Electrical Impedance
The measurement of charges in electrical resistance produced by cells as they traverse a small aperture; cells are suspended in conductive diluent then pulled through aperture in a tube.
Electrical Impedance Histogram
Determination of RBC count and MCV; allows calculation of Hct, MCH, and MCHC
2-20fL
The amount of platelets the histogram can separate
>36 fL
The amount of RBCs the histogram can separate
35-90fL
The amount of lymphs the histogram can separate
90-160fL
The amount of mononuclear cells the histogram can separate
160-450fL
The amount of granulocytes the histogram can separate
Protein Buildup
Lower cell counts and falsely elevated cell volumes in electrical impedance
Coincident passage of more than one cell
Falsely increased cell volumes and decreased cell counts
Flow Cytometers Light Scattering
Results based off process of absorption, diffraction, refraction, and reflection
Orthogonal Light
Scatter/side scatters allows furthers characterizing of cells giving differentiation of 5 types of WBCs based on volume and internal features
Cythochemical and fluorescent staining
Distinguishes segs, eos, and monos that are peroxidase positive from lymphs, basos, and PLTS that are peroxidase negative; dyes bind to DNA and RNA in WBCs, NRBCs, PLTs, and retics for identification
Cyanmetheomoglobin
The reference method for measuring Hgb
Beckman-Coulter, Sysmex, Abbott, and Siemens Healthineers
The multiple analyzer manufacturers
Hydraulics
The aspirating unit, dispensers, diluters, mixing chambers, aperture baths or flow cells or both, and hemoglobinometer
Pneumatics
Generates vacuums and pressures required for operating the valves and moving specimen through the hydraulics system
Electrical
Controls operational sequences of system and includes electronic analyzers and computing circuitry for data processing as well as a data display
Beckman Coulter
UniCel DxH 900 uses spectrophotometry, impedance, RF conductivity, & multiangle laser light scatter
Sysmex
XN-3100 system uses spectrophotometry, DC detection, & fluorescence flow cytometry with multiangle scatter & fluorescence detection
Abbott
CELL-DYN Sapphire uses spectrophotometry, impedance, & multiangle laser light scatter with fluorescence detection
Siemens Healthineers
ADVIA 2120/2020i uses spectrophotometry, light scatter, & cytochemistry
International Council for Standardization in Hematology (ICSH)
Published the guidelines for evaluation of blood cell analyzers
Limitations of Impedance Technology
Analyzer’s inability to reliably distinguish cells from other particles or cell fragments
Errors in Hemoglobin Determination
Due to spectrophotometric measurement, turbidity of specimen after RBC lysing the Hgb can be falsely increased
Errors in Retic Counting
Falesely increased retic counts can be due to RNA stain reacting with RBC inclusions, malarial parasites, giant PLT, and Hgb S & C
Specimen Integrity & Stability
Specimens should be properly collected, stability of CBC parameters is about 24 hours at RT & 48 hours refrigerated