quality assurance
deals with people
QUALITY CONTROL
deals with instrumentation and equipment
➢ more tangible and obvious than QA
➢ designed to ensure that the Radiologist is provided with an optimal image produce through good equipment performance and resulting in minimal patient dose
DARK ROOM
primary function is to protect the film from white
light and ionizing radiation during handling and
processing
8X6 ft (2.6 X 2 m)
ideally with a floor area of not less than
40 TO 60 %
humidity must be maintained within
65 TO 75 DEGREES
temp must be maintained within
10
the ventilation system should include a source of fresh air
and yield __ air changes per hour
SAFELIGHT
should provide adequate illumination without
risk of exposure to film.
✓Amber safelight filters (WRATTEN 6B)
can be used with blue‐violet‐sensitive film.
Pure red filters (GBX)
must be used with ortho film.
AMBER FILTER
transmits light with wavelengths of 550nm and
longer
RED FILTER
transmits light with wavelengths above 600nm
both green and blue sensitive film
40%
Static artifacts possible when humidity dips below__
EXPOSURE
Latent image created
DEVELOPMENT
Converts latent image to black
metallic silver.
WASH
Removes excess developer
LATENT IMAGE FORMATION
formed by deposits of free ( ioinized) Ag ions.
FIXING
Dissolves out unexposed silver halide crystals.
WASHING
Removes products of processing
DRY
Removes water.
RETICULATION
uneven expansion and contractions of the emulsion layer.