why do we do QA
to mantain radiation doses to a minimum
consistent image quality
ensure equipment is maintained
keep retake low
reduce servicer user error

what is the one used for
low contrast detail detectability (large circles)
sensitometric measurements (medium circles)
high contrast small detail detectability (smallest circles)
resolution (lines in the middle)

how do we use
SID and exposure should be repeatbale and reproducable
use 1mm Cu filter
place TOR CDR so line test is 45 degree angle to beam
expose
input

threshold for TOR CDR
+-2
resolution
-2
both no suspension level

which one is this and what does it measure
leed 18fg
low contrast large detail (circles at top and bottom)
circular geometry (circle around edge)
monitor brightness level and contrast level (sqaures)
resolution (centre)

how to use
same as TOR CDR

threshold
same as TOR CDR

whats this and
beam aligment and collimation test tool

how to use this
SID and exposure repeatable and reproducabkle
place in centre of IR
collimate to lead within test tool
input

threshold for this device
remedial ± 1cm @1m SID
suspension level ± 3cm @1 SID

what does this measure
dose
dose rate
kvp
hvl
exposure time
puleses

how to use
SID AND EXPOSURE BLAH BLAH
place test in centre
collimate to it
take a range and input (repeat)

threshold
remedial basline ± 10 percent
suspension baseline ± 20 percent
reproducibility
remadial +- 20 perc
suspension ± 50 perc
what is QA
monitor aspect of project or service to make sure its up to standards
QC definiton
maintain or improve quality and includes monitoring
purpose do we do qa and qc
reduce dose cost
mainitinag imaging standards
why do we do it qc and qa
to make sure we catch problems early on
ALARP principle
safety
meet IRR 1999 AND IR(ME)R 2017
WHATS IRR 2017 about section 32
commiosing - testing before first use (MP)
rotuine testing after major maintence (MP ABD STAFF)
radiation protecting dosimetry - regualr testing stuff
section 15 IRR 2017
Implemet and maintain qa
presrve instalations and provide it to who ask
whats ALARP section 12
keep dose low
medical physics expert section 14
medical physics is appointed and be involved in optimisatiomn
give advice on dosimetry
whats acceptance testing
before commsioning
one month before warranty end
to make sure image quality is satisfactoru
HCPC radiographers must
understand qa and qc
is relevant
participate
monitor
whats level a and level b staff
level a radiographers
b is physics team and engineers or specially trained radiographers
remedial level is
correction
suspenion
levrel to where we remove machine