SIGLENT SDG1032X Plus

0.0(0)
studied byStudied by 0 people
full-widthCall with Kai
GameKnowt Play
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
Card Sorting

1/37

encourage image

There's no tags or description

Looks like no tags are added yet.

Study Analytics
Name
Mastery
Learn
Test
Matching
Spaced

No study sessions yet.

38 Terms

1
New cards

Power

Turns the generator on/off. When off, the unit is fully powered down.

2
New cards

USB Host (front)

Plug a USB flash drive to save/recall setups/waveforms and update firmware.

3
New cards

Display Area

Shows menus, parameter fields, status, and softkey options aligned with the buttons below

4
New cards

Numeric Keypad

Enters values directly (frequency, amplitude, offset, etc.). Use units from on-screen popups

5
New cards

Knob

Adjusts the highlighted value; in file dialogs, rotates to select items.

6
New cards

Direction (Arrow) Keys

Moves the cursor/highlight between fields and menu items.

7
New cards

Output (per selected CH)

Toggles the selected channel’s BNC output on/off. Long-press quickly toggles the output load setting between 50 Ω and Hi-Z.

8
New cards

CH1 / CH2 (Channel Switch)

Selects which channel you’re editing; parameters and the Output key apply to that channel.

9
New cards

Mode / Auxiliary Function Keys

Shortcuts into advanced menus: Modulation, Sweep, Pulse Train, File Manager, and System Info

10
New cards

Waveforms

Opens the waveform selector (Sine, Square, Pulse, Ramp, Noise, Arb, Harmonics, etc.).

11
New cards

Menu Softkeys (under display)

Context keys mapped 1-to-1 with titles on the bottom of the screen; press to open/edit that item.

12
New cards

Rear Panel - Counter / Aux / 10 MHz / USB-Device / LAN

Frequency-counter input; trigger/sync/mod I/O; 10 MHz ref In/Out; PC control via USB-Device; LAN for remote control. (Not buttons, but handy I/O.)

13
New cards

1 kHz, 2 Vpp sine on CH1 (Hi-Z load)

CH1WaveformsSineParameterFrequency = 1 kHzAmplitude = 2 VppOffset = 0 V → (optional: UtilityOutput SetupLoad = HighZ or long-press Output 2 s to toggle 50 Ω/Hi-Z) → Output (CH1) ON.

14
New cards

AM (1 kHz modulation) on CH1

Set carrier first: CH1WaveformsSineParameter → set Frequency/Amplitude → then ModulationAMSource = InternalMod Freq = 1 kHzDepth (e.g., 50%) → Output ON.

15
New cards

Linear sweep 500 Hz → 1.5 kHz in 1 s (CH1)

Set base waveform (e.g., Sine) → SweepType = FrequencyMode = Linear (Up)Start = 500 HzStop = 1.5 kHzSweep Time = 1 sTrigger Source = InternalOutput ON.

16
New cards

N-cycle burst (5 cycles) on CH1

Pick waveform & carrier params → BurstMode = NCycleN cycles = 5Trig Source = Internal (or Manual for front-panel trigger) → (optional Burst Period) → Output ON.

17
New cards

Gated burst using CH2 as the gate (CH1 = carrier)

CH1: set waveform/freq/amp → BurstMode = GatedSource = External → CH2: make a square gate (0–5 V typical) → cable CH2 → Rear Ext Trig/Gate InOutputs ON. (Gate high = output, gate low = off.)

18
New cards

Toggle output load (50 Ω Hi-Z) quickly

Press & hold the channel’s Output key for ~2 s to switch between 50 Ω and Hi-Z. (Or: UtilityOutput SetupLoad.) Match this to your actual load so amplitude readings are correct.

19
New cards

Save current setup to USB

Insert USB → Save/Recall (or File Manager) → choose Save Setup → pick location/filename → Save. (You can also save waveforms and screenshots.)

20
New cards

Quick 10 MHz, 20 Vpp sine w/ gated burst (example)

WaveformsSineParameterFreq = 10 MHz, Amp = 20 VppBurstGated, Source = External, Polarity = Positive → wire external gate → Output ON.

21
New cards

Sanity check when no signal appears

Verify channel Output = ON, BNC and load OK, and (if amplitudes look off) confirm Load = HighZ/50 Ω is correct for your measurement path.

