1/12
Looks like no tags are added yet.
Name | Mastery | Learn | Test | Matching | Spaced |
|---|
No study sessions yet.
Cobalt leakage limits with beam ON
0.1% of the primary beam at 1 meter distance
Linac leakage limits with beam ON
0.1% of the primary beam at 1 meter distance (same as cobalt)
What would be the maximum allowable leakage from a Cobalt source that emitted 193.3cGy/min?
0.1933cGy/ min (move decimal 3 places to the left)
A beam stopper is not to exceed __% of useful beam
0.1%
Transmission with beam OFF.
For linacs
zero
Transmission with beam OFF: for cobalt
We take readings at 14 different locations all 1 meter from the source
Dose shall not exceed:
2mR/ hr average
10 mR/ hr in any one direction
Cobalt and Cesium sources for teletherapy must be put on what to prevent leakage? What do they leak?
double encapsulated in stainless steel
beta particles
T/F: Since Cobalt and Cesium sources for teletherapy are double encapsulated they do not need to be tested for leakage
False, they do
When do cobalt and cesium sources used for teletherapy tested for leakage
by manufacturer before delivery
by hospital upon delivery
then every 6 months
What type of test is done to check for leakage of cobalt sources for teletherapy?
leak test aka wipe test
Steps for a Leak Test/ Wipe Test
Wipe collimator leaves with a long-handle cotton applicator moistened in a detergent solution
Measure the cotton applicator with a GM counter
FYI: the leakage, beta particles, will be transferred to the cotton applicator so you test that
Who licenses Co60 sources?
NRC
Calibration for Co60:
Spot checks
Full calibration
monthly spot checks
yearly calibration