16 Fluoroscopy pt 2

0.0(0)
studied byStudied by 0 people
0.0(0)
full-widthCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/34

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced
Call with Kai

No study sessions yet.

35 Terms

1
New cards
Digital fluoroscopy
Uses an analog-to-digital converter (ADC); ADC placed between TV camera and monitor; may use flat-panel detectors; some units operate at higher mA similar to radiographic mode; x-ray beam automatically pulsed to reduce heat loading on tube.
2
New cards
Analog-to-digital converter (ADC)
Placed between TV camera and display monitor; converts analog signal from camera into digital signal for processing.
3
New cards
Flat-panel detector (FPD)
More recent advancement; replaces image intensifier; considered the primary barrier; lighter weight; flexible positioning; FOV up to 43 cm × 43 cm; can rotate to match patient anatomy; more sensitive than image intensifiers; improved image quality.
4
New cards
Digital fluoroscopy standard of care
Digital flat-panel detector systems are the current standard, replacing traditional image intensifiers.
5
New cards
Advantages of flat-panel detectors
Reduce size/bulk/weight of fluoroscopy tower; easier manipulation; improved patient access; replace spot filming and recording devices; can operate in radiographic mode eliminating need for extra radiographs; images stored in PACS/MIMPS.
6
New cards
FPD technology (solid-state digital output)
Produces digital signal directly; no fiberoptic coupling; no “end of life” degradation; flat detector surface eliminates vignetting, pincushioning, distortion; improved contrast resolution.
7
New cards
FPD scintillator materials
Uses indirect design with scintillator layer: Cesium iodide (CsI) or Gadolinium oxysulfide (Gd2O2S).
8
New cards
FPD dose reduction
Flat-panel detectors provide significant radiation dose reduction compared to image intensifiers.
9
New cards
Automatic Exposure Rate Control (AERC)
Maintains image brightness by adjusting mA, kVp, filtration, and pulse width; provides quicker, more stable response in FPD systems.
10
New cards
RF system designs
Two designs: x-ray tube above patient OR x-ray tube below patient; both fully functional but scatter radiation patterns differ.
11
New cards
X-ray tube above patient
FPD located under tabletop; variable SID up to 72 inches; table tilt up to 90°; single detector used for both fluoroscopy and radiography; system often remotely operated.
12
New cards
X-ray tube below patient
Traditional design; facilitates table-side fluoroscopy; x-ray tube under patient and digital detector above; system may require additional digital detectors (table bucky, upright holder).
13
New cards
Scatter radiation distribution
Scatter most intense at waist level and above (upper trunk, neck, head); patient is primary source of operator exposure.
14
New cards
Scatter with x-ray tube below patient
Creates scatter area mainly at waist level and below.
15
New cards
Last image hold (LIH)
Displays last fluoroscopic image; visible lines indicate collimator plate position.
16
New cards
Virtual collimation
Allows operator to adjust collimation electronically without exposing the patient again.
17
New cards
Last frame hold / fluoro loop save
Saves a single frame or fluoro sequence; helps review without additional radiation.
18
New cards
Frame averaging
Reduces patient dose and image noise by averaging multiple frames.
19
New cards
AERC (fluoroscopy)
Automatically adjusts tube mA, kVp, filtration, pulse width to maintain constant exposure to detector.
20
New cards
Dose rate settings
Low, medium, high settings that adjust dose, often by set percentages (e.g., 50%).
21
New cards
Magnification with flat panel
Flat panel detectors improve image quality in magnification mode WITHOUT increasing patient dose (unlike image intensifiers).
22
New cards
Continuous fluoroscopy
Continuous x-ray exposure; increases patient dose.
23
New cards
Pulsed fluoroscopy
X-ray exposure occurs in pulses with gaps; decreases patient dose.
24
New cards
Pulsed fluoro optimization
Designed to provide good image quality at acceptable dose; uses high-frequency generator; controls pulse height (mA), pulse width, pulse interval, and pulse rate.
25
New cards
FPD operation (fluoroscopy)
Uses amorphous silicon photodetector layer; arranged in pixel array; operates using pulsed exposures; requires pixel binning during fast sequences.
26
New cards
Pixel binning
Combines detector elements (DELs) to speed frame rate during fluoroscopy at the cost of reduced spatial resolution.
27
New cards
Air kerma
Measures kinetic energy released in air; represents x-ray intensity at a point.
28
New cards
Cumulative air kerma
Reference dose measured in Gray (Gy); estimates patient skin dose during procedure; chamber intercepts entire beam for total calculation.
29
New cards
Kerma-area product (KAP/DAP)
Product of cumulative air kerma and x-ray beam area at patient entrance; measured in mGy·cm² or microgray; must be recorded along with fluoro time.
30
New cards
Display monitors
Flat-screen LCD monitors; black-and-white; high resolution (up to 5 megapixels); contain layered LCD structure with liquid crystal between polarizers.
31
New cards
Mobile C-arm units
Provide fluoroscopic imaging; used in OR and ED; C-shaped design holds tube and intensifier/detector; allows multiple viewing angles.
32
New cards
O-arm unit (closed design)
Provides static and dynamic imaging; supports 2D and 3D imaging; uses two FPDs in orthogonal orientation.
33
New cards
Radiation safety in fluoroscopy
Intermittent pulsing reduces exposure; use lowest pulsed mode and lowest dose setting; use time, distance, shielding; tabletop exposure must not exceed 10 R/min (≈88 mGy/min) with ABS/AERC; SSD ≥ 38 cm (15 in) for stationary units and ≥ 30 cm (12 in) for mobile units.
34
New cards
Shielding requirements
Lead aprons: 0.5 mm Pb; bucky slot cover: 0.25 mm Pb; protective curtain: 0.25 mm Pb placed between patient and operator.
35
New cards
Quality control program
Includes equipment performance inspection and testing; QC radiographer inspects equipment at least every 6 months; includes room condition inspection, cables, control booth structure, and wear/deterioration.