22
New cards

FM (Frequency Modulation) - set up

CHx → pick carrier (e.g., Sine; set freq/amp) → ModOnModulation Type = FMSource = Internal (or External/CHy) → set FM Freq (modulating frequency) → set FM Dev (Δf). Output ON. (FM Dev = max ±freq deviation around carrier.)

23
New cards

PM (Phase Modulation) - set up

CHx → set carrier → ModOnType = PMSource (Internal/External/CHy) → set PM Freq and Phase Dev (Δφ). Output ON. (Δφ is max phase swing.)

24
New cards

FSK (Binary Frequency-Shift Keying) — set up

CHx → set carrier (this is f₀) → ModOnType = FSK → set Key Freq (bit rate/clock) → set Hop Freq = f₁ (the “1” frequency; “0” uses the carrier f₀). Choose Source = Internal (clock) or External (rear port). Output ON.

25
New cards

N-cycle burst every 10 ms (example)

Set carrier (e.g., Sine 10 kHz) → BurstOnBurst Type = N-loopTrigger Source = InternalBurst Period = 10 msLoops = 5 → (optional Trig Out = Rising) → Output ON.

26
New cards

Manual burst of fixed pulses (example)

Carrier Pulse 10 kHz, Width = 20 µsBurstN-loopTrigger Source = ManualBurst Period = 10 msCycles = 10Pulse Trains = 3Trig Out = RisingOutput ON → press on-screen Trigger to fire.

27
New cards

External gate or trigger for burst

BurstTrigger Source = External → select Edge (rising/falling). Wire gate/trigger to rear Aux In/Out BNC. Output ON (gate high = output if in gated mode).

28
New cards

Copy one channel to the other (quick clone)

UtilityCH Copy CouplingChannel CopyCH1⇒CH2 (or CH2⇒CH1) → Accept. (Excludes coupling/tracking while active.)

29
New cards

Frequency/Amplitude/Phase coupling (lock relationship)

UtilityCH Copy CouplingChannel Coupling → enable FreqCoup / AmplCoup / PhaseCoup → choose Mode = Ratio or Deviation → enter value(s). Now when you change the reference channel, the other follows per the set ratio/offset. (Works with basic waves: Sine/Square/Ramp/Pulse/Arc.)

30
New cards

90° phase-locked dual-sine (example)

Set CH1 Sine 1 kHz → UtilityCH Copy CouplingChannel CouplingPhaseCoup = OnMode = DeviationPhaseDev = +90° → set CH2 to Sine; tweak CH1 phase and frequency—CH2 stays +90°.

31
New cards

Track mode (make CH2 mirror CH1)

UtilityCH Copy CouplingTrack = On. Now CH2 auto-matches CH1 (same waveform/params/status). (Mutually exclusive with Copy/Coupling menus.)

32
New cards

Sync output (rear)

UtilitySync = On → rear Aux In/Out outputs a CMOS sync at the basic/modulation frequency (≤ 10 MHz; excludes Noise/DC/ext-mod). Use it to trigger your scope.

33
New cards

Load arbitrary waveform from USB (CSV/DAT)

Insert USB → Store/RecallFile Type = Data File → browse USB Device (0:) → select .csv or .datRecall/Load. The generator converts to .bin and opens the Arbitrary waveform screen. Output ON.

34
New cards

Copy arbitrary file from USB to internal memory

Store/RecallFile Type = Data → browse USB Device (0:)Copy → browse Local (C:)Paste. (Use English letters/numbers/underscores in filenames.)

35
New cards

Pick Arbitrary waveform mode (DDS vs TrueArb)

In Arbitrary screen → Mode = DDS (traditional) or TrueArb (point-by-point with variable sample rate; lower jitter). Then set freq/amp/offset as needed.

36
New cards

Make/import ARB with EasyWaveX (PC)

Create/edit waveform in EasyWaveX → save/transfer to instrument via USB-Device/LAN → the SDG stores as .bin and lists it under Local (C:). Then select it in Arbitrary and Output ON.

37
New cards

Set the right output load

Match the Load (50 Ω vs Hi-Z) to your measurement path for correct Vpp. Quick toggle: press & hold Output (per-channel).

38
New cards

File gotchas

USB filenames should use A–Z, 0–9, underscore only; other characters may not display.