatter and annhilation
Pregnant woman risk
50% increase in childhood caner per RAD, no risk fo birth defects at diagnostic levels
MIRD
used to calculate absorbed dose from internally administer radiopharm. each organ is a source and target
In filtered back projection reconstruction of SPECT images
as the cut-off frequency Vc is reduced, the images become "smoother" (less noisy) but at the cost of poorer spatial resolution