1/46
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
---|
No study sessions yet.
pair production occurs at
1.022 Mev
Cosmic Radiation annually
0.3 mSv
Terrestrial radiation annually
2 mSv
Internal radiation annually
0.7 mSv
DX exposure annually
0.6 mSv
Dose recieved during air travel
0.005-0.01 mSv/hr
Quality factor of x ray and gamma
1
Quality factor of beta
1 msv
quality factor of fast neutrons
20
QF for alpha particle
20
weighing factor for gonads
0.2
Annual occupational effective dose limit
50 msv
CummEFD
10 msv x age
Occupatonal exposure to lens of ey
150 msv
occupational exposure to skin, hands, feet
500 msv
Frequent Effective dose limit
1 msv
infrequent effective dose limit
5 msv
infrequent skin, hand, feet
50 msv
Educational effective annual limit
1 msv
educational exposure to lens of eye
15 msv
embryo-fetus exposure limit (monthly)
0.5 msv
Entire gestation
5.0 msv
Shields must have _________ lead
0.5 mm
Entrance to skin dose
0.05 msv per procedure
GSD
0.20 msv
max leakage
1 mGy/hr at 1 meter
Skin to surface distance requirement
15 CM
Inherent filtration
0.5 mm al
Total filtration
50-70Kvp= 1.5 mmAl, 70+kvp= 2.5 mmAl, mobile unit= 2.5 mmAl
Source to skin distance for mobile
12 IN
Min permanently installed Filtration
2.5 mm Al ( 3.0 prefferred)
Source to skin distance for fluoro unit
15 IN
Primary protective barrier eqv
2 mm lead
scattered x ray intensity
1/1000 (0.1%)
apron
0.5 mm lead
Primary protective barrier
1/16 in
secondary protective barrier
1/32 in lead
control booth barrier
7 ft up from floor
lead glass window
1.5 mm lead
Gloves
0.25 mm lead
Thyroid shield
0.5 mm lead
glasses
0.35 mm lead
curtain
0.25 mm
bucky slot shielding
0.25mm
Exposure to primary beam use factor
1/4
ceiling use factor
1/16
Max weekly permitted EQD
Controlled: 1msv
Uncontrolled: 20 msv