stage micrometer function
to calibrate an eyepiece graticule so that it can be used to make measurements
calibrating an eyepiece graticule
select desired objective lens and magnification
line up scales of micrometer and graticule
count number of eyepiece graticule equivalent to each division in the stage micrometer
calculate the length of one division of the stage graticule
procedure for preparing a slide
stain sample with dye
mount sample on slide
place a coverslip on slide, avoiding air bubbles
magnification calculation
size of image / size of object
eyepiece graticule length calculation
1 Eyepiece graticule length = visible stage micrometer length / no. equivalent eyepiece divisions (that fit)